1
Take the tri-filter assembly and load a Hydrogen
Sulphide filter in the base of the assembly.
2
Take a red filter paper holder and
load with a destruction paper.
3
Take a black filter paper holder
and load with a detection paper.
4
Switch the Arsenator on by pressing any key.
5
Place the loaded black filter paper holder into the
Arsenator to blank and start the 20 minute timer.
6
Load the tri-filter
assembly with the
red and black filter
paper holders.
7.1 Operating Method
7.0 Introduction
The
Digital Arsenator
is
designed for field or laboratory
determination of Arsenic
in the range 2-100µg/l As.
Using the modified Gutzeit
method and the unique
tri-stage filter assembly,
the Digital Arsenator will
provide accurate and safe
measurement of Arsenic.
51
7
Digital Arsenator