5. To begin your treatment, sit in a relaxed, upright position. Press and

hold the On/Off button for one second (Figure 17). You will hear one
“beep” and the status light will illuminate green. The Trio Electronic
Nebulizer is now ON.

6. After a few seconds, aerosol mist will begin to flow into the Aerosol

Chamber.

7. Place the Mouthpiece on top of your bottom lip and tongue, and close

your lips around it (Figure 18). Hold the Trio Electronic Nebulizer level.
If held at an angle over 45 degrees, the Trio Electronic Nebulizer will
sound two beeps (and two green lights) and shut off after approximately
30 seconds. If this occurs, hold the Trio Electronic Nebulizer level and
press the start button to complete the treatment.

8. Take slow, deep breaths in and out through the Mouthpiece. Each

breath should be a little deeper than normal, quiet breathing, but not
to your full capacity. Do Not breathe through your nose. Continue to
inhale and exhale comfortably until the treatment is finished.

9. When all of the medication has been delivered:

a. You will hear the “Treatment Complete Alarm” (2 beeps and

2 green lights).

b. The Trio Electronic Nebulizer will automatically shut off.

10. If for any reason you must stop the treatment before completion,

press and hold the On/Off Button for one second. A 1 beep alarm will
sound with 1 red light. To re-start the treatment, press and hold the
On/Off Button for one second and then resume the treatment.

11. After you have completed your treatment, disassemble the Trio Nebulizer

Handset for cleaning. See Section H for full cleaning instructions.

G

TAKING A TREATMENT (CONT)

IMPORTANT: Your medication should be at room temperature before use to
ensure that the Trio Electronic Nebulizer operates properly.

2. Pour only the medication prescribed by your physician specifically

for the Trio Electronic Nebulizer into the Medication Reservoir (Figure
15). 

DO NOT USE MEDICATIONS PRESCRIBED FOR OTHER NEBULIZERS

IN THE TRIO ELECTRONIC NEBULIZER. NOTE: MAXIMUM VOLUME FILL
IS 4 ml. DO NOT OVERFILL THE MEDICATION RESERVOIR.

3. Close the Medication Reservoir by aligning the tabs on the inside of

the Medication Cap with the tab slots in the Medication Reservoir.
Turn the cap clockwise as far as it will go (Figure 16).

15

Figure

16

Figure

17

Figure

18

Figure

                                              STATUS LIGHT

Summary of Contents for Trio

Page 1: ...ral law restricts this devicetosalebyoronorderofaphysician Read and understand these Instructions for Use and all safety precautions it contains Improperusecancauseserious or fatal injury and or illne...

Page 2: ...on and safety precautions for the Trio Electronic Nebulizer Save these Instructions For Use for future reference BeforeusingyourTrioElectronicNebulizer youshouldreadandunderstand the entire Instructio...

Page 3: ...NEVER operate the Controller if it is improperly or incompletely assembled or damaged Section G Taking a treatment shows the alerts to these situations when the Trio Electronic Nebulizer is improperl...

Page 4: ...o not have all the items please call the number listed below FOUR AA BATTERIES MEDICATION RESERVOIR AEROSOL CHAMBER MOUTHPIECE WITH EXHALATION VALVE AC POWER SUPPLY AEROSOL HEAD INHALATION VALVE CARRY...

Page 5: ...eries will provide approximately 2 hours of continuous treatment You may also use AA rechargeablebatteries Pleasesee Specifications ontheback cover of these Instructions For Use NOTE IT IS STRONGLY RE...

Page 6: ...fingers Figure 8 F TRIO NEBULIZER HANDSET ASSEMBLY WARNING TheTrioNebulizerHandsetisnotsterile CleanandDisinfecttheTrio NebulizerHandsetbeforethefirstuse seeSectionHforCleaningand SectionIforDisinfect...

Page 7: ...withTrioElectronic Nebulizer a Trio Nebulizer Handset Attached To Controller Place the Controller on a clean flat stable surface Slide the Trio Nebulizer Handset into the Controller cradle as far as i...

Page 8: ...en delivered a You will hear the Treatment Complete Alarm 2 beeps and 2 green lights b The Trio Electronic Nebulizer will automatically shut off 10 If for any reason you must stop the treatment before...

Page 9: ...bulizerHandsetimmediatelyaftereveryuse Donot wash the Controller Nebulizer Connection Cord AC Power Supply or PlugConverters Useliquidsoapdesignedforwashingdishes Donot use white liquiddishsoaps e g I...

Page 10: ...f clear liquid dish soap to a clean bowl or basin and add warm tap water Figure 28 5 Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under running warmtapwater Figure 29 If Disinfect...

Page 11: ...ain after the treatmentintheMedicationReservoir Ifmorethanafewdropsremain cleantheAerosolHead SectionH If afterpropercleaning morethan afewdropsremainintheMedicationReservoir call1 888 335 6946 QUESTI...

Page 12: ...batteries following the procedure in Section D Please note that the Controller has a built in automatic shutoff after 10minutes anditwillturnoffatthattime Ifthissituationcontinues three or four times...

Reviews: