ID05449 - Rev. 0.1 - 01/12/2020
7
2. GENERAL WARNINGS
2.1 Manufacturer
The item described in this manual is manufactured by:
2.2 Intended use
The device is intended to be used exclusively as a stretcher for the transportation, diagnosis,
treatment and monitoring of patients under the close supervision and surveillance of medical
personnel.
The device cannot be used for inpatient purposes
Environment of use: within healthcare and health facilities.
The device cannot be used in a potentially explosive or flammable atmosphere.
Personnel intended for use of the product: specialist operators and doctors.
2.3 Essential performance
The aspects of essential performance of the stretcher include:
▪
Height adjustment:
by means of a button control it is possible to electrically adjust the
height of the bed surface.
▪
Backrest inclination adjustment:
the stretcher is equipped with a manually operated
mechanism with millimetre adjustment which, once released, allows adjustment of the
inclination of the backrest section.
▪
Trendelenburg/Reverse Trendelenburg:
by means of a button control it is possible to
electrically adjust the inclination of the bed surface until the Trendelenburg/Reverse
Trendelenburg position is obtained.
▪
Longitudinal translation:
the stretcher is equipped with a manually operated mechanism
with millimetre adjustment which, once released, allows sliding of the bed surface in order to
maximise the examination area.
2.4 Environmental limits of use
WARNING! The stretcher cannot be used in a potentially explosive or flammable
atmosphere.
The environmental working conditions of the stretcher must respect these indications:
▪
Temperature: 0°C ÷ +40°C
▪
Humidity: 10% ÷ 70% (non-condensing).
The device must be placed in an absolutely dry environment.
Environmental conditions other than those indicated may cause serious damage to the stretcher.
Positioning of the stretcher in environments that do not correspond to what is indicated will
invalidate the warranty.
2.5 Expected lifespan
The stretcher has been designed and built to operate without risk to property and persons under
the ordinary conditions of use defined in this manual for 10 years. However, this duration can
only be achieved by complying with the requirements set out in this manual and by contacting
Pam Mobility s.r.l. assistance whenever a malfunction occurs on the stretcher. After 10 years of
use it is advisable to replace the entire stretcher.