Steps of repairing
When the table is not working please make the following steps:
Cables: Please check that all cables are not squeezed and they must be connected correctly.
New battery: Check if a new battery can solve the problem.
Switches: Check if all cables are correctly connected to the control box. Also, please check and
be sure that a cable is not squeezed or destroyed. Before you disassemble the switches,
disconnect the switches from the control box. Connect a new switch and check if the table is
working or not. Please do not forget to install the battery when you check the table function. If
the table is still not working please go to the next step.
If one of the switches is the problem please disassemble the parts, which cover the switch and
unscrew the switch.
Cables connected
Cable for switches
Remove axle
Dismounted parts