7
© Copyright 2008 Panoramic Corporation
Pre-Installation Check
Electrical Requirements
Optimally, the PC-1000 and the PC-1000/Laser 1000 should have a dedicated 105-125 VAC, 20 A circuit
with line regulation of 5% or better. If a dedicated circuit is not available, a regular 20 A circuit will work
as long as it is not taxed by other loads beyond 5 A. A standard 115 VAC, three-wire, grounded, electrical
outlet should be installed by an electrician behind the machine.
Control Panel Orientation
The control panel on the PC-1000 and the PC-1000/Laser 1000 is mounted directly on the side of the
machine and requires no separate installation, wiring, or mounting. Although the control panel is mounted
on the patient's left side from the factory, it can be easily relocated to the opposite side during installation.
Verify with the doctor on which side the control panel should be located.
Cephalometric Arm Orientation
The cephalometric arm for the PC-1000/Laser 1000 is shipped in a separate box and can be mounted on
either side of the machine. Verify with the doctor on which side the cephalometric arm should be located.
Remote Switch
Some states and local governments require that the exposure switch to be remotely installed. A remote
switch kit is available from Panoramic Corporation. Refer to Appendix G for detailed procedures.
Darkroom Requirements
Proper darkroom facilities must be present before installing the PC-1000 or PC-1000/Laser 1000. The
darkroom will be used to process test fi lms at the time of installation to verify proper calibration. Refer to
Appendix A for darkroom requirements and processing procedures.
Summary of Contents for PC-1000
Page 39: ...39 Copyright 2008 Panoramic Corporation Notes...
Page 56: ...Appendix Copyright 2008 Panoramic Corporation G Final Assembly...
Page 58: ...Appendix Copyright 2008 Panoramic Corporation G Base Assembly...
Page 60: ...Appendix Copyright 2008 Panoramic Corporation G Main Chassis Assembly...
Page 62: ...Appendix Copyright 2008 Panoramic Corporation Chinrest Arm Assembly G...
Page 64: ...Appendix Copyright 2008 Panoramic Corporation Head Support Assembly G...
Page 66: ...Appendix Copyright 2008 Panoramic Corporation Film Drum Assembly G...
Page 68: ...Appendix Copyright 2008 Panoramic Corporation Upper Belt Drive Assembly G...
Page 70: ...Appendix Copyright 2008 Panoramic Corporation Belt Drive Assembly G...
Page 72: ...Appendix Copyright 2008 Panoramic Corporation Roller Orbit Drive Assembly G...
Page 74: ...Appendix Copyright 2008 Panoramic Corporation Rotating Arm Assembly G...
Page 76: ...Appendix Copyright 2008 Panoramic Corporation Screw Motor Assembly G...
Page 78: ...Appendix Copyright 2008 Panoramic Corporation Tubehead Assembly G...
Page 80: ...Appendix Copyright 2008 Panoramic Corporation Electronics Assembly G...
Page 82: ...Appendix Copyright 2008 Panoramic Corporation Control Panel Assembly G...
Page 84: ...Appendix Copyright 2008 Panoramic Corporation Laser Collimator Assembly G...
Page 86: ...Appendix Copyright 2008 Panoramic Corporation Electromagnet Assembly G...
Page 88: ...Appendix Copyright 2008 Panoramic Corporation H System Wiring Schematic...
Page 89: ...Appendix Copyright 2008 Panoramic Corporation Control Panel Schematic H...
Page 90: ...Appendix Copyright 2008 Panoramic Corporation Timer Circuit Board Schematic H...
Page 91: ...Appendix Copyright 2008 Panoramic Corporation X ray Control Schematic 1 3 H...
Page 92: ...Appendix Copyright 2008 Panoramic Corporation X ray Control Schematic 2 3 H...
Page 93: ...Appendix Copyright 2008 Panoramic Corporation X ray Control Schematic 3 3 H...
Page 94: ...Appendix Copyright 2008 Panoramic Corporation Laser Power Control Schematic H...