9. FAULTS AND TROUBLESHOOTING
10.CONDITION FOR TRANSPORTATION AND STORAGE
If your chamber fails to operate properly, please refer to the troubleshooting chart on the following pages for possible
causes and solutions. If the problems still can’t be solved, please contact your euipment provider.
Environment temperature scale : 10 °C to 40 °C
Comparative humidity scale : 30%-85%
Air pressure scale : 700hPa-1060hPa
14
Chamber pressurization very
slow
1. Air compressor joint or air inlet joint
quick connect may be loose.
2. Air filters full of dusts.
3. Air inlet on the air compressor is
clogged, The Air intake is no clear.
4. Air deflate valve is loose.
5. Tubing may not be fully inserted.
1. Didn’t close the zippers.
2. Didn’t close the air deflate valve or
didn’t close it fully.
3. Air tube is leaking air at one of the
joints.
4. The blue silicon sealing strip
between inside and outside zippers
is not flat.
1. Air filter is clogged.
2. Air inlet on the air compressor is
clogged, no smooth intake.
3. Air compressor screws loose.
4. Internal parts loose.
1. Air tube is leaking at the joint.
2. Didn’t close the air deflate valve or
didn’t tighten up.
3. Did not close one of the Zippers
completely.
1. Place your hand over all joints
to feel for an air leak.
2. Reinsert tubing if the connector
is leaking.
3. Replace new filters or clean the
dusts on the filters.
4. Clean the air inlet part of the air
compressor.
5. Tighten up the air deflate valve.
1. Depressurize the chamber first,
then, close the zippers and restart
2. Make sure no particles in
the valve holes, then close the
valve and tighten it up.
3. Stick the air tube in the connector.
4. Make the blue flap between
zippers is flat.
1. Make sure there are no cracks
or Replace the filters.
2. Clean the air inlet part of the air
compressor.
3. Tighten up the screws on air
compressor.
4. Check for looses screws in the
air compressor
1. Find out the leakage source,
reinsert the air tube in the
connector.
2. Close the pressure deflate valve
fully.
3. Close all zippers fully.
Pressure does not go to or hold
at proper KPa
Noisy sound or intense vibration
of air compressor
Chamber does not inflate at all.
TROUBLE
CAUSES
SOLUTION