![Panametrics DewPro MMR101 User Manual Download Page 15](http://html1.mh-extra.com/html/panametrics/dewpro-mmr101/dewpro-mmr101_user-manual_3403851015.webp)
DewPro MMR101 User’s Manual
9
Chapter 2. Installing the DewPro MMR101
2.3.3
Using a Bracket
When measuring ambient air, the MMR101 can be mounted directly on a wall, panel or pipe using an optional
mounting kit. Mounting kits can be ordered from the factory. Use Figure 3, below to properly install your unit in the
desired configuration.
Figure 3: Mounting Kit Configurations
2.3.4
Making Electrical Connections
The DewPro MMR101 is a loop-powered moisture and temperature transmitter for two- or four-wire connection being
supplied via the signal line (moisture signal). The first circuit transmits the moisture unit selected, the second the
process temperature. The DewPro can be optionally supplied from a common or two separate power sources
(1 2 ... 28 V DC). Additional power supply connections are not required.
6.50 in.
(165 mm)
2.76 in.
(70 mm)
4.45 in.
(113 mm)
3.23 in.
(82 mm)
4.45 in.
(113 mm)
2.95 in.
(75 mm)
0.79 in.
(20 mm)
4.45 in.
(113 mm)
Panel Mounting
Vertical Pipe Mounting
Horizontal Pipe Mounting
Horizontal Pipe Mounting