ENTERING THE CALIBRATED LEVEL
32
N2 Compact Nitrogen Gas Generator - User Guide. ©2020.
-
Navigate to menu 3.2 and press .
-
Using the and keys enter the purity of the calibration gas.
-
Press to send the calibration level to the O Analyser.
-
On successful completion of the calibration the new O reading will be shown on the bottom line of the display.
-
If the calibration is not successful the original reading from the analyser will be loaded. Should this occur repeat the above steps.
-
On completion of the calibration, return the ball valves back to their original position and remove the regulated calibration gas
supply as applicable.
-
Navigate to menu 3.1 and disable the O Alarm Over-ride.
-
When returning to the main operating menu, “CAL” will be shown on the top line of the display. Th
is will remain for a period of
twenty minutes after the calibration. Th
roughout this time period the O alarm is overridden, to allow the sensor(s) to return to the
required level.
Summary of Contents for N2 NITROSource Compact N2C-2
Page 4: ...04 N2 Compact Nitrogen Gas Generator User Guide 2020 SAFETY...
Page 7: ...07 N2 Compact Nitrogen Gas Generator User Guide 2020 RECEIVING AND INSPECTING THE EQUIPMENT...
Page 10: ...10 N2 Compact Nitrogen Gas Generator User Guide 2020 OVERVIEW OF THE EQUIPMENT...
Page 12: ...12 N2 Compact Nitrogen Gas Generator User Guide 2020 LOCATING THE EQUIPMENT...
Page 14: ...14 N2 Compact Nitrogen Gas Generator User Guide 2020 INSTALLATION AND COMMISSIONING...
Page 20: ...20 N2 Compact Nitrogen Gas Generator User Guide 2020 OPERATING THE GENERATOR...
Page 27: ...27 N2 Compact Nitrogen Gas Generator User Guide 2020 SERVICING THE GENERATOR...
Page 35: ...35 N2 Compact Nitrogen Gas Generator User Guide 2020 TECHNICAL SPECIFICATION...