Parker Zero Air Generators 

Models HPZA-3500, HPZA-7000, HPZA-18000, HPZA-30000

 

 

Bulletin TI-HPZA-3500H

 

 

     

 

 

 

Operation 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Shutdown

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
The Parker Zero Air Generator is designed to operate continuously, 24 hours per day as long as the 
compressed air or power supply is not interrupted. The LED indicator lights on the front panel of the 
Parker Zero Air Generator give the operator instantaneous feedback regarding system function. The 
“Start-Up” (yellow) and “THC” (yellow and green) LEDs show system operational status; the “Overflow” 
(yellow) and 

“Check System” (red) LEDs show the need for operator attention or system maintenance. 

 

During normal operation, the display on the Parker Zero Air Generator will have the 

green “THC” LED 

illuminated.  If the 

“Overflow” LED, “Start-Up” LED, “Check System” LED, or yellow “THC” LEDs illuminate 

during routine operation, please consult the Troubleshooting Guide at the end of this bulletin. 

 

The 

“Check System” LED will blink occasionally during normal operation. The blink rate will increase as 

the system approaches 3 years of operation. At that time, the catalyst module should be replaced. 

 

If the 

“Check System” LED is solid red, please consult troubleshooting guide. 

 
The equipment shall not be placed too close to the wall so that it does not hinder the removal of the power 
cord. Do not use a power cord below the specifications of the provided power cord. 
 
To prevent excess demand on the Parker Zero Air Generator, use a flow control device downstream from 
the unit. If the flow capacity of the generator is exceeded, the 

“Overflow” LED and yellow “THC” LED will 

illuminate, indicating that hydrocarbon content exceeds 0.1ppm. 

 

Note:  To ensure hydrocarbon content of less than 0.05 ppm (Models HPZA-3500,

 

HPZA-7000, and HPZA-18000), a flow control device must be installed downstream from the 
generator to prevent downstream demand in excess of the 

generator’s maximum flow rate.

 

 

To shut down the Parker Zero Air Generator, simply turn the power switch to the off position and turn off 
the compressed air supply to the generator. 
 
 

Maintenance 

All maintenance procedures for Parker Zero Air Generators should be performed by suitable 
personnel using reasonable care.

 

Prior to servicing the Parker Zero Air Generator, turn off the compressed air and power 
supplies to the generator, and ensure that the system is depressurized.

 

To ensure consistent product performance and reliability use only genuine Parker replacement 
parts and filter cartridges. 

The primary maintenance tasks required by the Parker Zero Air Generator are changing the prefilter 
cartridges (annually), replacing the final membrane filter cartridge (annually), and replacing the catalyst 
module (when the 

“Check System” LED illuminates or blinks). A summary of the replacement part 

numbers and recommended service frequency is shown at the end of this Maintenance section. See 
Figure 6 for the locations of the various maintenance items.

 

Replacement prefilter cartridges, final filter replacement cartridges, and catalyst modules may be 
ordered through a local representative. For convenience, a one-year supply of these replacement 
cartridges has been assembled into a maintenance kit (P/N MK7840). The kit contains one first stage 
coalescing prefilter cartridge (P/N 100-12-DX), one second stage coalescing filter cartridge (P/N 100-
12-BX), and one Grade GS final membrane filter (P/N GS-050-05-95).

 

If  necessary,  the  Zero  Air  Generator  may  be  wiped  clean  with  a  dry  cloth  on  an  as  needed 
basis.  Do  not  use  water,  aerosols, or  other  cleaning  agents  to  clean  the  units.  Use  of  any 
liquid detergent to clean the generator could present an electrical hazard. 

 

 
 
 

Summary of Contents for HPZA-18000

Page 1: ...e 1 are completely engineered systems which replace the use of inconvenient high pressure gas cylinders as a source of hydrocarbon free air The HPZA 30000 generators purify on site compressed air to a...

Page 2: ...rs are equipped with float drains which automatically open to empty any liquids from inside the filter housing The drains are connected to 1 4 O D plastic tubing which discharges to atmosphere onthe r...

Page 3: ...re taken into consideration Approximately 6 of space should be allowed along the top and sides of the generator to ensure proper ventilation Compressed Air The Parker Zero Air Generator requires a sou...

Page 4: ...ion only Piping Components The inlet and outlet ports for the Parker Zero Air Generator are 1 4 NPT female and located on right side of the unit Inlet tubing and fittings should be clean and rated for...

Page 5: ...stream from the generator Exposure of the generator to halogenated hydrocarbons will damage the catalyst module and void the warranty For additional information request the Analytical Gas Supplies cat...

Page 6: ...carbon content of less than 0 05 ppm Models HPZA 3500 HPZA 7000 and HPZA 18000 a flow control device must be installed downstream from the generator to prevent downstream demand in excess of the gener...

Page 7: ...Replace the GS membrane cartridge GS 050 05 95 and re assemble The time required to service all three stages of filtration integral to the Parker Zero Air Generator is approximately 15 minutes To prev...

Page 8: ...tor may burn out The fuses are located in the power receptacle on the left hand side of the generator Before servicing the fuses turn the generator off and disconnect the power cord from both the powe...

Page 9: ...generator is exceeded Overflow LED and yellow THC LEDs will illuminate An excessive pressure drop greater than 4 psig 0 3 bar may accompany an overflow situation If demand exceeds capacity install a f...

Page 10: ...arg 40 psig 125 psig 2 8 barg 8 6 barg 40 psig 125 psig 2 8 barg 8 6 barg Min Max ambient temperature 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C Max relative h...

Page 11: ...calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or i...

Page 12: ...800 343 4048 Symbol Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 242 Neck Road Haverhill MA 01835 www parker com labgas Description Caution Refer to accompanying docum...

Reviews: