PHYSIO-PORT
User Manual Version 2017-11-12
- 30 -
6.4
Programming recorder
After you have selected the
patient, a window will appear
for configuring and
programming the PHYSIO-PORT
Recorder.
Figure 41: Configuration of the measurement
Number of measurement intervals:
Up to 4 different measurement intervals can be selected. The standard setting
is at 2 intervals (day and night phase) .
Power supply:
The type (normal battery or rechargeable battery) that is used for long term
measurements is set.
From.. To..:
The beginning and end of selected intervals can be set in minutes.
Measurement interval:
Exact: measurements are taken at the exact intervals in minutes that have
been set.
Approx.: measurements randomly vary at approx. +/- 2 minutes around the
times set
Minutes:
The interval between subsequent measurements can be programmed to
range between 2 and 120 minutes.
Max. pump pressure:
Limits the max. cuff pressure to the value set between 200 mmHg and 280
mmHg.
Number:
The number of all programmed measurements will be displayed for each
measurement interval and a summary is provided for each 24 hours.
Display readings:
The blood pressure values are displayed on the LCD of the recorder.
Activate key lock:
Day and night phase can be switched.