Chapter 2. Installing Pipe Nozzles
10
Model T17 Installation Guide
2.2.3 Determining and Marking the Second Nozzle Location, Diametrical (cont.)
7.
Place the strip of film on the pipe again. This time, line up the overlap mark with the horizontal and vertical scribe
lines. Again, make sure you wrap the strip of film squarely around the pipe.
8.
The new position of the center of the second nozzle is now identified as the intersection of the fold line and the
second edge of the strip of film. Center punch this location prior to removing the strip of film.
9.
Remove the film.
10.
Scribe 6” long vertical and horizontal lines that intersect at the center-punch mark.
Overlap Mark
First Edge
Side View
Fold Mark
Side View
Punch Here
Second Edge
Scribe Lines
Side View