Software description
& working principle
•
It is necessary for the rapid test manufacturer
to use software to set the detection range of
the C line and T line, and conduct experiments
through the cooperation of both parties; after
the experiment, the results are converted into
data settings in the reader, and then the
threshold for judging negative and positive is
defined.
•
The software itself has the function of analyzing
and processing images, and the interpretation
time is about 7 to 15 seconds; when the user
presses the analysis button, the reader will
shoot the image of the rapid test, and analyze
the C line in the reading range according to the
manufacturer's settings Line and T-line values
are compared to the threshold to give a
positive or negative result.