background image

020-17469-00 Instructions For Use Jet Mixer RevD

 


 

  Push the driver into a horizontal position 

 

 

Figure 10: Correctly installed drive unit 

 

  Lock the 2 screws to fixate the system. 

 

 

Figure 11: Locking screws 

 

Figure 12: Correctly installed drive unit in final position 

 

  Place the TELEMODUL control unit holder on the side of the tank. As the non-

jacketed and jacketed tanks have a different edge design, the holder has a 
positioning tool. The 2 discs on the side of the holder can be rotated. In this way the 
drive unit can always be positioned vertical. 

 

Summary of Contents for Jet Mixer

Page 1: ...020 17469 00 Instructions For Use Jet Mixer RevD Filtration Separation Solution SM Instructions For Use Jet Mixer...

Page 2: ...of the mixing biocontainer 13 5 1 Installing the mixing biocontainer 13 5 2 Filling the mixing biocontainer 14 5 7 Starting the mixing cycle 15 5 8 Emptying the mixing biocontainer 15 6 Cleaning main...

Page 3: ...ly followed and where appropriate they should be incorporated into the user s standard operating procedures If some of the procedures do not suit your needs please consult your Pall distributor before...

Page 4: ...drive unit and control unit be disassembled from the system to avoid damage due to vibrations For limited on site transportation this is not required and the control unit and drive unit may remain on...

Page 5: ...magnetic drive unit is positioned Aligns the drive unit in the correct position underneath the tank Control unit holder Stainless steel holder in which the control unit is placed Houses the control u...

Page 6: ...020 17469 00 Instructions For Use Jet Mixer RevD www pall com biopharm 6 Figure 2 Control unit holder Figure 3 Drive unit Figure 4 Control unit...

Page 7: ...gure 6 Back of control unit 6 Master power On Off button 7 Power cable female socket 8 Communication cable female socket Assembly To assemble the equipment follow the next steps Lock the wheel brakes...

Page 8: ...nit into the drive unit holder with the cable on the open end Figure 8 Inserted drive unit The driver has 4 rubber stoppers at the bottom that fit into the 4 square openings of the holding plate Inser...

Page 9: ...stem Figure 11 Locking screws Figure 12 Correctly installed drive unit in final position Place the TELEMODUL control unit holder on the side of the tank As the non jacketed and jacketed tanks have a d...

Page 10: ...se Jet Mixer RevD www pall com biopharm 10 Figure 13 Control unit holder on a jacketed tank Figure 14 Control unit holder on a non jacketed tank Insert the control unit into the control unit holder Fi...

Page 11: ...the communication cable from the drive unit to the control unit and connect the power cable Figure 16 Connected power and communication cable Plug the power cable into the mains electricity and switch...

Page 12: ...and stop button adjustable startup time 0 60sec continuous operation or timed operation 0 60min It is advised to use the function of the start up time especially when high speed is used A gentle accel...

Page 13: ...ms such as y pieces or filters the small drain cover plate around the drain hole can be removed If the mixing biocontainer incorporates an EZ Drain fitment position it in the drain port cutout install...

Page 14: ...sterile application should be done only via a sterilizing grade vent filter integral to the mixing biocontainer OPTIONAL Any sensors pH conductivity probe etc should be installed before filling the m...

Page 15: ...ot be OK and the impeller will not rotate correctly If this is observed at this stage one can still gently manipulate and reposition the biocontainer When activating the impeller start at low speed an...

Page 16: ...liquid can be recovered by gently lifting the mixing biocontainer so as to direct the residual liquid to the drain Close all clamps and detach the mixing biocontainer from all external connections Rem...

Page 17: ...and maintenance considerations of the drive unit Mobil 200 Thermo Scientific and control unit Telemodul 80 Thermo Scientific please consult the Operating Manual of these systems 6 3 Transport The tan...

Page 18: ...7 4 Operating temperature The operating temperature range is 0o C to 60o C at max 80 humidity For more details please consult the specific equipment manuals doc ref 020 17473 00 8 Service The mixing...

Page 19: ...red and installed and operated as per the specifications and design conditions stated in this document will be free from defects in material and workmanship during their shelf life Pall liability unde...

Reviews: