Step F4:
Gently pull up on the tab to open the
Handset and remove the Blue Valve (F-4).
Step F5:
Being careful to touch only the plastic outer ring
of the Aerosol Head, press the 2 flexible
plastic tabs on the side of the Aerosol Head
towards each other and remove (F-5). After
the Aerosol Head has been used 7 times, throw
away (dispose of) and replace with a new one.
Step F6:
Gently wipe away any drops of medicine
from the medication reservoir (F-6a), aerosol
chamber (F-6b) and mouthpiece (F6c) before
rinsing to reduce antibiotics added to water
systems.
Use only plain, dry paper towels or wipes.
Do not use towels or wipes that have any
chemicals added to them such as alcohol,
lotion, or baby wipes.
Be careful not to harm the parts.
Do not wipe the Aerosol Head.
Throw away paper towels by disposing in
trash with solid waste.
Step F7:
Rinse each of the parts under warm running
tap water for 10 seconds. Pay special attention
to rinsing the Aerosol Head and rinse each
side of the Aerosol Head for 10 seconds
(F-7). Thorough rinsing of both sides of the
Aerosol Head helps to prevent clogging and
makes sure the Aerosol Head works properly.
Never use a brush or any other object to clean
the Aerosol Head.
Cleaning Your Lamira (continued)
F
10 seconds
F-4
F-5
F-7
Do not touch the center silver part of the
Aerosol Head.
TAB
FLEXIBLE PLASTIC
TABS
F-6a
F-6b
F-6c
Reference ID: 4327567