9. FAULTS AND TROUBLESHOOTING

10.CONDITION FOR TRANSPORTATION AND STORAGE

If your chamber fails to operate properly, please refer to the troubleshooting chart on the following pages for possible 
causes and solutions. If the problems still can’t be solved, please contact your euipment provider.

Environment temperature scale : 10 °C to 40 °C
Comparative humidity scale : 30%-85%
Air pressure scale : 700hPa-1060hPa

14

Chamber pressurization very

slow

1. Air compressor joint or air inlet joint
    quick connect may be loose.
2. Air filters full of dusts.
3. Air inlet on the air compressor is
    clogged, The Air intake is no clear.
4. Air deflate valve is loose.
5. Tubing may not be fully inserted.

1. Didn’t close the zippers.
2. Didn’t close the air deflate valve or
    didn’t close it fully.
3. Air tube is leaking air at one of the
    joints.
4. The blue silicon sealing strip
    between inside and outside zippers
    is not flat.

1. Air filter is clogged.
2. Air inlet on the air compressor is
    clogged, no smooth intake.
3. Air compressor screws loose.
4. Internal parts loose.

1. Air tube is leaking at the joint.
2. Didn’t close the air deflate valve or
    didn’t tighten up.
3. Did not close one of the Zippers
    completely.

1. Place your hand over all joints
    to feel for an air leak.
2. Reinsert tubing if the connector
    is leaking.
3. Replace new filters or clean the
    dusts on the filters.
4. Clean the air inlet part of the air
    compressor.
5. Tighten up the air deflate valve.

1. Depressurize the chamber first,
    then, close the zippers and restart
2. Make sure no particles in
    the valve holes, then close the
    valve and tighten it up.
3. Stick the air tube in the connector.
4. Make the blue flap between
    zippers is flat.

1. Make sure there are no cracks
    or Replace the filters.
2. Clean the air inlet part of the air
    compressor.
3. Tighten up the screws on air
    compressor.
4. Check for looses screws in the
    air compressor

1. Find out the leakage source,

    reinsert the air tube in the

    connector.
2. Close the pressure deflate valve
    fully.
3. Close all zippers fully.

Pressure does not go to or hold

at proper KPa

Noisy sound or intense vibration
of air compressor

Chamber does not inflate at all.

TROUBLE

CAUSES

SOLUTION

Summary of Contents for Wheel Chair Ready Type Chamber

Page 1: ...PO Box 2301 Naperville IL 60540 USA 630 452 8431 SalesTeam HyperbaricPro com www HyperbaricPro com AIR COMPRESSOR AIR DEHUMIDIFIER...

Page 2: ...tem for the first time Keep this manual handy for your reference Please ensure that all items are removed from the box before trowing the box away Some customers have thrown away the protective paddin...

Page 3: ...aintenance and Cultivation 8 FAQ 9 Faults and Troubleshooting 10 Condition for Transportation and Storage 4 4 5 6 7 9 11 12 13 14 14 CONTENTS Installation Working Scheme Accessories Air Compressor and...

Page 4: ...e following table is a list of symbols and definitions that used with the Hyperbaric Chamber MILD HYPERBARIC OXYGEN THERAPY Authorized Representative In The European Community HOT Surface Do Not Touch...

Page 5: ...people with severe infectious diseases and patients 2 Bad colds Pneumothorax patients 3 Patients with severe emphysema 4 Hypertension exceeds 160 100ml 5 Severe heart problems 6 Severe COPD Mild COPD...

Page 6: ...de 10 Muffler Silencer 11 Blank Port reserved for future upgrades 12 Automatic Pressure Relief Valves 4 PRODUCT SPECIFICATIONS AND CONSTRUCTION 2 AIR COMPRESSOR Voltage 110 220V Air Flow 72 L Minute P...

Page 7: ...LY AND OPERATION 4 Dehumidifier Voltage 110 220V Capability 150 L Minute Pressure 0 8mPa Power 300 Watt Weight 5kg 12 LPS Working Scheme Installation 1 Please open the accessory box before installatio...

Page 8: ...Note Please install internal and external pressure gauges before use the internal pressure gauge should be installed inside the chamber external pressure gauge should be installed outside the chamber...

Page 9: ...the oxygen generator 1 LPM below the maximum setting 2 Get into the chamber and pull outside zippers and then the inside zippers closed and make sure to flatten out the silicone strip between the zip...

Page 10: ...ally and stablize the pressure betwen 30 KPa and 32 KPa 5 It is recommended to use 40 80 mins per session Pressure gauge error tolerance 3 KPa 40 80mins 6 After use please open the Air Deflate Valve t...

Page 11: ...t time if there is slight discomfort or pain in your ears This is a normal phenomenon caused by pressure increase You can improve your discomfort by yawning chewing or swallowing You can also slow pre...

Page 12: ...ure has stabilized When a session is finished unscrew the air deflate valve slowly to depressurize the chamber the pressure will drop slowly from 30 KPa to 0 KPa this depressurization process should b...

Page 13: ...ssure and avoid the feeling of fullness in your ears you will need to clear your ears by chewing gum or food etc swallowing yawning During the pressurization of the chamber your mouth may become dry O...

Page 14: ...h intake 3 Air compressor screws loose 4 Internal parts loose 1 Air tube is leaking at the joint 2 Didn t close the air deflate valve or didn t tighten up 3 Did not close one of the Zippers completely...

Page 15: ...To help you choose the right Hyperbaric Chamber that fits your needs One of our hyperbaric specialists will assist you Please Call 630 451 8821...

Page 16: ...This page intentionally left blank...

Reviews: