palintest Pooltest 3 Manual Download Page 5

 

 

10

Care and Maintenance 

The  handling  of  photometer  tubes  is  important  to 

ensure  continuing  accuracy.  Scratches,  finger-prints 

and  water  droplets  on  the  tube  or  inside  the  light 

chamber  can  cause  inaccurate  results.  It  is  imperative 

that  the  tubes  and  light  chamber  are  clean  and  dry. 

The glassware must be clean and defect-free. Scratches 

and  abrasions  will  permanently  affect  the  accuracy  of  the 

readings. Tubes can be acid washed periodically. 

Here  are  some  hints  on  keeping  the  photometer  clean, 

free from contamination and in good working order :- 

1

Prepare  your  workplace  before  use.  Make  sure  that 

you  have  enough  space  to  work  with  the  photometer 

and with the reagent systems. 

2

Do  not  pour  out  samples  or  prepare  the  tests 

directly over the instrument.  

3

Always  cap  the  test  tubes  after  preparing  the  blank 

and test sample. 

4

Wipe  test  tubes  on  a  clean  tissue  to  remove  drips 

or condensation before placing in the photometer. 

5

Do  not  leave  tubes  standing  in  the  photometer  test 

chamber. Remove the tubes immediately after each test. 

6

Immediately  wipe  up  any  drips  or  spillages  onto  the 

instrument or into the test chamber with a clean tissue. 

7

Keep  the  instrument  clean.  Clean  the  test  chamber 

regularly using a moistened tissue or cotton bud. 

 

 

11

 

8

Keep  the  instrument  in  a  clean,  dry  place  when  it  is 

not  in  use.  Keep  it  on  a  clean,  dry  bench  away  from 

chemicals, place it in a storage cupboard or keep it  in 

a carrying case. 

  

Viewing the Instrument Memory 

To  view  the  previous  results  (10  results  are  stored 

within the instrument memory), hold the ‘menu’ button 

down 

for  at  least  three  seconds

  whilst  the  display 

showing the selected test or result is on the screen.  
 

To scroll through the 10 different results, simply 

press  the  ‘menu’  button.    To  exit  the  results 

menu,  press  the ‘menu’  button again for at least 

three seconds.

 

 

 

Backlight Operation 

The  backlight  can  be  turned  on  and  off  by  holding  the 

‘On/Off’ button for a two second period during power on.   

 

Summary of Contents for Pooltest 3

Page 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Page 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Page 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Page 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Page 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Page 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Page 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Page 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Page 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Page 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Page 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Page 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Page 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Page 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Page 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Page 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Page 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Reviews: