27
TEMPERATURE RECORDER (OPTION)
WARNING
Always disconnect the power supply to the unit prior to attachment of a temperature recorder
in
order to prevent electric shock or injury.
A temperature recorders is available for the freezer as the optional component. The type of the
temperature recorder is MTR-G85A or MTR-G85C and MTR-4015LH. For the attachment, the recorder
fixing is necessary. Following shows the combination of temperature recorder with the recorder fixing
Contact our sales representative or agent for the attachment of a temperature recorder. :
Temperature recorder
Mounting kit
MTR-4015LH MPR-S30
MTR-G85A or MTR-G85C (circular chart type)
MPR-S7
Setting of MTR-4015LH
Pull the knob on the upper part of the temperature recorder
forward to change the recording paper or battery.
Setting of recording paper
1.
The information on the temperature recording paper is
shown in Fig. 1.
2.
Pull the cartridge up after opening the top lid. The lid
can be opened by turning the knob counterclockwise.
See Fig. 2.
3.
As shown in Fig. 3, insert the recording paper with the
“begin” tab placed in the cartridge. Check that the printed
side is facing out.
4.
Place the recording paper beneath the arm and between
the plate spring and guide plate in the direction of the
arrow.
Note:
Do not scratch or put pressure on the recording paper.
Do not bend the recording paper.
Do not reverse the recording paper manually.
The used recording paper left in the used recording paper
compartment can cause a malfunction. Be sure to
remove it. See Fig. 4.