BWT™ System Installation Guide
15
Chapter 2. Standard Installation (FTPA Buffers)
3.
Turn the nut on the second stud 1/8 turn. The second stud should be diametrically opposite the first stud. Proceed
to tighten the remaining studs in the order shown below, or in a similar manner.
4.
Turn each nut an additional 1/8 turn. Be sure to
TURN THE NUTS IN THE SEQUENCE SHOWN
.
5.
Table 1 below to determine the appropriate final torque for your studs.
Table 1: Final Torques
Flange Size
and Rating
Stud Rating with
4 Studs/Flange
Stud Diameter (in.)
x Length (in.)
Torque in
ft-lb (N-m)
316 Stainless Steel
1 1/2" ANSI 150#
B8M,C,1
1/2" x 3 1/4"
56 (76)
1 1/2" ANSI 300#
B8M,C,1
3/4" x 4 3/4"
82 (111)
1 1/2" ANSI 600#
B8M,C,1
3/4" x 4 3/4"
82 (111)
1 1/2" ANSI 900#
B8M,C,1
1" x 6 "
107 (145)
1 1/2" ANSI 1500#
B8M,C,1
1" x 6"
107 (145)
Carbon Steel
1 1/2" ANSI 150#
B7
1/2" x 3 1/4"
56 (76)
1 1/2" ANSI 300#
B7
3/4" x 4 3/4"
82 (111)
1 1/2" ANSI 600#
B7
3/4" x 4 3/4"
82 (111)
1
1
2
2
3
3
4
4
5
6
7
8
4-Hole Flange
8-Hole Flange
Turn first nut 1/8 turn.
1
1
2
2
3
3
4
4
5
6
7
8
4-Hole Flange
8-Hole Flange
Turn second nut 1/8 turn.
Summary of Contents for BWT System
Page 1: ...panametrics com Flow 916 058 F July 2021 BWT System Installation Guide ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 8: ...Contents BWT System Installation Guide ...
Page 52: ...Chapter 5 Specifications 44 BWT System Installation Guide no content intended for this page ...
Page 54: ...46 BWT System Installation Guide no content intended for this page ...
Page 57: ......