33
Features and Technical Specification
Application
For application in swimming pool and spa testing
using Palintest tablet reagent systems
Instrument Type
Single-beam colorimeter with built-in colour filters
and pre-programmed test calibrations
Peak Wavelengths
445
5nm, 495
5nm, 555
5nm, 570
5nm, 605
5nm and 655
5nm
Range
1 - 100%T
Accuracy
1%T
Display
Large graphic display with option of backlight.
Language
Test identification and prompts in English, French,
German, Spanish, Italian or Turkish
Chlorine Test Options
Selection of 0 – 5 or 0 – 10 ppm ranges
Water Balance Options Palintest BW Index of Langelier Index
Timer
Clock and timer feature to log test results and audible
alarm for timing test procedure
Units
Direct-reading of test results in mg/l or ppm
User Selectable
10 digit sample number entry, dilution factor, time/
date, date format, system lock, temperature and
rounding of results
Date Format
Date format selectable as day/month/year or
month/day/ year
Zeroing
Automatic zeroing on blank tube and hold blank
facility for series of tests. Continuation test facility
without the need for reblanking
Internal Memory
Stores 1000 previous readings with option to view
logged results on screen, or download to computer
USB Interface
USB 1.1 full-speed, bus-powered device. Software
selectable between either emulation of a removable
hard-drive or emulation of a serial device connected
via a virtual COM port
Power
3 x 1.5V ‘AA’ alkaline batteries or via USB interface.
Power management system with variable length auto
switch-off or 'continuous' operation
Size
250 x 150 x 70 mm
Options
Summary of Contents for Pooltest 25