pH/chlorine regulator AUTODOS 5000 - 2005-01-21
Eng_Autodos5000.doc ver 3.2 - page 7
ENTER CODE LEVEL
CODE
When the button
'CODE'
is pressed down, the opening code can be entered by using the buttons
’0’
–
’9’, ’RESET’
clears the display. If the entered code is the same as opening code 1, you will enter
code level 1and if the entered code is the same as opening code 2, you will enter code level 2 when
the button
'CODE'
is released. This is indicated by LED ’CODE’, if the LED ’CODE’ is flashing
you are in code level 2 and if there is steady light, you are in code level 1. The opening code can be
altered in calibration mode, this is described in the calibration chapter. If you have forgotten the
code, it can be displayed during the startup sequence. If the button
’CODE’
is pressed during the
count-down time, the code for code level 2 is displayed. If no codes are used, code level 2 is reached
directly when the button
‘CODE’
is pressed.
VIEW/ALTER SET VALUES
SET
VALUE
When the button
'SET VALUE'
is pressed down, the set values can be altered by using the buttons
'
↓
' or '
↑
' below the displays for the respective channels. At least code level 1 is required for alteration
of the values.
VIEW/ALTER ALARM LIMITS
ALARM
HIGH
When the button
'HIGH ALARM'
is pressed down, the high alarm limits can be altered by using
the buttons '
↓
' or '
↑
' below the displays for the respective channels. At least code level 1 is required
for alteration of the values.
ALARM
LOW
When the button
'LOW ALARM'
is pressed down, the low alarm limits can be altered by using the
buttons '
↓
' or '
↑
' below the displays for the respective channels. At least code level 1 is required for
alteration of the values.
OTHER DISPLAY
VIEW
UNCAL
The uncalibrated value for all electrodes can be shown by pressing the
'FUNCTION'
and
'LOW
ALARM'
buttons simultaneously.
VIEW
FLOW
GUARD
To display the current status of the flow guard input (’
F-0
’ = flow OFF, ’
F-1
’ = flow ON) , press the
'SET VALUE'
and
'HIGH ALARM'
buttons simultaneously. If no flow guard is selected, there
will be no display.
VIEW
VERSION
To display the current program version press the
'HIGH ALARM'
and
'LOW ALARM'
buttons
simultaneously.
VIEW
SERIAL
NUMBER
To display the factory programmed serial number, press the
'FUNCTION'
and
'HIGH ALARM'
buttons simultaneously.