Bulletin TI-72730J 

Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

1

1-800-343-4048

These instructions must be thoroughly read and understood before installing and oper-
ating this product.  Failure to operate this product in accordance with the instructions 
set forth in this manual and by other safety governing bodies will void the safety certifi-
cation of this product.  If you have any questions or concerns, please call the Technical 
Services Department at 800-343-4048, 8AM to 5PM Eastern Time (North America only).  
For other locations, please contact your local representative.  Send email to:   
[email protected]

The Balston 72-730 and 72-O2730NA Oxygen Analyzers are designed to monitor the oxygen 
concentration in a process stream, display this concentration in digital form, and provide 
appropriate alarms and controls for protecting a process against undesirable oxygen con-
centrations.   The oxygen analyzer is a valuable accessory for food packaging, solvent and 
chemical blanketing, and purging applications.

Note:  In hazardous applications where the oxygen content is critical (i.e., blanket-
ing explosive chemicals or packaging food for extended shelf life), an oxygen moni-
tor   and/or trace oxygen analyzer should be used in conjunction with safety interlocks       
and/or alarm systems to assure proper nitrogen purity levels at all times.

The Balston Oxygen Analyzer is a self-contained unit which may be wall-mounted, bench-
mounted, or ordered as an integral accessory to any Nitrogen Generation System.  The unit is 
powered by 120 VAC or 240 VAC, 50/60Hz. (

Note:

 Main supply voltage must be within 10% 

of nominal rated voltage for the generator.)

The Balston oxygen analyzer has been certified to IEC 1010 Standards (CSA 22.2 No. 
1010.1-92) and bears the CSA safety marking on the product label.

The sensing device designed into the Balston 72-730 and 72-O2730NA Oxygen Analyzers 
is a galvanic cell.  The Oxygen Analyzer has an internal temperature compensation circuit to 
provide accurate readings within a specified temperature range, with an accuracy of ±1% of 
the calibration gas concentration.

The Balston 72-730 and 72-O2730NA Oxygen Analyzers have all the controls necessary to 
assure safe and accurate monitoring of the oxygen concentration in a process stream.  The 
analyzers are equipped with the following controls and features (see Figure 1).

Description

Sensor

Controls

Figure 1 - Front Panel Controls

Installation, Operation, and Maintenance Manual

Balston

®

 72-730 and 72-O2730NA 

Oxygen Analyzer

Summary of Contents for Balston 72-730

Page 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Page 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Page 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Page 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Page 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Page 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Page 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Page 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...