Before attempting to connect or operate this product,

please read these instructions carefully and save this manual for future use.

Model No.

BM-ED500

Control Unit

Operating Instructions

Summary of Contents for BMED500 - CAMERA CONTROL UNIT

Page 1: ...Before attempting to connect or operate this product please read these instructions carefully and save this manual for future use Model No BM ED500 Control Unit Operating Instructions ...

Page 2: ...l devices Any changes or modifications not expressly approved by the party responsible for compli ance could void the user s authority to operate this equip ment For U S A WARNING To prevent fire or electric shock hazard do not expose this appliance to rain or moisture The apparatus shall not be exposed to dripping or splashing and that no objects filled with liquids such as vases shall be placed ...

Page 3: ...es and a third grounding prong The wide blade or the third prong are provided for your safety If the provided plug does not fit into your outlet consult an electrician for replacement of the obsolete outlet 10 Protect the power cord from being walked on or pinched particularly at plugs convenience receptacles and the points where they exit from the apparatus 11 Only use attachments accessories spe...

Page 4: ...Description 18 To Do Setting without the Buzzer Sound 18 Operation Mode Setting 18 OPERATING PROCEDURES 21 Turning the Power On 21 Turning the Power Off 21 When the Door Trouble is Automatically Unrecoverable 22 When the Door Trouble is Unrecoverable 22 ADDITIONAL SYSTEM COMPONENTS 23 Iris Camera BM ET500 23 Administration Software BM ES500E 23 SPECIFICATIONS OF ELECTRICAL LOCK CONNECTORS GATE 1 2...

Page 5: ...di tions 3 The iris camera for enrollment captures the iris images of both eyes at the same time but the iris data of each eye is enrolled separately 4 Capturing diagnosis is made by the administration software at the time of enrollment 5 Up to 127 gate systems an iris camera and control unit can be connected to the adminis tration PC in the LAN The con trol unit can also be used as both for enrol...

Page 6: ...ted you will be recognized as an enrolled user The recognition time will be approx 3 seconds approx 2 seconds for capturing and 1 second for recognition 1 1 The recognition time may differ depending on capturing conditions Access control system can be composed Iris cameras control units and the administration PC can compose an access control system in combination with access control panels 2 and e...

Page 7: ...ppearing in these operating instructions are registered trademarks or trade marks of the company concerned LIMITATION OF LIABILITY This Product is used to recognize an individual person by using Iris data and is not designed to protect against theft or crime independently IN NO EVENT SHALL MATSUSHITA ELECTRIC INDUSTRIAL CO LTD BE LIABLE TO ANY PARTY OR ANY PERSON EXCEPT FOR REPLACEMENT OR REASONAB...

Page 8: ...iris cameras Distribution To transfer enrolled data from the administration PC to control units and to save the data in the control units Enrollment To save a user s iris data associated with the individual information name and ID data etc in the administration PC Iris data is enrolled using the administration software Iris camera Panasonic Iris Camera BM ET500 Iris data Data that are generated fr...

Page 9: ...ce at temperatures between 0 C and 40 C between 32 F and 104 F and humidity between 30 and 80 The input power source for this appliance is 120 V AC 60 Hz Do not share an AC outlet with devices that consume much electricity such as a copy machine or air conditioner Do not operate the appliance while the door is open Power switch Power supply is not shut down only by turning off the power switch To ...

Page 10: ... Output board e Power board This board is shielded for safety Before the connec tions do the following 1 Loosen the 5 screws fixing the shield 2 Remove the shield After the connections attach the shield to the control unit as it was r Earth terminal t Clamping bar This bar is used for clamping the cables After the connections fix them with the supplied clamp ing tools y AC power cable Do not conne...

Page 11: ...o reset the disabled status use the administration software w Setup indicator Shows the setting data or which numeric button has been pressed e Operation mode switch DSW3 These switches are used for setting the operation mode of the control unit r Buzzer Beeps when the door is open or trouble has occurred to the control unit t Buzzer level adjustment knob You can adjust the buzzer volume with this...

Page 12: ...00Base TX This is a port to connect the control unit to LAN Local Area Network Note Route all power limited wiring away from non power limited wiring Power board q Camera power connectors Power limited terminal These are connectors to supply iris cameras with 32 V DC power One connector can supply power to one camera Note Power supply from one connector to more than one camera may cause trouble w ...

Page 13: ...ion It may cause invalid recognition or injury noise for example places near air conditioners or ventilators electricity This unit is for indoor use only Coaxial cable BNC Coaxial cable BNC Camera power cable General purpose power supply 24 V DC BM ET500 For recognition Camera 1 Camera 2 System Instruction Maximum System OK NG BM ET500 For recognition In the case of Wiegand RS 485 use OK NG OK NG ...

Page 14: ...Necessary for mounting the control unit over the wall with the unicorn anchor Unicorn anchor UC 830 or equivalents Necessary for mounting the control unit over the wall Preparation Prepare the following items before the installation The necessary items and their lengths differ depending on condition Preparing the necessary items 2 Cables with connector covers are unusable Note When you use the GAT...

Page 15: ...he supplied mounting bracket to the anchor Use an anchor bolt which meets the standards shown in the following figure 10 mm 2 5 in or less Wall Mounting bracket 2 mm 2 25 in D Hit the anchor into the wall by referring the following figure 30 250 362 4 2 4M 8 9 342 280 324 2 13 13 4 φ20 4 R 4 5 13 Unit of length mm Wall Type Screw Type Concrete wall TAPMARK concrete screw anchors Part No 21230 Size...

Page 16: ...ough the upper cable holes and then mount the control unit over the wall Make two holes to attach the supplied bushing or a flexible metal conduit described in the figure Use a metal conduit Use a metal conduit Cable holes Flexible metal conduit Trade size mm In Internal diameter mm In Total area 100 mm2 In 2 2 wires 31 mm2 In 2 Over 2 wires 40 mm2 In 2 1 wire 53 mm2 In 2 19 3 4 25 4 1 32 1 1 4 21...

Page 17: ... Note Alarm related operations and alarm output timing are configurable through the administration PC Refer to the installation guide of the administration software 5 After removing the power board s cover connect the camera power cable to the control unit as described in the figure ALARM OUT G 8 G 7 G 6 G 5 G 4 G 3 G 2 G 1 Note Connect the power cable and camera power cable after connecting elect...

Page 18: ...tion is stopped 2 Pull the door sensor 3 Restart the recognition from the administration PC You can do setting without the buzzer sound and you can continue the recognition with iris camera during the setting Setting changes will not be reflected until turn ing on the power again 4 After the setting close the door The buzzer will beep again 5 Restart the recognition from the administration PC When...

Page 19: ...ed of numbers up to 255 Note Enter 0 before the numerics from 0 to 99 to make it 3 digit Example When you enter an IP address 192 168 1 20 enter 192168001020 5 Enter the write command 113 000 Address will be checked After the checking the result will appear on the Setup indicator Completed 0 Failed 1 The address may be wrong Parameter error E The digits may be wrong or may be used in the address 6...

Page 20: ... The numeric shows the setting switch number Be sure to set either one of 1 or 2 to ON Solenoid polarity setting of electric locks It is settable with Switch 3 of DSW 2 POWER ALARM S3 S2 S1 CN1 IC102 E102 E101 CN504 CN503 CN201 CN101 1 2 3 4 5 6 7 8 9 RESET 0 MIN DSW1 DSW2 BZ1 VR1 MAX R1 R7 R6 R4 R5 1 1 9 OFF FSK 2 3 4 5 6 7 8 1 1 9 OFF FSK 2 3 4 5 6 7 8 1 9 3 OFF FSK 2 3 4 5 6 7 8 1 OFF FSK 2 3 4...

Page 21: ...gnal for oper ation start from the administration PC 4 Confirm the door closed state has been reset and READY is lighting on the iris camera 5 Turn off the control unit and close the door Notes When the power of the control unit is turned off no power is supplied to the electric lock In this case the status may vary depending on the lock type If you are using a Fail Safe type refer to p 20 the doo...

Page 22: ...ouble status In this case turn off the power immediately and refer to service personnel Refer to p 21 for how to turn on off the power Note When door trouble has occurred the electric lock control will be stopped In this case the status may be different depending on the lock type If you are using a Fail Safe type refer to p 20 the door can be opened with the lock deactivation However be careful wh...

Page 23: ...oduct is used for recognition enrollment of iris data ID and password authorization is also available Administration Software BM ES500E This product is necessary to administer iris data ID data and passwords ACCEPT REJECT READY Administration Software BM ES500E ...

Page 24: ...ivated SPECIFICATIONS OF ELECTRICAL LOCK CONNECTORS GATE 1 I O Signal Description Remarks 2 1 OUT RS 485 OUT RS 485 3 OUT CH1 Wiegand DATA1 4 IN CH1 Wiegand power 5 OUT CH1 Wiegand DATA0 6 OUT CH2 Wiegand DATA1 7 IN CH2 Wiegand power 8 OUT CH2 Wiegand DATA0 9 Wiegand GND SPECIFICATIONS OF WIEGAND OUTPUT RS 485 OUTPUT GATE 2 I O Signal Description 0 V to 5 V 0 V to 5 V 0 V to 5 V Wiegand I F 5 V 0 ...

Page 25: ...Alarm output 3 GND Open collector output Alarm output 4 Open collector output Alarm output 4 GND Open collector output Alarm output 5 Open collector output Alarm output GND Open collector output Alarm output 6 Open collector output Alarm output 6 GND Open collector output Alarm output 7 Open collector output Alarm output 7 GND Open collector output Alarm output 8 Open collector output Alarm output...

Page 26: ...lied to the power input connectors Check if the screws of the power input connectors are not loosened The power of the control unit or the power breaker of the external power control device may be turned off Turn on the power of the control unit or the external power control device Refer to p 8 The power of the external devices may be turned off Turn on the power of the external devices Refer to p...

Page 27: ...put contact signal 24 V DC 100 mA or less x 1 Emergency unlocking activation input contact signal 24 V DC 100 mA or less x 1 Wiegand interface Output x 2 5 V TTL level 25 mA or less Data Length 26 128 bit Pulse Width Time 30 40 µs Pulse Interval Time 1 2 ms RS 485 interface Output x 1 Baud Rate 2400 4800 9600 19200 38400 Parity none odd even Stop Bits 1 2 bit Data Bits 7 8 bit Alarm output Open co...

Page 28: ...Gabriel Industrial Park 65th Infantry Ave KM 9 5 Carolina P R 00985 809 750 4300 Panasonic Digital Communications Security Company Unit of Matsushita Electric Corporation of America Security Systems Group www panasonic com cctv Executive Office One Panasonic Way 3E 7 Secaucus New Jersey 07094 Zone Office Eastern One Panasonic Way Secaucus NJ 07094 201 348 7303 Central 1707 N Randal Road Elgin IL 6...