Press the FPM/MPS key to switch the measurement units from FPM
(feet per minute, 1 FPM resolution) to MPS (meters per second, 0.01
MPS resolution).
Press the SAMPLE RATE key to change the measurement averaging
rate (“sample rate”) of the unit:
2ΣΕΧ
An average value of airspeed measurements during the
preceding 2 seconds is displayed.
4ΣΕΧ
An average value of airspeed measurements during the
preceding 4 seconds is displayed.
8ΣΕΧ
An average value of airspeed measurements during the
preceding 8 seconds is displayed.
16 Σ
An average value of airspeed measurements during the
preceding 16 seconds is displayed.
Press the MAX/MIN key to record and display the maximum airspeed
reading. The maximum airspeed reading display will alternate with the
letter “H” displayed with the sample rate. Press the MAX/MIN key
again to record and hold the minimum airspeed reading. The minimum airspeed
reading display will alternate with the letter “L” displayed with the sample rate.
For example:
1065
alternating with
Η 8
signifies that 1065 is the highest airspeed
reading since the MAX/MIN key was pressed, and the sample rate is set to 8
seconds.
82
alternating with
Λ 16
signifies that 82 is the lowest airspeed
reading since the
MAX/MIN key was pressed, and the sample rate is set to 16
seconds.
To exit MAX/MIN mode, press the SAMPLE RATE key.