![Pari SINUS Instructions For Use Manual Download Page 49](http://html1.mh-extra.com/html/pari/sinus/sinus_instructions-for-use-manual_3717420049.webp)
- 49 -
en
6
TECHNICAL DATA
6.1 General
Compressor type
Double compressor:
Piston for generation of
compressed air
Diaphragm for generation of
bolus
Supply voltage [V]
Mains frequency [Hz]
Power consumption [A]
see back cover page
max. compressor flow (FreeFlow)
9.5 l/min.
Compressor flow
1)
4.8 l/min.
Sound pressure level
2)
approx. 56 dBA
Operating pressure
3)
150 kPa (1.5 bar)
Bolus pressure amplitude, absolute 20 kPa (0.2 bar)
Frequency of bolus pressure
amplitude
44 Hz
Housing dimensions [cm]
19.2 × 14.5 × 15.0
(W × H × D)
Weight
1.8 kg
Flow
4.5 l/min.
1) towards PARI nebuliser nozzle (Ø 0.48 mm)
2) Maximum A-evaluated sound pressure level of the compressor
(in accordance with DIN EN 13544-1, section 26)
3) with LC SPRINT SINUS nebuliser