14
TEMPERATURE SETTING
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is -25
o
C.
Note:
The unit is set at the factory that the chamber temperature is -30
o
C.
Table 1. Basic operation sequence (Example: Chamber temperature -25
o
C)
Description of operation
Key operated
Indication after operation
1 Turn the power switch ON.
----
The current chamber temperature is
displayed.
2 Press set key.
SET
The second digit is flashed.
3
Set to -25 with the numerical value
shift key and digit shift key.
When pressed, the figure of settable
digit changes.
When pressed, the settable digit
is shifted.
4 Press set key.
SET
Set temperature is memorized and the
current chamber temperature is
displayed.
Note:
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
• Although the value of the chamber temperature setting can range from -18
o
C to -35
o
C, the guaranteed
temperature with no load is -30
o
C when the ambient temperature is 35
o
C.
Key lock function
This unit is provided with the key lock function. When the key lock is ON, change of temperature setting
through the key pad is not possible. The key lock is set to OFF at the factory.
Display Mode
Function
L 0
Key lock is OFF
Enable to change temperature setting
L 1
Key lock is ON
Disable to change temperature setting
Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2 Press digit shift key for 5 seconds.
The first digit is flashed.
3
Press numerical value shift key
and scroll the figure to 1.
When pressed, the figure of settable
digit changes.
4 Press set key.
SET
The key lock is set to ON.
The current chamber temperature is
displayed.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.