I
NSTRUCTION
M
ANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: [email protected] Web: http://www.metallographic.com
11
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
NANO 1000T / FEMTO 1500
Polishing Head
Installation (continued)
3.6 Height Adjustment for FEMTO Head /Specimen Holder
Specimen
holder height
adjustment:
The specimen holder height needs to be adjusted to prevent the pistons from
grinding during the operation of the machine.
Step 1: Connect air and lift up head.
Step 2: Place the working wheel with the highest profile abrasive/polishing pad
on the wheel. Using a single wall box as a spacer, place the cardboard on top of
the working wheel/abrasive combination.
Step 3: Loosen holding ring collar on the FEMTO post and slide down out of the
way.
Step 1: To lift the
FEMTO head, rotate
the lever to engage
the piston
Step 2: Lower the
FEMTO head onto a
cardboard spacer
Step 3: Loosen and
lower post the
holding collar