51
Defrosting of chamber
The frost is built at the upper portion of the chamber and inner door. The excessive frost possibly make
some gap between the cabinet and the magnetic door gasket, which may cause poor cooling. Remove
the frost on the inner door with a scraper enclosed with the unit. Following shows the procedure for
removing the chamber frost.
Note:
For removing the frost, do not use a tool with sharp edge such as a knife or a screw driver.
1.
Turn off the backup cooling kit if it is installed.
2.
Take out and transfer all the contents to another freezer or a container which is refrigerated by liquid
carbon dioxide or dry ice.
3.
Turn off the power switch and battery switch of the
freezer.
4.
Open the outer door and inner door. Remove the
inner door by lifting up as shown in the figure.
5.
Leave the freezer as it is.
6.
The water accumulated on the bottom of the
chamber should be wiped up with a dry cloth.
7.
After cleaning the chamber and inner door, replace
the inner door and start up the unit according to the
procedure on page 17.
8.
Put back the articles into the sufficiently cooled
freezer compartment.
9.
Turn on the backup cooling kit if it is installed.
WARNING
Always mount and/or remove the inner door by two people.
The work by oneself may cause injury
from falling.
Always wear gloves when mounting and/or removing the inner door to prevent injury.