ters) and the PARI compressor, as recommended below in ac
cordance with the maximum output power of the communica
tion device.
Safety distance depending on transmitting frequency:
Rated power
of the trans
mitter [W]
150 kHz to
80 MHz
D = (3.5/3) ×
SQRT(P) [m]
80 MHz to
800 MHz
D = (3.5/3) ×
SQRT(P) [m]
800 MHz to
2.5 GHz
D = (7/3) ×
SQRT(P) [m]
0.01
0.2
0.2
0.3
0.1
0.4
0.4
0.8
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters whose rated power does not appear in the
table above, the safety distance can be calculated using the
equation associated with the respective column, where "P" is
the rated power of the transmitter in Watt (W) according to the
information supplied by the device manufacturer.
6.5 Ambient conditions for operation
Ambient temperature
+10 °C to +30 °C
Relative humidity
30% to 75%
(noncondensing)
Atmospheric pressure
700 hPa to 1,060 hPa
– 44 –