Screen 

“Read Sample” button 

“Menu” button 

“Blank Sample” button  

“On/Off” button 

 

 

7

 

2  

BACKGROUND INFORMATION 

Introduction to Colorimetric Measurement 

Palintest  tests  are  based  on  measuring  the  intensity  of 

colours 

produced 

by 

Palintest 

reagents 

and 

using 

Palintest photometers to measure that intensity of colour. 

This  is  colorimetry  and  can  be  defined  as  any  technique 

used  to  evaluate  an  unknown  colour  in  reference  to 

known colours.  

To  avoid  subjective  measurement  between  test  samples 

and  colour  standards,  a  colorimeter  can  be  used  for 

quantitative  measurement  of  the  amount  of  coloured  light 

absorbed  by  a  sample  (with  reagents  added)  in  reference 

to a untreated sample (blank). 

White  light  is  made  up  of  many  different  wavelengths  of 

light. 

A  colorimeter  passes  a  white  light  beam  through  an 

optical  filter  which  transmits  only  one  particular  band 

of  wavelength  of  light  to  the  photodetector  where  it  is 

measured. 

The difference in  the amount of  coloured  light  transmitted 

by a colourless sample (blank) and the amount of coloured 

light  transmitted  by  a  coloured  sample  is  a  measurement 

of the amount of coloured light absorbed by the sample.  

Summary of Contents for Pooltest 3

Page 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Page 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Page 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Page 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Page 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Page 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Page 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Page 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Page 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Page 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Page 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Page 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Page 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Page 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Page 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Page 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Page 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Reviews: