7.23  Equipotential connection 

The electric beds are equipped with equipotential connecting clamp placed on the side of 

the bed head; the clamp is necessary for the equalization of the electrical potentials of all 

the metal parts without protection. 

 

WARNING! DANGER OF ELECTRICAL DEVIATIONS. It is necessaryto usefully 

equipotential connection cauldron if the patient is connected to intravascular or intracardiac 

equipment. The cable must be connected to the equipotential connecting clamp located on 

the bed; then it is necessary to connect the latter to an appropriate equipotential terminal. 

 

7.24  Pull-out network plan 

Network top in technopolymer easily removable for thorough disinfection and cleaning. 

 

 

 

Summary of Contents for SYNAGO

Page 1: ...SYNAGO BED WITH VARIABLE HEIGHT 727T0045 727T0045C...

Page 2: ...nical data sides lower section of the legs 02 07 2018 Update general description technical data sides lower section of the legs 0 2 23 10 2019 Insertion variant 727T0045S 0 3 14 04 2021 CE marking acc...

Page 3: ...ying 9 2 7 Identification of commands 9 3 SAFETY 10 3 1 Safety standards 10 3 2 General provisions 10 4 GENERAL DESCRIPTION 11 4 1 Description of the project 11 4 2 Technical characteristics 14 5 PUSH...

Page 4: ...the upper leg section 33 7 10 Simultaneously raise and lower the back section and the upper leg section Autocontour 35 7 11 Trendelenburg e reverse trendelenburg 37 7 12 Comfort position chair 39 7 13...

Page 5: ...food additives 52 9 4 Intervallinification 53 9 5 Automatic sanitization 53 9 6 Manual sanitization 53 10 MAINTENANCE 54 10 1 Periodic check 54 10 2 Technical assistance 54 10 3 Provision 54 10 4 Demo...

Page 6: ...6 ID08435 Rev 1 2 19 04 2023...

Page 7: ...oses without committing to updating this documentation The illustrations and images in this manual are intended only as examples and may differ from practical situations The content of this Manual com...

Page 8: ...dy mass index BMI of 17 or more 2 3 Essential performance of the bed The essential benefits of the medical bed are Trendelenburg location the Trendelenburg location can be reached in any condition via...

Page 9: ...ID08435 Rev 1 2 19 04 2023 requirements of this manual and contacting the assistance of Pam Mobility s r l whenever a bed failure occurs After 10 years of use it is recommended to replacethe interolet...

Page 10: ...of production G Manufacturing data H Warning Read the installation and operating instructions I Electrical data L Vector UDI not present M Basic UDI DI number N Serial number O CE marking P Name and...

Page 11: ...re during use of the bed follow the maintenance instructions in this manual in addition to the safety requirements for personnel established by the Head of the Emergency Medical Service Operators shou...

Page 12: ...elenburg 3 joints 4 sections electrically adjustable top in technopolymer shoulder side head independent The electric bed has been designed for use in specialist or semi intensive care units For the p...

Page 13: ...82700018 3 Bilateral pedals optional cod 82700077 4 Control panel 5 Base 6 Telaio rete 7 Headboard 8 Footboard 9 Back section 10 Basin section 11 Upper section of the legs 12 Lower section of the legs...

Page 14: ......

Page 15: ...oxy painted steel covered by an ABS casing 6 Frame net Longitudinal metal structure on which the different sections rest 7 8 Headboard and footboard Headboard and footboard are the end part of the bed...

Page 16: ...deg 0 15 Bottom section of the frame deg 0 30 Safe workload kg 270 Patient weight kg 230 Standard wheel diameter mm 150 Suggested dimensions mattress mm 2000 x 850 x 180 h Weight kg 130 3 4 2 Electric...

Page 17: ...follows Raise the back To Lower the backrest B Raise the upper section of the shankseC Lower the upper leg section D Raise the lower section of the legs E Lower lower leg section F Simultaneous raisi...

Page 18: ...reen led on the edge button turns off Inhibition commands To unlock commands from console proceed as follows Press the A button in with the lock symbol closed and holding the Open button also the move...

Page 19: ...euvre to be carried out Description Lock inhibition unlocking movements A Control of movement B Control of movement C descent movement of the upper section of the backrest D movement of the upper leg...

Page 20: ...18 ID08435 Rev 1 2 19 04 2023 4 5 Control panels for kit 4 foldable sides NOT PROVIDED UNFORESEEN 4 5 1 Inhibition of commands from the control panel UNFORESEEN...

Page 21: ...hat the material is intact has not been damaged during transport or has not been opened voluntarily to remove parts inside Check that the delivery corresponds to the specifications of the order and ve...

Page 22: ...install the bed without the support of a qualified technician of Pam Mobility s r l Similarly it is absolutely forbidden to disassemble the bed for a subsequent reinstallation without the support of...

Page 23: ...ith compass proceed as follows mount the clamp 1 on the traverse 2 of the bed frame fixing it with screws 3 washers 4 and self locking nuts 5 included Insert the tailgate into the pins of the clamps 6...

Page 24: ...G Verify that the voltage and the network frequency correspond to the one to which the article has been prepared see identification plate prepare a SCHUKO socket connect the plug to the power supply W...

Page 25: ...of the bed when there is a patient above During the stop always lock the bed by applying the brakes When the PATIENT s condition such as disorientation due to medication or particular clinical conditi...

Page 26: ...n all the sections of the bed are zeroed and the bed base moves to the Trendelenburg position with the sides lowered To bring the bed to the EMERGENCY POSITION 1 proceed as follows press the button of...

Page 27: ...he bed sections press the button of the push button to bring the Treendelenburg into position lower the sides see reference paragraph FROM A CONTROL PANEL press the button of the M button until the po...

Page 28: ...s follows ensure that the sides are raised remove the power plug and wrap the cable release the brakes Push or pull the handle by grabbing the headboard or footboard at the end of the journey lock the...

Page 29: ...means of an audible alarm if the brakes are not inserted 6 7 Courtesy light device accessory 82700086 or included in the panel kit sides 82700087 LED night courtesy light placed under the bed frame o...

Page 30: ...the height of the bed it is necessary that the dedicated command is enabled see dedicated paragraphs in the chapter PUSH BUTTONS WARNING Do not place hands or objects between the base and the moving...

Page 31: ...ecessary that the dedicated command is enabled see dedicated paragraphs in the chapter PUSH BUTTONS WARNING Do not interpose the hands or objects between the backrest and the frame of the bed base Do...

Page 32: ...the dedicated command is enabled see dedicated paragraphs in the chapter PUSH BUTTONS WARNING Do not interpose the hands or objects between the upper section of the legs and the frame of the bed base...

Page 33: ...command is enabled see dedicated paragraphs in the chapter PUSH BUTTONS WARNING Do not interpose the hands or objects between the sections and the frame of the network plane Do not manually intervene...

Page 34: ...l box a control panel and external control panels for the sides To adjust the position of Trendelenburg the dedicated command must be enabled see dedicated paragraphs in the chapter PUSH BUTTONS To ad...

Page 35: ...ernal control panels for sides To regulate the position of the control panel it is necessary for the dedicated command to be enabled see dedicated paragraphs in the chapter CONTROL PANEL To adjust the...

Page 36: ...ctric actuators controlled by a control panel and internal and external control panels for sides WARNING Do not insert the hands or objects between the moving sections and the frame of the network pla...

Page 37: ...ernal control panels for sides and for the exclusive use of the operator WARNING Do not insert the hands or objects between the moving sections and the frame of the network plane Do not manually inter...

Page 38: ...control panels for sides and for the exclusive use of the operator ATTENTION The emergency switch can be activated ONLY if the override switch is in the unlock all controls position To adjust the pos...

Page 39: ...t to be lowered quickly in case of emergency and can be activated by means of a bilateral lever positioned under the bed base To lower the backrest proceed as follows unlock the backrest by grabbing i...

Page 40: ...external control panels for the exclusive use of the operator ATTENTION The safe position is activated SOLO if the override selector is in the unlock all commands position To bring the bed to the sche...

Page 41: ...ection of the legs to the desired position verifying the perfect locking The rack offers the possibility to adjust the lower section of the legs up to 6 different positions To lower the lower section...

Page 42: ...unscrewing the knobs 18 placed on both sides of the bed and pull the section outwards until the end of the stroke Fig 2 Turn the knobs 18 to lock the section Note The knobs need not be unscrewed compl...

Page 43: ...ts To lower the sides proceed as follows grab the bank with one hand and with the other lower the lever in the direction indicated by the arrow 15 lower the bank by turning it downwards To lift the si...

Page 44: ...dal letto Per sollevare le sponde procedere come segue Assicurarsi che il letto sia frenato vedi pag 15 del presente manuale Afferrare il tubo superiore della sponda K e sollevarlo nella direzione ind...

Page 45: ...the clamp pins have left the seat repeat for the other bank place the sides in a safe place WARNING Incompatible banks can create risks WARNING it is absolutely forbidden to mount and install the lift...

Page 46: ...protection WARNING DANGER OF ELECTRICAL DEVIATIONS It is necessaryto usefully equipotential connection cauldron if the patient is connected to intravascular or intracardiac equipment The cable must b...

Page 47: ...atient lifting rod made of chromed steel tube Adjustable strap and molded trapeze Cod 82700025 HEIGHT ADJUSTABLE DRIP ROD 4 HOOKS IV rod adjustable in height to 4 hooks Cod 82700016 HEIGHT ADJUSTABLE...

Page 48: ...accurate checks before using these solutions If the device is to be sanitised using halogen e g chlorine sanitising products the following requirements shall be followed the pH must be above 10 the te...

Page 49: ...at least two handling cycles for each of the adjustments made on the bed 9 2 Technical assistance Requests for assistance from the customer service department must be sent by fax or e mail to the foll...

Page 50: ...0 12 hours with completely discharged battery It is advisable to make sure that the batteries do not drain completely but recharge them frequently in order to achieve a longer service life Batteries m...

Page 51: ...ge the battery pack Defective battery Replace the battery pack A control unit output allows only one direction of advance to the actuator connected It s a faulty keypad Send the keypad to repair It s...

Page 52: ...ON to Control unit TC21 b Buffer battery TBB2 4398 001 0 c ACO TNP6 4398 001 d Push button panel TH12 and Trolley lift actuator TA23 f Actuator back TA1 g Basin actuator TA31 h Foot actuator where app...

Page 53: ...Via Verdi 39 42043 Gattatico RE Tel 39 0522 473859 Fax 39 0522 1548244 info pammobility com www pammobility com...

Reviews: