42
Technical data
EW 3004
Warning function
Flashing figures on the display signal
Method of measurement:
Intelligent measuring technology
(oscillometric)
(FUZZY LOGIC)
Measuring range:
Pressure: 0–280 mmH
g
The device is vented automatically at
pressures exceeding 280 mmH
g
and
an error message is shown.
Pulse: 30–160 beats per minute
Measuring accuracy:
Pressure: ±3 mmH
g
Pulse rate: ±5 %
Leakage of pneumatic system:
≤
6 mmH
g
/min
Inflation/deflation system:
Automatic measurement during
inflation, and rapid deflation
Wrist circumference:
12.5–22.0 cm (4 7/8–8 5/8 inches)
Batteries:
2
#
1.5V alkaline (LR03/AAA/Micro)
(Control sign for battery change)
Battery symbol appears when the
batteries have to be replaced.
Automatic switch-off:
Approx. 5 minutes after measurement
Reference method for clinical trials:
Auscultatory measurement
Weight:
95
g
(without batteries)
Storage and transport conditions:
– 20˚C to + 60˚C, 10% to 95% RH
Operating conditions:
+ 10˚C to + 40˚C, 30% to 85% RH
EW3004 was proven clinically by sequential comparison with auculatatory
measurement. The waiting period before EW3004 evaluation requires 1.5-2
minutes, and before auscultatory measurement of 0.5-1 minute.