4.4 System Menu
The System menu allows the user to set the
Compact Turbimeter preferences and review
results and calibration data.
4.4.1 Log
Select
‘Results’
or
‘Calibrations’
.
Once the log has been chosen, select
‘View’
or
‘Clear’
.
The
‘Readings’
log holds up to 100 data points
with date, time, Sample ID, Operator ID, Reading
Mode and Result.
The oldest result will automatically be
overwritten when the log is full.
The
‘Calibrations’
log stores the date, time,
method and Operator ID (if set) for the last
12 good calibrations.
To view data select
‘View’
and use the
Up/Down
buttons to scroll in either direction.
Use
‘Clear’
in either log to delete the entire log.
Select
‘Clear’
then
‘Yes’
in the following screen
to delete.
4.4.2 Operator ID
Optionally select or edit the username using
alphanumeric characters.
To create a new Operator ID select a blank
line and press
‘OK’
.
Press
‘New’
to create the ID.
Use the
Up/Down
buttons to
Show/Change
characters.
When the correct character is shown, press and
hold the [+] key for one second to move to the
next character.
Correct mistakes by holding the
[Del]
key for
one second. Complete the process by pressing
‘Done’
briefly. Select
‘OK’
to accept the entry
or
‘Edit’
to modify the entry.
To modify or delete an existing Operator ID,
highlight the ID and select
‘Edit’
Choose either
‘Edit’
to modify the existing entry or
‘Delete’
to remove it from the list.
Deleting the Operator ID will not affect results
stored in the log. Up to 12 Operator IDs can be
stored and recalled from memory as required.
To return to the System menu, either select the
Operator ID required using
‘OK’
or highlight a
blank entry (if one is available) and press
‘Back’
.
4.4.3 Sample ID
Optionally set the Sample ID using the same
method as Operator ID.
Up to 24 Sample IDs can be stored and
recalled as required.
Both Operator ID and Sample ID are stored
in the log with result and calibration data.
4.4.4 Units
Results can be displayed in NTU, FTU and FNU.
4.4.5 Language
Select English, French, German, Spanish or Italian
and press
‘OK’
to switch to alternative languages.
4.4.6 Set Time
Increase/decrease the time by using the
Up/Down
buttons. When the correct time is shown select
‘OK’
.
4.4.7 Set Date
Increase/decrease the date by using the
Up/Down
buttons. When correct select
‘OK’
.
4.4.8 Date Format
Select
DD/MM/YYYY
or
MM/DD/YYYY
as required.
4.4.9 Version
The serial number of the instrument and the
software version will be displayed.
The instrument serial number is required for
support and warranty claim.
4.4.10 Backlight
Select either
‘Off’
or
‘On’
as required. The
Compact Turbimeter will maintain the current
status until changed.
4.4.11 Start Page
The
‘Start Page’
allows the user to define a choice
of initial screens/modes on powering the Compact
Turbimeter. To change start page, highlight the
favoured choice for initial screen - Mode Menu,
Normal Reading Mode or Last Reading Mode.
The new Start Page choice will appear when the
instrument is next powered on.