The example shown below will configure Channel #2 Low Setpoint Relay to 12 K. Begin by pressing MENU
MENU
MENU
MENU
to navigate to the “Select Diode” display shown below.
The first menu is the diode selection
menu. Press
MENU
button
to move to
the “Channel #1 Low Temperature
Setpoint.
Press
MENU
button
to move to the
“Channel #1 High Temperature
Setpoint.
Press
MENU
button
to move to the
“Channel #1 SPARE Temperature
Setpoint.
Press
MENU
button
to move to the
“Channel #2 Low Temperature Setpoint.
Note that the cursor is on the
HUNDREDS digit. Press
ENTER
to
move to the TENS digit.
The cursor is now on the TENS digit, so
press
ENTER
again to move to the ONES
digit.
Once the cursor is on the ONES digit,
use the
UP
/
DOWN
keys to increment
the value under the desired value of
12K is reached.
After the value has been set, either
press
MENU
until the Main Display is
reached, or let the display timeout after
10 seconds.
Set SPARE #1
2 7 0 K
Set LO Temp #2
1 5 K
Set HI Temp #1
2 0 K
Set LO Temp #1
1 5 K
Select Diode
ASC Diode
Set LO Temp #2
1 5
Set LO Temp #2
1 5 K
Set LO Temp #2
1 2 K
Press
MENU
button
to move to the
“Channel #2 Low Temperature Setpoint.