Version 1.4 rev 13 Oct 2017
19
Newton
3.6 c
onnecting
tHe
n
eWton
to
tHe
p
oWeR
s
Upply
•
Connect the power cable between the Camera power input (rear panel) and the PSU.
•
Connect the power cable of the PSU to the mains supply.
3.7 c
onnecting
a
c
ooling
s
ysteM
The Newton can use either air cooling, or optional liquid cooling for deep cooling.
3.7.1 i
MpoRtant
c
onsideRations
WHen
Using
l
iqUid
c
ooling
s
ysteMs
• Before attempting to remove the coolant hoses, ensure that all coolant has been drained from the hoses and
integral coolant channel within the camera head.
• Care must be taken to avoid permanent damage to the camera system resulting from either leakage of
coolant during connection/removal of hoses or spillage of any residual coolant contained within the camera
head once the hoses have been removed.
• Always ensure that the temperature of the liquid coolant circulated through the camera head is above the
d
ew point of the camera ambient temperature and humidity conditions. Refer to the Dew Point graph in
Appendix C for guidance.
•
Use of coolant at or below the dew point can result in permanent damage to the camera head, due to
formation of condensation on internal components.
• When using a liquid cooling system it is recommended that an overpressure device is fitted to avoid
leaks that may find their way to the mains electricity supply and create a hazard.
•
Never use damaged, split or worn hoses.
• The coolant connectors are factory fitted and sealed, they cannot be removed.
• Always remove residual coolant from the camera head if the camera is to be stored after operation-
especially if the storage conditions are below the freezing point of the coolant.
3.7.2 c
ooling
H
ose
c
onnectoRs
There are two cooling hose connectors that are located on the top plate of the camera for connection of the Newton to
a water cooler, or re-circulator system. One end is inserted into the camera body, the other end is barbed for push-fit
connection to suitable tubing.
• Coolant Hose Connectors:
Two push-fit hose connectors are factory fitted and sealed, suitable for connection
to 6 mm [0.25”] internal diameter soft PVC tubing / hose
• Recommended tubing
: 10 mm [0.4”] outside diameter, i.e. a wall thickness of 2 mm [0.08”]. Alternative hose
dimensions and materials should be thoroughly tested to ensure a leak tight seal is achieved with the barbed
connectors.
PSU
10 mm
[
0.4”
] Ø OD
2 mm [
0.08”
]
6 mm
[
0.25”
] Ø ID
Coolant Hose Connector
Coolant Tubing
Mains Supply
Newton
6 mm
[
0.25”
] Ø OD
Barbed end for
connection to
Coolant Tubing
Threaded end
is factory fitted
and sealed