4
E.5.
Check for signs of excessive dust around the cooling air intakes. Check the room
filters and replace the inserts before dust may obstruct the flow.
E.6.
Follow the instructions for the oxygen generator.
ANNUALLY:
Disconnect the ozone generators from power and check for excessive dust
inside. Use a long hairbrush and vacuum cleaner to gently remove. Failures caused by
dust are not covered under the warranty. Follow the instruction for the oxygen source.
LATER:
The cooling fans are rated for up to 100,000 hours of continues duty under ideal
conditions. Excessive dust may decrease the life span considerably.
Troubleshooting:
-
FUSE 1:
The lamp indicates a need for service (Main Fuse is blown).
-
FUSE 2:
The lamp is lit with short interruptions after ~ 20 seconds, indicating
a problem in the HV circuits. Ozone is not being produced.
The most likely reasons are:
. Water is inside the ozone electrodes
. Dirt or moisture is accumulated over a long period of operation around the HV circuits
causing HV discharges.
. The pressure or vacuum inside the ozone-generating cell are extremely high. This may
only be possible if the Balance Barometer BBL is not installed.
. The HV coil may have failed.
Except for the last item the digital fuse will recover once proper conditions for operation
are established.
-
FUSE 2:
The lamp flashes,
. Voltage in the power line is greater than 132V
The digital controls will restore the operation once proper operating line-voltage is
restored.