Hysteresis
4 BPM to 60 BPM
Hysteresis
Refractory Periods
From 200 to 500ms in steps of 25 ms
Refractory
Upper Rate
From 80 BPM to 180 BPM in steps of 2
BPM.(Behavior will be Wenckebach)
Trigger Upper Rate
PVARP
RATE RESPONSE TO ACCELERATION
Upper Rate
From 80 to 180 BPM in steps of 2 BPM.
Upper Rate
Coefficient of rate
response to acceleration
From 1 to 15
Slope
Reaction time
From 10 to 60 Sec. in steps of 10 Sec
Up Time
Recovery time
From 1 to 10 Min. steps of 1 Min.
Dn Time
Automatic Refractory
Change
On/Off. When programmed On, if the rate
is greater than 120 BPM, the refractory
switch to the minimum value between the
programmed one and 250ms
ARC
A Tachy Therapy
Enable:
On/Off
A Trigger Rate:
From 80 min-1 to 180
min-1 in steps of 2 min-1
Number of events for A-Tachy Entry:
4,
8, 16, 32
Number of events for A-Tachy exit:
4, 8,
16, 32
A Tachy. Therapy
Auto Polarity Switch
ON - OFF
Auto Pol. SW