Fault finding
Symptom
Diagnosis and suggestions
“Lo N” appears on the
controller
There is less than 6 cm of liquid in the dewar. Do not
operate the Cryojet with less than 10 cm of liquid.
Refill the dewar, then switch the controller off and on
again.
“Hot 1” appears on the
controller
The nozzle temperature is above the hot limit, set by
default to 320 K. The Cryojet is only specified to
operate up to 300 K. Reduce the set temperature,
and switch the controller off and on again.
Alternatively, the nozzle sensor is faulty, or the cable
is not connected.
Controller will not stay in Auto
mode (Heater light will not
stay on)
Sample flow must be at least 2 l/min.
RAISE, LOWER and Heater
On/Off do not work
Controller is set to REMOTE. Press LOC/REM to restore
to LOCAL control (unless controller is LOCKED in
REMOTE mode).
The controller does not
respond to computer control.
The controller may be in LOCAL. If it is, use the
DISCONNECT and then the CONNECT command in the
instrument driver. This will put the controller into
REMOTE.