AFX SERIES® OPERATION MANUAL
SECTION 8: REMOTE PROGRAMMING
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 171 of 274
8.7
Source Commands
Source commands control the settings of the power source. This includes all operating
modes, voltages, frequency, current and power limits and phase angles as well as transient
operation. Since the SOURCE subsystem is the primary system, the SOURce portion of these
commands is optional. The following status commands are supported broken down by
SOURCE sub groups.
8.7.1
Source Configuration Programming Commands
Command Syntax
[SOURce:]INITial <n>
Description
Sets source output settings initialization mode. This command in
combination with the “OUTPut[:STATe]:AUTO” command allows the
unattended resumption of a test station after a power failure.
Note:
This condition is potentially hazardous and should be used with
caution.
[SOURce:]INITial? = ON Set points of the unit at power on will be the last set after power off.
The set points affected by this command are:
- Frequency
- Voltage AC1
- Voltage AC2
- Voltage AC3
- Voltage DC1
- Voltage DC2
- Voltage DC3
- Form (THREE , SPLIT or SINGLE)
- Range (High or Low)
- Compatibility Mode (UPC or NORMAL)
[SOURce:]INITial? = OFF Set points of the unit at power on will have a default value of:
- Frequency = 60Hz
- Voltage AC1 = 0V
- Voltage AC2 = 0V
- Voltage AC3 = 0V
- Voltage DC1 = 0V
- Voltage DC2 = 0V
- Voltage DC3 = 0V
- Form = THREE
- Range = HIGH
- Compatibility = NORMAL
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
:INIT ON
Query Format
[SOURce:]INITial?
Returned Data Format
<b>
Query Example
INIT?
0