When LOCK is lit, the instrument is locked into either local or remote control and the
LOC/REM button has no effect. At power up, the controller is locked in LOCAL.
When the controller is in REMOTE, many of the front panel controls are disabled. Those
controls that only affect the display will still work, but those that could change the
operation of the instrument will not. If LOCK is lit whilst in REMOTE, all the front panel
controls are inoperative.
Note that there is no indication on the instrument driver as to whether the controller is in
LOCAL or REMOTE. If the controller is in LOCAL, it will not respond to computer
commands. To rectify this, use the DISCONNECT and then the CONNECT command in the
instrument driver. This will put the controller into REMOTE.
6.4
Restoring Standard Calibration Parameters
If any of the set-up parameters are changed by mistake, or the EEPROM (non-volatile
memory) becomes corrupted, the controller can be restored to the state in which it left the
factory, by using the back-up disc provided, and ObjectBench software. Use the following
sequence of operations and menu selections.
This feature is not available from the Cryojet instrument driver. Instead, you must add the
instrument driver called ITC503.drv (using Configure / Add/remove instrument / Add).
DISCONNECT the Cryojet driver if it is connected.
If the controller ISOBUS address is not the default value (1), or you are using a COM port
(serial port) other than COM1 on your computer, set up the ITC503 driver to the correct
address and COM port using Configure / Interface.
Press and hold the Shift key while clicking CONNECT in the ITC503 driver. 0.000 will
appear in the driver display.
In the ITC503 driver window, select SETUP/MEMORY.
LOAD ->
Enter the filename of the back-up file: a:
nnnnn
.ram, where
nnnnn
is the Oxford
Instruments Project Number of your Cryojet.
OPEN
PUT->
OK
Then STORE the restored data in the controller EEPROM. (Press and hold CAL/STORE and
press LOC/REM. NB This will interrupt the flow momentarily.)
DISCONNECT
If you want, remove the ITC503 driver (using Configure / Add/remove instrument).