23
10.
Draining
Draining can be performed in a number of ways according to the specific application requirements.
Table 1 below gives drainage approaches according to application requirements.
Before draining the mixer, follow the below instructions:
1. Connect the outlet port to the desired downstream equipment or receptacle according to the specific
fitting on the single-use system (e.g. Kleenpak sterile connector or quick connector).
2. Open the outlet port valve by gently rotating the flush port bayonet valve to a stop, allowing fluid to
flow out of the Allegro mixer biocontainer.
CAUTION!
Take care to not over-rotate, apply excessive force, or pull the flush outlet valve during
operation. This may cause it to detach from the mixer. Use MINIMUM force when rotating the bayonet
valve to open the outlet line.
Table 1
Draining the mixer system
Mixer rotation
‘Bag Inflation’ Setting
When to Use
Comments
On
ON
To keep the mixer inflated
Once the mixer is empty, ‘Bag Inflation’
Then OFF
Then VENT
and to continue mixing
can be switched to ‘Vent’ to allow
during dispensing the
collapse of the bag prior to removal.
mixed volume
Once ‘Vent’ has been selected, the top
of the bag should be gently pushed down
into the mixer tote to free the top hat
from the upper gasket to vent the
biocontainer and allow removal.
OFF
ON (until fluid dispensed)
To keep the mixer inflated
As above
Then VENT
during dispensing the mixed
No agitation of fluid during dispensing
volume
OFF
OFF
To allow the mixer to collapse
The top of the mixer SU system should
during dispensing the mixed
be gently pushed down into the mixer
volume
tote to free the top hat from the upper
gasket at the start of the fluid dispensing
phase. Allows for removal of the single-
use mixer system as soon as fluid
volume is fully dispensed from the system.
Note: Fluid dispensing is normally done using a peristaltic pump on the outlet tubing, especially when a filter is being used down-
stream of the mixer chamber.
The floor of the Perspex tote is specifically designed to
provide very high yield/fluid recovery during draining.
Summary of Contents for Allegro 50 L
Page 1: ...USD2915a Instructions For Use Allegro 50 L Single Use Mixer ...
Page 30: ...29 20 Appendix 2 Wiring Diagrams 21 Appendix 3 General Assembly Drawings ...
Page 31: ...www pall com biopharm 30 ...
Page 32: ...31 ...
Page 33: ...www pall com biopharm 32 ...
Page 34: ...33 ...
Page 35: ...www pall com biopharm 34 ...