INDICATIONS FOR USE: PARI TREK
®
S PORTABLE AEROSOL SYSTEM
The PARI Trek
®
S nebulizer compressor is a portable, DC powered air compressor intended to provide a source
of compressed air for use with jet nebulizers. The Trek
®
S is intended for adult and pediatric patients. The device is
nonsterile, prescription-use only, intended for use in hospital, clinic, or home environments.
WARNING
Use only with nebulizers and tubing authorized by PARI for
this device. Other nebulizers and tubing may result in improper
treatment.
When using electrical and medical products, especially when
children are present, basic safety precautions must
be followed.
DANGER
To reduce the risk of serious or fatal injury from electrocution:
1. Do not place or store the product where it can fall or be
knocked into bathtub, sink, water, or other liquid. Do not
place or drop into water or other liquid. Do not use while
bathing.
2. Do not reach for a product that has fallen into water or
other liquid. Unplug immediately.
WARNING
To reduce the risk of serious or fatal injury from electrocution,
fire or burns and to reduce the risk of damage and malfunction
to the product:
1. Before use, check the serial number label on the bottom of
the compressor, battery and on the Multi-Voltage
AC Adapter to ensure that the voltage and current
indicated on the unit correspond with the voltage and
current available.
2. Never operate this product if it has a damaged cord or
plug, if it is not working properly, if it has been dropped or
damaged, or if it has been exposed to any liquids inside the
case. Return the product to PARI for examination
and repair.
3. Keep all electrical cords away from heated surfaces. Do
not overload wall outlets or use extension cords.
4. Do not operate where oxygen is being administered in a
closed environment such as an oxygen tent.
5. Never drop or insert any object into any openings.
6. Be sure the PARI TREK
®
S compressor and the Multi-
Voltage AC Adapter are unplugged prior to cleaning.
7. Do not open or disassemble the PARI TREK
®
S
Compressor. Refer all servicing to an authorized PARI
Service Center.
8. Use only adapters that are authorized for the PARI TREK
®
S Compressor. Use of unapproved accessories can lead
to improper treatment or damage to the compressor,
adapters, or the battery.
9. Never spray liquids onto the compressor housing or
battery. Fluid could cause damage to the electrical and
compressor parts and could lead to a malfunction. In the
event that fluids enter the unit, return the product to PARI
for examination and repair.
10. Do not cover up or use the PARI Trek
®
S compressor
inside the carry case. Covering the device may cause
excessive heat buildup in the compressor housing.
WARNING
11. The compressor has auto-resettable thermal protection.
If the compressor shuts down due to temperature, it will
restart automatically when it cools off if the switch is not
set to the off position. Ensure the switch is set to the off
position if the compressor shuts off due to temperature.
12. Always unplug immediately after using.
13. Do not use the PARI TREK
®
S Compressor or the PARI
Reusable Nebulizer while operating a vehicle.
14. Portable and mobile RF communication devices can disrupt
electrical medical equipment. The electrical field generated
when the compressor is running can interfere with screens
and monitors (flickering). For this reason, the compressor
must be placed far enough away from screens/monitors
to prevent such interference. Technical electromagnetic
compatibility data is available in table form upon request
from PARI Respiratory Equipment Inc./Patient Connection
1.800.FAST.NEB (327.8632), press 2.
BATTERY SAFETY PRECAUTIONS
WARNING
1. Use only the PARI TREK
®
S Rechargeable Battery described
in these Instructions For Use for operating your PARI
TREK
®
S Compressor.
2. Do not immerse the PARI TREK
®
S Rechargeable Battery in
water.
3. Do not touch the battery with wet or moist hands.
4. Do not expose battery to direct sunlight, humidity, extreme
temperatures or chemicals.
5. Make sure battery poles do not come into contact with keys
or other metal objects, causing a “short circuit”.
6. Do not attempt to open the plastic case or to replace any
components. Do not disassemble Rechargeable Battery. The
Rechargeable Battery is a sealed unit and can not
be repaired.
7. Do not incinerate (burn).
CAUTION
1. To charge the PARI TREK
®
S Rechargeable Battery use only
the PARI TREK
®
S Multi-Voltage AC Adapter or optional
12 V DC Adapter. Using other adapters may damage the
battery.
2. In case of failure in charging or operating, unplug the
Rechargeable Battery from the Multi-Voltage AC Adapter
and turn the unit off. Should the Rechargeable Battery fail to
charge or operate properly, even if the Instructions For Use
are followed, notify the Authorized Service Center by calling
PARI Patient Connection toll free (USA) at 1.800.FAST.NEB
(327.8632), then press 2.
3. Protect the environment: when replacing the Rechargeable
Battery, recycle old battery or dispose of properly in
accordance with local laws.
B. SAFETY PRECAUTIONS - TREK
®
S PORTABLE AEROSOL SYSTEM
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/TrekS
EN-2