EmStat3-MUX16 for OEM
3
EmStat3-MUX16 for OEM
Important before you start
unpacking
Please read this quick start guide carefully before unpacking the boards from their
antistatic bags or using the kit, as it contains important information on how to avoid
damage to the PCB’s.
1
EmStat3 PCB
The EmStat-series are potentiostats with an embedded microcontroller. They provide all the major
potentiostatic techniques with automatic current ranging and peripheral control.
2
Connecting EmStat board to MUX16 multiplexer
EmStat3 PCB layout
The board has two pin headers: CON1 and CON2
CON1 is used to connect for instance the MUX multiplexer or any other peripheral.
CON2 is used when serial TTL communications is used and provides an external analog input and analog
output line, with a range of 0 - 4.095 V and a digital input line.
The EmStat and MUX boards can be damaged by static electricity
(electrostatic discharge or ESD). Please take adequate precautions against
static discharge during handling.
See document “EmStat3 for OEM.pdf” for more information about interfacing with the
EmStat3 board.