• Press the On/Off button on the controller to
begin rinsing.
►
The device emits a short acoustic signal.
►
The LED on the button lights up green.
►
The rinsing function is executed.
• As soon as all of the rinsing fluid has run
through, switch the device off by pressing
the On/Off button again.
►
The device emits a short acoustic signal.
►
The LED in the button goes out.
• Release the tab on the underside of the
device.
• Disconnect the VELOXcare from the con-
troller.
• Release the tabs on the VELOXcare and open it.
• Take the medication reservoir out.
• Rinse the VELOXcare and the medication reservoir thoroughly with tap water.
The medication reservoir must always be disinfected immediately after rinsing in
the VELOXcare.
If necessary, the VELOXcare can be cleaned and disinfected together with the
nebuliser parts.
Disinfection
The nebuliser, the mouthpiece and the mask must be disinfected
once a week
imme-
diately after cleaning. Only parts that have been cleaned can be disinfected effect-
ively.
CAUTION
A damp environment may encourage the growth of bacteria. Therefore, remove all
parts from the pot or disinfector as soon as disinfection has finished. Dry the parts.
The
risk of infection
is reduced when the parts are dried completely.
– 50 –