background image

51

 

 

 
5.

 On the Alarm Export screen, select the period to 

export.

 

 

To export the saved alarm log data over the entire 

period, press All radio button. 
 

To export the alarm log data for the specified days 

(The newest period containing that day), press Last 
XX Days radio button and input days. 
Settable range: 1 day

45 days. 

Note:

 The error of about 1 minute may be observed 

during 1 month. Refer to page 53 for the procedure 
of setting time. 
 

6.

 Press Export key. 

 
 
 

 
 
 
 
 
 
 
Note:  

When USB memory is not inserted in the USB port, 

Notice dialog box is displayed. Press OK key and 
insert an USB memory into the USB port. 
 
 
 
 
 
 
 
 

When alarm log data doesn’t exist in the specified 

days, Notice dialog box is displayed. Press OK key 
and specify days again as shown in the procedure 

5

 
 
 
 
 
 
 

Summary of Contents for MCO-230AIC

Page 1: ...carefully before using this product and save the operating instructions for future use See page 96 for all model numbers MCO 230AIC MCO 230AICUV MCO 230AICUVH Series CO2 Incubator Operating Instructions MCO 230AIC MCO 230AICUV MCO 230AICUVH ...

Page 2: ......

Page 3: ... inner attachments P 24 Filling humidifying pan P 25 FOR BETTER CULTIVATION Precautions for cultures P 26 Preventing contamination P 27 CORRECT OPERATION P 28 BASIC OPERATION ON LCD TOUCH PANEL P 29 BASIC PARAMETERS Numerical input to input window P 31 Setting temperature CO2 density and high limit temperature alarm P 32 Setting key lock P 34 Removing key lock P 36 ALARM PARAMETERS P 38 OPERATION ...

Page 4: ...ION P 68 Setting User ID P 68 Setting auto lock P 70 Using unlock key P 73 Removing auto lock P 74 GAS AUTO CHANGER OPTION P 75 Connecting CO2 gas cylinder P 75 Automatic CO2 gas supply line changeover P 76 Manual CO2 gas supply line changeover P 78 STD GAS AUTO CALIBRATION KIT OPTION P 79 ROUTINE MAINTENANCE P 82 ALARMS SAFETY AND SELF DIAGNOSIS P 83 TROUBLESHOOTING P 86 DISPOSAL OF UNIT P 89 SPE...

Page 5: ... part of the operating instructions may be reproduced in any form without the expressed written permission of our company CAUTION Our company guarantees the product under certain warranty conditions Our company in no way shall be responsible for any loss of content or damage of content INTENDED USE AND PRECAUTIONS This equipment is designed for culture of cell tissues organs embryos The adapted cu...

Page 6: ... Symbol shows This symbol means caution This symbol means an action is prohibited This symbol means an instruction must be followed Be sure to keep the operating instructions in a place accessible to users of this unit Label on the unit This mark is labeled on the cover in which the electrical components of high voltage are enclosed to prevent the electric shock The cover should be removed by a qu...

Page 7: ...et is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than that on the rating...

Page 8: ...ire by short circuit Disconnect the power supply plug before moving the unit Take care not to damage the power cord A damaged cord may cause electric shock or fire Disconnect the power plug when the unit is not used for long periods Keeping the connection may cause electric shock current leakage or fire due to the deterioration of insulation If the unit is to be stored unused in an unsupervised ar...

Page 9: ...d items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet copy the last page when you request any repair or maintenance for the safety of service personnel Wear rubber gloves when handling the H2O2 reagent Direct contact with the H2O2 reagent may result in inflammation of the skin H2O2 decontamin...

Page 10: ...sing the door switch Do not press the door switch because the UV light is hazardous Personal injury Hazardous UV Light Interior The UV light is hazardous Never turn on the UV lamp without the cover Personal injury Gas Poisoning or Oxygen Deprivation Environment When using CO2 gas for control make sure that there is adequate ventilation Using CO2 gas in a small room without adequate ventilation may...

Page 11: ...This symbol indicates an earth This symbol means ON for a power switch This symbol means OFF for a power switch ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions based on the IEC 61010 1 Indoor use Altitude up to 2000 m Temperature 5o C to 40o C Maximum relative humidity 80 for temperature up to 31o C decreasing linearly to 50 relative humidity ...

Page 12: ...tem set MCO 170UVS is installed 2 MCO 230AICUVH or when an optional electric lock MCO 170EL is installed When some inner attachments are removed 19 Leveling feet Leveling Foot Handles When 13 is removed Sample air outlet cap 18 17 16 1 15 14 Removal power supply cord port Lower right side 20 21 Access port cover Silicon cap Rear left ...

Page 13: ...an Fill the humidifying pan with sterile distilled water and set the humidifying pan with the inner side flush against the back Install the humidifying pan in a longitudinal direction as its shorter side is placed in the back 11 Door switch Detects the door opening closing and stops the fan and electromagnetic valve for CO2 when the door is open The UV lamp 1 is also deactivated by the door openin...

Page 14: ...ayed when CO2 density is set 0 5 Set CO2 density value display field The set value of the chamber CO2 density is displayed Default setting 0 6 CO2 gas injection indicator This lamp lights when CO2 gas is being injected 7 CO2 gas supply line indicator A and B 1 Current supplying CO2 gas supply line connecting port for CO2 gas pipe is displayed The connecting port A B for gas pipe that is currently ...

Page 15: ...tely in normal characters and reverse video Close Door Closed is displayed Locked Door Locked is displayed 2 13 Message display field Alarms errors or messages are displayed when fault occurs Refer to page 83 85 Note When there are a number of alarms errors the display shows the message For example if 2 alarms errors occur in total the display shows 1 2 14 Message select key When there are a numbe...

Page 16: ... at a remote location from this product by connecting the external alarm unit to the remote alarm terminals For the type and behavior of remote alarm output refer to page 83 to 85 The terminal of the remote alarm is installed at the rear upper right of the unit See the figure on the point The alarm is outputted from this terminal Contact capacity is DC 30 V 2 A When Buzzer key is pressed the behav...

Page 17: ...oilers ovens or autoclaves Heat will adversely affect the performance of the incubator Ambient temperature at least 5 C lower than set temperature The control temperature of the incubator is at least 5 C higher than the ambient temperature For example if the chamber is controlled at 37 C the ambient temperature must be 32 C or less Do not allow the ambient temperature to become too high Strong and...

Page 18: ...o water In addition do not install it near water or steam pipes Moisture can cause the insulation to deteriorate which may result in current leakage or electric shock No inflammable or corrosive gas Never install the incubator in a location where it will be exposed to inflammable or corrosive gas Doing so may result in explosion or fire In addition insulation may deteriorate due to corrosion of pr...

Page 19: ...he instruction of the detergent Wipe off the residual detergent with a wet cloth and then wipe off any moisture Note Remove the cable tie banding the power supply cord Prolonged banding may cause the corrosion of the cord coating WARNING Do not leave the plastic wrapping bags within reach of children If the bag is placed over a child s head it can block the mouth and nose and cause suffocation 2 S...

Page 20: ...ock always ground the incubator If grounding is not possible then have a ground installed by qualified personnel If the incubator is not grounded it may result in electric shock Never connect the ground wire to a gas pipe water pipe lightning rod or telephone ground wire Doing so may cause electric shock Installing a ground fault circuit breaker If using the incubator in the location with moisture...

Page 21: ...ator or O2 CO2 incubator other than this product use the stacking plate MCO 230SB Refer to table 11 on page 96 When the incubator is not in use Empty the water from the humidifying pan and remove moisture from the chamber Make sure that the chamber is completely dry before closing the doors Failure to do so may result in damage Before moving the incubator Before moving the incubator empty the wate...

Page 22: ...sing 2 tube bands that is provided connect the gas tube tightly to prevent it from coming off Make sure that the gas tube is not folded If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder a CO2 solid will be formed in the gas regulator The gas regulator safety valve will operate and there may an explosive sound 3 After connecting the gas tube make sure that no gas is l...

Page 23: ...on corrosion or rusting Be careful to keep the diluted detergent or water out of the temperature sensor the CO2 gas injection port the inner sample air access port the fan motor shaft bearing and the inner sample air outlet Fig 2 Also do not wash the temperature sensor and the UV lamp using detergent Doing so may cause failure Fig 2 3 Soak a gauze or unwoven cloth in distilled water and wring it t...

Page 24: ... cleaning to keep the intended performance Be careful not to damage the UV lamp in the duct MCO 230AICUVH 230AICUV or when an optional UV system set MCO 170UVS is installed 1 Turn OFF the power to the incubator 2 Open the outer and inner doors and pull out all the trays Fig 1 3 Pull out the humidifying pan Fig 2 4 Pull out the humidifying pan cover Fig 3 5 Pull out the fan cover Fig 4 Humidifying ...

Page 25: ...uct and remove it from the pins on the rear side Fig 5 7 Pull out the fan Fig 6 8 Remove the silicon caps of each access port from interior Fig 7 and exterior Fig 8 Duct Silicon cap Fig 5 Fig 6 Fig 7 Fig 8 Silicon cap ...

Page 26: ...ling the fan cover position the long hole of duct with the projection of fan cover and insert directly Fig 3 Same applies for the humidifying pan cover The fan cover may lean by strongly pushing the head of it in the back Make sure that there is no space in the lower fan cover after installing because the leaning fan cover may have a bad influence on the camber temperature distribution CAUTION If ...

Page 27: ... as its shorter side is placed in the back Refill the humidifying pan with water early when the volume of water is decreased Mixing any reagent in the humidifying water may have a bad influence on the cultivation Especially when using the UV lamp do not use any reagent Because the UV light may deteriorate the reagent mixed with the humidifying water 5 Set the humidifying pan with the inner side fl...

Page 28: ... the UV lamp is OFF if the MCO 230AICUVH 230AICUV or the optional MCO 170UVS is installed Be careful when closing the outer door Use the handle when closing the outer door Holding the door in other places may cause injury by getting fingers caught in the door Do not lean on the outer door Doing so may result in injury from the outer door coming loose or the incubator falling over or it may cause c...

Page 29: ... them in and out Keep the chamber clean Wipe off any fingerprints If water spills from the humidifying pan or if the doors are left open for a long time condensation may form on the inside of the doors If that occurs wipe off the condensation with a dry sterile gauze In particular clean and disinfect the chamber if the culture medium is spilled For details refer to ROUTINE MAINTENANCE on page 82 U...

Page 30: ...wer switch on the lower right side of the incubator 7 MCO 230AICUVH 230AICUV or when an optional UV system set MCO 170UVS is installed Set the frequency of a power supply on the LCD touch panel Refer to page 58 59 WARNING Always use the removal power supply cord that is provided Other power supply cord may cause electric shock or fire The provided removal power supply cord is only for this product...

Page 31: ... Alarm Export Export Alarm log 50 52 Lock Lock screen Auto Lock Auto Lock screen 1 Setting Auto lock ON OFF User ID ON OFF 70 74 KeyLock Key Lock screen Setting Key lock ON OFF password 34 37 Auto Lock User Auto Lock User screen Setting User ID password 68 70 Tools 1 Tools 1 screen Temp CO2 Calibration Don t push See Next page Alarm Setting 1 Setting Temp alarm CO2 density alarm 38 39 Alarm Settin...

Page 32: ...ccurate temperature or CO2 density may be displayed Operation from other than Menu key CO2 gas supply line select key Operate Manual changeover between CO2 gas supply line A and B Page 78 Message select key Operate Changing some messages H2O2 key 5 Operate H2O2 decontamination Page 62 67 Unlock key 6 Operate Unlock the outer door Page 68 72 Buzzer key Operate Silencing the buzzer Alarm is not canc...

Page 33: ...ue and press OK key Key description Numerical key 0 9 Input numerical values Up Down key Increases or decreases the numerical value displayed in the numerical input box Clear key Deletes the numerical value displayed on the numerical input box Cancel key Stops inputting on the numerical input box and closes the input window Note While the input window is open it is not possible to operate Top key ...

Page 34: ...ature the CO2 density and the temperature of the high limit temperature alarm for normal operation according to the following procedure The incubator automatically starts operation using these settings after power on 1 Press Menu key to lead the Menu screen 2 Press Set key to lead the Stand by Setting screen ...

Page 35: ... less than 45 0 o C to 45 0 o C or higher the incubator readjusts the CO2 sensor During readjusting Status Gas sensor initializing is displayed in the Message display field and is displayed in the Present CO2 density display field After 15 minutes in the shortest the incubator returns to the normal operation When operating the incubator for the first time or after not using it for an extended peri...

Page 36: ...34 BASIC PARAMETERS Setting key lock 1 Press Menu key to lead the Menu screen 2 Press Lock key to lead the Lock screen 3 Press KeyLock key to lead the Key Lock screen ...

Page 37: ...y Lock turns to ON Password The number Max 6 digit inputted here are registered the release password of Key Lock Confirm Password To prevent erroneous input input the same password as Password input box When inputting different password Notice dialog box is displayed Press OK key and input the correct password Note To prevent abuse of the release password of Keylock manage properly by limited admi...

Page 38: ...d and input of the release password of Key Lock is required When the inputted password is incorrect Notice dialog box is displayed Press OK key and then input the correct password Removing key lock 1 By pressing Menu key the Password input window is displayed 2 On Password input box input the set release password of Key Lock and press OK key to lead the Menu screen ...

Page 39: ...reen 5 On the Key Lock screen by holding Key Lock slide key and sliding to the left change to OFF Press Apply key to turn the key lock OFF The display returns to the Lock screen Note The release password of key Lock is deleted 6 On the Lock screen press Top key to return to the Top screen ...

Page 40: ...m delay ring back and remote alarm go to procedure 5 4 On the Alarm Setting 1 screen input each parameter Press Apply key to save the input value The display returns to the Tools 1 screen Each parameter setting Temp Alarm When the chamber temperature exceeds the scope the set temperature the set value of Automatic set temperature alarm the alarm is activated Settable range 1 0 o C 5 0 o C factory ...

Page 41: ...overed from the alarm state within the door alarm delay time the buzzer doesn t sound after the elapse of the door alarm delay Ring Back The function is that the alarm buzzer sounds again when the alarm state still continues after the alarm delay set time passed even though the alarm buzzer was stopped by pressing Buzzer key By holding and sliding Ring Back slide key to the right the Ring Back is ...

Page 42: ...ey For the behavior at the time of pressing Buzzer key and the re activation of alarm under each setting condition refer to Table 5 7 on page 84 Resolve the cause of the alarm in reference to page 83 85 because the alarm itself is not deactivated by pressing Buzzer key except for some alarms Note The buzzer for the high limit temperature alarm can t be silenced ...

Page 43: ...ration log data chamber temperature CO2 density and open close state of outer door Use the following procedure to set the log interval interval of acquiring the operation log 1 Press Menu key to lead the Menu screen 2 Press Log key to lead the Log screen 3 Press Setting key to lead the Setting screen ...

Page 44: ...between log interval and the estimated amount of data that can be saved Log interval 2 minutes Approx 46 days Log interval 6 minutes Approx 135 days Log interval 30 minutes Approx 664 days When saving data more than the above and the data is overwritten and the old data is delated 5 Press Top key to return to the Top screen Displaying operation log Operation log saved in the incubator can be displ...

Page 45: ...aphically 5 On the Chart screen by pressing Show key after pressing the item you want to display graphically the graph of each operation log is displayed Actual Temp Chamber temperature log graph Go to procedure 6 Actual CO2 Level CO2 density log graph Go to procedure 7 Door Opening Open close state of outer door log graph Go to procedure 8 ...

Page 46: ...p key to return to the Top screen 8 Door Opening log graph is displayed Press Back key to return to the Chart screen Press Top key to return to the Top screen Note The error of about 1 minute may be observed during a month Refer to page 53 for the procedure of setting time To Actual CO2 Level log To previous day To next day To Actual Temp log To Door Opening log To previous day To next day To Actu...

Page 47: ...SB memory inserted into the USB port 1 Insert the USB memory into the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu key to lead the Menu screen 3 Press Log key to lead the Log screen 4 Press Data Export key to lead the Export screen ...

Page 48: ...select the type of operation log data you want to export To export all types of operation log data press All Ch key To export only operation log data you want to export select operation log data you want to export and then press Selected Ch key Actual Temp Chamber temperature log data Actual CO2 Level CO2 density log data Door Opening Open close state of outer door log data When both of the Auto l...

Page 49: ...d in the USB memory and the exported file is saved in it in CSV format The exported file name is in date 8 digits type of data format e g When exporting all types of data using All from Oct 1st 2013 to Jan 1st 2014 20131001 20140101_AllCh csv 20131001 20140101_Door csv e g When exporting Actual Temp using 1 Day Jan 1st 2014 20140101_Temp csv On the beginning of the exported file MCO 230AIC is writ...

Page 50: ...ax 256 logs Alarm log saved in the incubator can be displayed graphically on the LCD touch panel Note When saving alarm logs more than 257 the oldest alarm log is deleted and then overwritten 1 Press Menu key to lead the Menu screen 2 Press Log key to lead the Log screen 3 Press Alarm key to lead the Alarm screen ...

Page 51: ...5 days Note The error of about 1 minute may be observed during 1 month Refer to page 53 for the procedure of setting time Press Back key to return to the Log screen Press Top key to return to the Top screen On the Alarm screen of procedure 4 or 5 alarm log data can be exported in CSV format to the USB memory inserted into the USB port 6 Insert the USB memory into the USB port Note It is not possib...

Page 52: ...nserted in the USB port by CSV format 1 Insert a USB memory in the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu key to lead the Menu screen 3 Press Log key to lead the Log screen 4 Press Alarm Export key to lead Alarm Export screen ...

Page 53: ...d input days Settable range 1 day 45 days Note The error of about 1 minute may be observed during 1 month Refer to page 53 for the procedure of setting time 6 Press Export key Note When USB memory is not inserted in the USB port Notice dialog box is displayed Press OK key and insert an USB memory into the USB port When alarm log data doesn t exist in the specified days Notice dialog box is display...

Page 54: ...ator is not deleted 8 Remove a USB memory from the USB port Note A log folder is created in a USB memory and an exported data file is saved in the log folder by CSV format Exported file name The first date during exported period 8 digits the last date 8 digits AlarmLog Example When exporting alarm log data for 7 days on January 7 2014 20140101 20140107_AlarmLog csv 9 Press Top key to return to the...

Page 55: ... key to lead the Date Time screen 4 On the Date Time screen input the present date and time Press Apply key to save the input value The display returns to the Tools 2 screen Note 24 hour clock It is recommended to set the time periodically since the error of about 1 minute may be observed during a month 5 Press Top key to return to the Top screen ...

Page 56: ...en 2 Press Tools 2 key to lead the Tools 2 screen 3 Press Brightness Sleep key to lead the Brightness Sleep screen 4 On the Brightness Sleep screen each setting of brightness and sleep is available Press Apply key to save the input value and setup The display returns to the Tools 2 screen ...

Page 57: ...he Sleep function is turned to ON Input the set value of time to change the Sleep state Settable range 1 minute 5 minutes factory setting 2 minutes Note It is not possible to operate any key in the Sleep state By touching the LCD touch panel the Sleep state is released and the LCD touch panel returns to the usual state Under this condition key operations are available Brightness Sleep Brightness o...

Page 58: ... 2 Press Tools 2 key to lead the Tools 2 screen 3 Press DAQ Setting key to lead the DAQ Setting screen 4 On the DAQ Setting screen each setting of DAQ is available Refer to the Procedure Setting Manual Data Acquisition System MTR 5000 for details of settings Press Apply key to save the input value and setup The display returns to the Tools 2 screen 5 Press Top key to return to the Top screen ...

Page 59: ...n the condensation in the chamber and chamber temperature distribution because the UV lamp generates heat for a long time It may also shorten the service life of the UV lamp The preset period can be changed when necessary as shown in the page 58 59 To check whether the UV lamp is lit open the outer door and then press the door switch with the inner door close Visible blue light can be confirmed fr...

Page 60: ...r information on replacing the UV lamp If the UV lamp burns out Err18 UV Lamp Abnormal is displayed in the message display field it is not possible to perform H2O2 decontamination Replace the UV lamp and the glow starter Setting UV lamp ON period Use the following procedure to change the setting of the UV lamp ON period 1 Press Menu key to lead the Menu screen 2 Press Tools 1 key to lead the Tools...

Page 61: ... lamp has turned on is displayed as the percentage to 5 000 hours which are recommendation time to replace It is impossible to set UV Timer Ext The more total time which UV lamp has turned on increases the more UV ray output declines In order to cover a decline of the UV ray output the lighting time of UV lamp is automatically extended with an increase of total lighting time of UV lamp The set val...

Page 62: ...m the chamber including the trays the fan cover the duct the fan the humidifying pan and the humidifying pan cover Clean all the attachments in an autoclave or with alcohol for disinfection 2 Clean and wipe off inside the chamber with alcohol for disinfection 3 Set the CO2 density to 0 Refer to page 32 33 4 Press Menu key to lead the Menu screen 5 Press Tools 1 key to lead the Tools 1 screen 6 Pre...

Page 63: ...on display when UV lamp is lighting Note The UV 24 hour mode may cause the automatic set temperature alarm because of a rising chamber temperature After procedure 8 by opening the outer door when UV lamp is lighting UV lamp is turned OFF and UV 24 hour mode is canceled by opening the outer door Redo from procedure 4 to start the UV 24 hour mode again 9 Press Top key to return to the Top screen 10 ...

Page 64: ...NG When performing H2O2 decontamination make sure that the outer and inner doors are securely closed During H2O2 decontamination plug the access hole with the silicon caps that are provided Failure to do so may be harmful to health due to leakage of H2O2 gas WARNING Never start H2O2 decontamination when Warning UV Bulb Life is displayed in the message display field The UV resolve is not sufficient...

Page 65: ...on the lower left side of the duct Fig 1 Note Make sure that the H2O2 generator is securely attached Uncertain attachment may result in insufficient decontamination 7 Remove a connector cap from connector on bottom right of the far side of the chamber Connect the H2O2 generator and connector by a cable that is provided Fig 2 Note Be sure to keep the connector cap 8 Insert 4 trays in the 2nd the 3r...

Page 66: ...hat close the inner door and the outer door 11 Press H2O2 key for 3 seconds to open Setting Position window Note When the model does not have H2O2 decontamination function H2O2 key is not displayed on the Top screen When key lock is ON Password input window is opened and input of the release password of Key Lock is required Refer to page 36 12 Finally make sure that the inner attachments and the H...

Page 67: ...zzer sounds when H2O2 decontamination is completed About 100 minutes later CAUTION The electric lock will remain locked if power supply is cut off during H2O2 decontamination After the power supply is recovers the H2O2 gas resolution process will start execution and finish automatically Execute the decontamination again because the decontamination is not completed WARNING Do not use the unlock key...

Page 68: ...r and inner doors Disconnect connector of the chamber cable and H2O2 generator and remove H2O2 generator and cable from the chamber Note When doing the above work put on protection glasses and rubber gloves 16 On the H2O2 Decontamination Step8 screen press OK key to return to the Top screen Note When H2O2 generator is connected with connector of the chamber by cable OK key is not workable 17 Dilut...

Page 69: ...prevent impurities from becoming mixed in the reagent Check the container to make sure that there is no damage corrosion or cracking Store the container with the inlet facing upwards and make sure that the container will not tip or be knocked over Precautions for disposal Dispose according to the rule in your country Using unlock key Unlocking when power is interrupted MCO 230AICUVH or when an opt...

Page 70: ...e MCO 230AICUV 230AIC The modes of unlocking the outer door are as follows Quick mode Press the Unlock key User ID mode Input the User ID and release password of Auto Lock after pressing the Unlock key Setting User ID Before turning the User ID mode to ON use the following procedure to register a User ID and a release password of Auto Lock 1 Press Menu key to lead the Menu screen 2 Press Lock key ...

Page 71: ...word as Password input box When inputting different password Notice dialog box is displayed Press OK key and input the correct password Note A release password of Auto Lock is for unlocking the outer door It is different from the release password of Key Lock refer to page 34 36 It is possible to input up to 8 digit alphanumeric characters as a User ID It is possible to input up to 6 digit numbers ...

Page 72: ... registered User ID into User ID input box and input its registered password into Password input box Press Delete key to delete the registered User ID and its password Note When deleting all registered User IDs the User ID mode is turned to OFF refer to page 70 5 On the Menu screen press Back key to return to the Top screen Setting auto lock 1 Press Menu key to lead the Menu screen 2 Press Lock ke...

Page 73: ... of unlocking the outer door between the quick mode or the User ID mode By holding the User ID slide key and sliding it right the User ID mode is turned to ON Factory setting OFF quick mode Note When no User ID is registered notice dialog box is displayed Press OK key and then register a User ID and its password in reference to page 68 70 In the User ID mode User ID which is inputted to unlock the...

Page 74: ...r ID and its release password of Auto Lock Note The User ID which is inputted at this time is saved as the open close state of outer door log data refer to page 45 46 When the inputted User ID or its password is incorrect Notice dialog box is displayed Press OK key and then input the correct User ID or its password Note When the unlocked outer door is closed and the setting time passes the unlocke...

Page 75: ... the power is interrupted use the unlock key that is provided To re lock the outer door turn the unlock key to the lock direction while the outer door is open Close the outer door after the out door is locked manually Note The outer door cannot be locked by using the unlock key while the outer door is closed Lock the outer door while it is open Attempting to turn the unlock key while the outer doo...

Page 76: ...ad the Lock screen 3 Press Auto Lock key to lead the Auto Lock screen 4 On the Auto Lock screen by holding the Auto lock slide key and sliding it left the Auto lock is turned to OFF Press Apply key to change Auto lock OFF and the display returns to the Lock screen 5 Press Top key to return to the Top screen ...

Page 77: ...s regulator of the CO2 gas cylinder A 3 Using a gas tube that is provided connect the connecting port B for CO2 gas pipe and the gas regulator of the CO2 gas cylinder B Note If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder a CO2 solid will be formed in the gas regulator The gas regulator safety valve will operate and it may make an explosive sound 4 After connecting...

Page 78: ...to changer MCO 21GC is installed CO2 gas supply line indicator A B and CO2 gas supply line select key are displayed in the Top screen CO2 gas supply line indicator A or B being used is lighted CO2 gas supply line indicator A and B CO2 gas supply line select key CO2 gas supply line being used ...

Page 79: ...gas cylinder A Table 2 CO2 gas supply line automatic changeover e g When CO2 gas cylinder A is empty it changes over CO2 gas cylinder B Situation CO2 gas CO2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 CO2 gas is supplying from valve A A Remaining Remaining Light on Light off 2 CO2 density in the chamber is not increased even if CO2 gas valve opens A Emp...

Page 80: ...anger function the empty CO2 gas is manually retuned to the CO2 gas supply line before the Buzzer key is pressed Table 3 After changed into CO2 gas cylinder B because CO2 gas cylinder A is empty if you changed into A manually Situation CO2 gas CO2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 Changed into CO2 gas supply line B automatically B Empty Remaini...

Page 81: ...O2 incubator Since a Standard gas cylinder is used as a standard of exact density during CO2 density calibration prepare a standard gas cylinder that is same as the set CO2 density Note There is not the problem to remain connected standard gas cylinder after finished CO2 gas density calibration 2 Press Menu key to lead the Menu screen 3 Press Tools 1 key to lead the Tools 1 screen 4 Press STD Gas ...

Page 82: ...ns to the Tools 1 screen Settable range 4 0 21 0 Factory setting 5 0 6 Press STD Gas Calibration key to lead the STD Gas Calibration screen 7 On the STD Gas Calibration Step1 screen system check starts automatically If the system is normal display leads the STD Gas Calibration Step2 screen 8 On the STD Gas Calibration Step2 screen press OK key to lead the STD Gas Calibration Step3 screen ...

Page 83: ...10 automatically 10 After completion of CO2 density calibration display leads the STD Gas Calibration Step5 screen CO2 incubator returns to the normal operation 11 On the STD Gas Calibration Step5 screen press OK key to return to the Tools 1 screen On the Tools 1 screen press Top key to return to the Top screen ...

Page 84: ...east once a month 1 Remove all the inner attachments by the procedures shown on page 22 2 Clean the chamber and all the inner attachments by the procedures shown on page 21 3 Install all the inner attachments by the procedures shown on page 24 When there is excessive dirt contact our sales representative or agent ...

Page 85: ...2 gas cylinder empty The CO2 density does not increase when the CO2 valve is opened Err01 CO2 Gas Empty is displayed in the message display field Chamber temperature sensor error The chamber temperature sensor is disconnected Err05 Temp Sensor Open is displayed in the message display field Heater OFF The chamber temperature sensor is short circuited Err06 Temp Sensor Short is displayed in the mess...

Page 86: ... Interlock with Buzzer key ON ON OFF Alarm is not canceled ON OFF OFF OFF Note Resolve the cause of the alarm in reference to page 83 85 because the alarm itself is not deactivated by pressing Buzzer key Table 6 In the cases of high limit temperature alarm or Err38 The outer door opens during H2O2 decontamination refer to next page Remote Alarm setting Ring Back setting Buzzer from CO2 incubator R...

Page 87: ...r is open ON when the outer door is open H2O2 mist generation is cancelled Resolve time is extended Failure during H2O2 gas resolve by UV The UV lamp failed during H2O2 gas resolve by UV Err36 UV Lamp Failure is displayed in the message display field Intermittent tone ON Resolve time is extended Power was interrupted during H2O2 gas resolve by UV After power is restored Err37 Power Failure is disp...

Page 88: ... kgf cm 2 G 551 psi G or lower it is a sign that there is little gas remaining Replace the cylinder soon Is the incubator operating beside the appliance that generates the electromagnetic wave The chamber temperature does not match the set value Is the ambient temperature less than 5 C different from the set value for the chamber temperature Is the outer door closed with the inner door left open I...

Page 89: ...ng If gas density is slow to recover when the CO2 gas pressure is normal it is possible that the HEPA filter may be clogged Contact our sales representative or agent Is there little gas remaining in the CO2 gas cylinder Is the secondary pressure for the gas regulator at the specified value of 0 03 MPa G 0 3 kgf cm 2 G 4 4 psi G Is the gas tube blocked Is the duct securely attached Attach the duct ...

Page 90: ...g 1 Is the fan attached properly Confirm if the fan is pushed all the way to the motor shaft Fig 2 Is the H2O2 generator securely installed Set the 2 pins of the H2O2 generator in the 2 holes at the left bottom of the duct Fig 3 Is it the end of life of the H2O2 generator If the total operating time exceeds 5 000 hours replace the H2O2 generator Fig 1 Fig 2 Fig 3 Note If the problem still has not ...

Page 91: ...or must be dismantled and disposed of by qualified personnel only If the CO2 incubator is left where outsiders enter it may result unexpected accident for example children to become locked inside Before disposing the CO2 incubator with biohazardous danger decontaminate the CO2 incubator to the extent possible by the user ...

Page 92: ...ase help us to conserve the environment we live in German Ihr Panasonic Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und Komponenten die recycelt und wiederverwendet werden können Dieses Symbol bedeutet daß elektrische und elektronische Geräte am Ende ihrer Nutzungsdauer von Hausmüll getrennt entsorgt werden sollen Bitte entsorgen Sie dieses Gerät bei Ihrer örtli...

Page 93: ...e manière inappropriée peuvent s avérer potentiellement dangereuses pour la santé humaine et pour l environnement Cependant ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine Pour cette raison il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé avec vos ordures ménagères Spanish Los productos Panasonic están diseñados y fabricado...

Page 94: ...sistemas de recolha separados para produtos eléctricos e electrónicos usados Por favor ajude nos a conservar o ambiente em que vivemos Italian Il vostro prodotto Panasonic è stato costruito da materiali e componenti di alta qualità che sono riutilizzabili o riciclabili Prodotti elettrici ed elettronici portando questo simbolo alla fine dell uso devono essere smaltiti separatamente dai rifiuti casa...

Page 95: ...tronische apparatuur bij de daarvoor wettelijke instanties aangeboden Alstublieft help allen mee om het milieu te beschermen Swedish Din Panasonic produkt är designad och tillverkad av material och komponenter med hög kvalitet som kan återvinnas och återanvändas Denna symbol betyder att elektriska och elektroniska produkter efter slutanvändande skall sorteras och lämnas separat från Ditt hushållsa...

Page 96: ...et system Heater 345 W Humidifying system Natural evaporation with humidifying pan Temperature controller PID control system Temperature display Digital display CO2 controller PID control system CO2 density display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or higher UV lamp 4 W x 1 ozone free emission Alarms Automatic set temperature alarm Automatic set CO2 den...

Page 97: ...erface board MCO 420MA Interface board MTR L03 For LAN Interface board MTR 480 For RS 232C RS 485 Optional product for using in the chamber Shaker for CO2 incubator MIR S100C Only for the Data acquisition system MTR 5000 user Note Refer to the updated catalog when ordering an optional component Designs and specifications are subject to change without notice Table 10 Optional accessories for each f...

Page 98: ...e 5 o C to 50 o C ambient temperature 5 o C to 35 o C Temperature distribution 0 25 o C ambient temperature 23 o C setting 37 o C CO2 5 no load Temperature variation 0 1 o C ambient temperature 23 o C setting 37 o C CO2 5 no load CO2 control range 0 to 20 CO2 variation 0 15 ambient temperature 23 o C setting 37 o C CO2 5 no load Chamber humidity 95 R H 5 R H Applicable environment condition Temper...

Page 99: ...f the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see below Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name CO2 incubator Model No MCO Serial number Date of Installation ...

Page 100: ... only 0123 Panasonic Marketing Europe GmbH Panasonic Testing Centre Winsbergring 15 22525 Hamburg Germany Panasonic Healthcare Co Ltd 2015 Printed in Japan LDCL001000 2 S0115 20916 2015 09 06 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 JAPAN ...

Reviews: