23
ID08435 - Rev. 1.2 - 19/04/2023
6. OPERATION AND USE
6.1 Warnings
Electric beds can not be used in potentially explosive or flammable atmosphere
(hyperbaric chamber type).
Before moving the bed, make sure that the power cord is disconnected and attached to the
bed.
Sanitize The reader according to the modalities described in the QUICK chapter.
It is the responsibility of the attending staff to authorize the patient to use the functions of the bed.
The electric part is designed for a continuous use of 2 minutes with an interval of 18
minutes between one use and the other as shown in the identification plate on the
control unit. Using the bed without respecting these constraints does not pose any
danger for either the patient or the operator but can damage the device.
Inform the patient whenever bed adjustments are made.
Always lift the safety rails of the bed when there is a patient above. During the stop always
lock the bed by applying the brakes.
When the PATIENT’s condition (such as disorientation due to medication or particular
clinical conditions) can lead to ENTRAPMENT OF THE PATIENT with THE
SIDES/SHOULDERS, The MATTRESS SUPPORT PLATFORM must be left in a safe
position with horizontal bed base and lowered when the patient is left alone (except when
requested to be provided to medical staff for special or special circumstances).
Do not use the bed for purposes other than those for which it was intended and for which it was
designed.
Beds should only be used with patients who meet the following parameters: weight
greater than or equal to 40 Kg, height greater than or equal to 146 cm and body mass
index greater than or equal to 17 (see label below).
6.2 Secure position
The bed is in a safe position when the bed base is in a horizontal position in the lowest
position with the sides raised, the controls by push-buttons disabled, the extension bed (if
available) closed and the brake inserted.