Resistance measurement 

W

 (See Figure 5)

UT220  OPERAT

ING MANUAL 

7. Do not measure the voltage or current higher than the allowable input value. Set the function range

switch at the maximum range position if the scope of measured value couldn't be defined. Prior to

measurement of on-line resistance and diode or the circuit on-off measurement, the power of circuits

being measured shall be powered off and all capacitors shall be completely discharged to assure the

measurement accuracy.

8. When LCD display shows the icon  " �"  it is required to replace the battery in time to ensure the

measurement accuracy. Take out battery when clamp meter is not in user for a long time.

9. Refrain from changing the internal wiring in the clamp ammeter to guard against damage to the meter 

and danger.

10. Refrain from storing or using the clamp ammeter in the explosive and flammable environment with

high temperature, high humidity and strong electromagnetic field.

11. Clean the clamp meter case with soft cloth and neutral detergent. To prevent causing corrosion to 

the case, or damage to the instrument and danger, it is forbidden to use abrasive material and solvent. 

IV El  t  . 

.

ec  nca  svm 

bl 

OS

IQ] 

Dual insulation 

Diode 

...J... 

Grounding 

Data hold

&. 

Warning prompt 

-

AC or DC (Alternating current or direct current) 

_,

AC (Alternating current) 

Danger! High voltage! 

-

DC (Direct current) 

CE 

Comply with EU standard 

---

.,,, 

Continuity

CATIV  Equipment is designed to protect against transients from the primary supply level (e.g. an 

electricity meter, primary overcurrent protection devices). 

Equipment is designed to protect against transients in equiftment in fixed installations (e.g.

CATIII  switches in fixed installation, industnal used equipments wi h permanent connection to fixed 

installations). 

UT220  OPERAT

ING 

MANUAL

V. External construction (See Figure 1)

1.

Clamp body: 

Safety  design to protect operator from

touching the dangerous area

2. Clamp head pulling handle: 

Press the trigger to open

the clamp head,  when release the trigger,  clamp head

will partially closed.

3. Functional key: 

Select basic functions

4. Measure input terminal: 

Measure the signal input

5. LCD display area: 

Display the measured data and

functional symbols.

6. Dial switch: 

Select the measurement function gear.

7. Clamp head: A 

sensing device used to measure AC/DC

current and covert current into voltage . 

Figure 1 

Resistance

Q

 Always make sure the power is off and 

all capacitors are fully discharged before 

measuring resistance.
A more accurate measurement result will be 

achieved by separating the component from 

the circuit.

1. 

Insert the black test lead into the COM 

input port and the red test lead into the 

Ω

” input port.

2.  Turn the dial function switch to “

Ω

“.

3.  Connect tip of the test leads to the 

points where the value of the resistance 

is needed.

4.  Read the result from the LCD panel.

Diode measurement (See Figure 6)

UT220  OPERAT

ING MANUAL 

7. Do not measure the voltage or current higher than the allowable input value. Set the function range

switch at the maximum range position if the scope of measured value couldn't be defined. Prior to

measurement of on-line resistance and diode or the circuit on-off measurement, the power of circuits

being measured shall be powered off and all capacitors shall be completely discharged to assure the

measurement accuracy.

8. When LCD display shows the icon  " �"  it is required to replace the battery in time to ensure the

measurement accuracy. Take out battery when clamp meter is not in user for a long time.

9. Refrain from changing the internal wiring in the clamp ammeter to guard against damage to the meter 

and danger.

10. Refrain from storing or using the clamp ammeter in the explosive and flammable environment with

high temperature, high humidity and strong electromagnetic field.

11. Clean the clamp meter case with soft cloth and neutral detergent. To prevent causing corrosion to 

the case, or damage to the instrument and danger, it is forbidden to use abrasive material and solvent. 

IV El  t  . 

.

ec  nca  svm 

bl 

OS

IQ] 

Dual insulation 

Diode 

...J... 

Grounding 

Data hold

&. 

Warning prompt 

-

AC or DC (Alternating current or direct current) 

_,

AC (Alternating current) 

Danger! High voltage! 

-

DC (Direct current) 

CE 

Comply with EU standard 

---

.,,, 

Continuity

CATIV  Equipment is designed to protect against transients from the primary supply level (e.g. an 

electricity meter, primary overcurrent protection devices). 

Equipment is designed to protect against transients in equiftment in fixed installations (e.g.

CATIII  switches in fixed installation, industnal used equipments wi h permanent connection to fixed 

installations). 

UT220  OPERAT

ING 

MANUAL

V. External construction (See Figure 1)

1.

Clamp body: 

Safety  design to protect operator from

touching the dangerous area

2. Clamp head pulling handle: 

Press the trigger to open

the clamp head,  when release the trigger,  clamp head

will partially closed.

3. Functional key: 

Select basic functions

4. Measure input terminal: 

Measure the signal input

5. LCD display area: 

Display the measured data and

functional symbols.

6. Dial switch: 

Select the measurement function gear.

7. Clamp head: A 

sensing device used to measure AC/DC

current and covert current into voltage . 

Figure 1 

Resistance

Q

 Always make sure the power is off and 

all capacitors are fully discharged before 

measuring resistance.
A more accurate measurement result will be 

achieved by separating the component from 
the circuit.

1. 

Insert the black test lead into the COM 

input port and the red test lead into the  

” input port.

2.  Turn the dial function switch to “

”, 

press the “SELECT” button once and  

” will appear on the LCD.

3.  Place the red and black test lead to the 

positive and negative terminals of the 

diode.

4.  The forward conduction voltage value 

will be indicated on the LCD.

5.  When the red and black test lead is 

inadvertently connected, the over range 

symbol “OL” is displayed.

Summary of Contents for OXD-516-1725D

Page 1: ...516 1725D CZ HU PL SK TH cromwell co uk Auto ranging AC current to 2000A DC voltage to 1000V AC voltage to 750V Audible continuity Resistance Diode testing 63mm jaw opening LCD backlight Maximum value...

Page 2: ...ing current f Danger High voltage DC Direct current CE Comply with EU standard Continuity CATIV Equipment is designed to protect against transients from the primary supply level e g an electricity met...

Page 3: ...ANUAL VII Key function and buzzer SELECT Function selection key Actuated by trigger method used for switching among the measurement functions at the gearQ i Q MAX Maximum value hold key Actuated by tr...

Page 4: ...easured 60mm diameter Electromagnetic field impact Using the Clamp Meter near an electromagnetic field may Result in an unstable display or an inaccurate reading Power 6F22 9V alkaline battery Dimensi...

Page 5: ...Input impedance 10M AC VOLTAGE Mean value response Input impedance 10M Frequency response 40 400Hz 400mV 50 100Hz RESISTANCE CONTINUITY TEST DIODE TEST AC CURRENT Mean value response Frequency respon...

Page 6: ...scope of measured value couldn t be defined Prior to measurement of on line resistance and diode or the circuit on off measurement the power of circuits being measured shall be powered off and all ca...

Page 7: ...or the circuit on off measurement the power of circuits being measured shall be powered off and all capacitors shall be completely discharged to assure the measurement accuracy 8 When LCD display sho...

Page 8: ...mary overcurrent protection devices Equipment is designed to protect against transients in equiftment i CATIII switches in fixed installation industnal used equipments wi h perm installations Resistan...

Page 9: ...UT220 OPERAT ING MANUAL 7 Do not measure the voltage or current higher than the allowable input value Set the function range switch at the maximum range position if the scope of measured value couldn...

Page 10: ...HandStrap See Figure 11 1 Open the hand strap 2 Slide the two tabs through the tether points 3 Fold the tabs back over themselves and press them down firmly 4 Close the strap by pressing it firmly in...

Page 11: ...11 NOTES...

Page 12: ...aw electrical goods that have reached the end of their life must be collected separately and returned to an environmentally compatible recycling facility Do not dispose of electrical goods with domest...

Reviews: