
BWT™ System Installation Guide
35
Chapter 4. Non-Flanged Installation (FSPA/FWPA Buffers)
Chapter 4. Non-Flanged Installation (FSPA/FWPA Buffers)
4.1 Introduction
The non-flanged
FSPA
and
FWPA
buffers (see
Figure 9 below) are available in either stainless steel or titanium. The
FSPA
is a threaded buffer that is screwed into the pipe coupling or nozzle, and the
FWPA
is a socket-weld buffer that
is welded into the pipe coupling. Both types of buffers enable you to install and remove transducers easily without
interrupting the process or emptying the pipe.
Installing the BWT transducers into an
FSPA
or
FWPA
buffer requires three steps:
• Installing the
FSPA
or
FWPA
• Installing and orienting the transducer junction box (see
Note:
For gas flow applications, a preamplifier is required. In such applications, the preamplifier may be mounted
in the transducer junction box.
• Inserting the BWT transducers (see
• Verifying the Installation (see
Figure 9: FSPA Buffer (Top) and FWPA Buffer (Bottom)
FSPA Threaded Mount
FWPA Socket-Weld Mount
Summary of Contents for BWT System
Page 1: ...panametrics com Flow 916 058 F July 2021 BWT System Installation Guide ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 8: ...Contents BWT System Installation Guide ...
Page 52: ...Chapter 5 Specifications 44 BWT System Installation Guide no content intended for this page ...
Page 54: ...46 BWT System Installation Guide no content intended for this page ...
Page 57: ......