Filling the nebuliser
NOTE
Nebuliser lid might break off
If the lid is twisted in the wrong direction, it may break off.
The nebuliser will then be unusable and irreparable.
• Never move the lid except in the direction allowed by the
hinge.
• Insert the nebuliser in the holder on the compressor
intended for this purpose.
• Open the nebuliser lid by pressing your thumb against the
underside of the lid.
• Pour the required quantity of
inhalation solution into the top
of the nebuliser.
Note the minimum and
maximum fill volumes
[see: General nebuliser data,
page 26]. If the nebuliser
contains too little or too much
liquid, the nebulisation and
consequently the therapy will
be less effective.
• Close the nebuliser lid.
Make sure that the lid snaps into place.
If several inhalation solutions are to be used one after the
other:
• Rinse the nebuliser out with drinking water between the
individual applications.
• Shake excess water out of the nebuliser.
• Fill the nebuliser with the next inhalation solution as
described.
– 17 –
Summary of Contents for LC SPRINT Junior
Page 2: ......