8.Communication with Android Device via Wi-Fi (Optional)

 

 

(10) 

Set 

IP

 and 

Port

 to the same value as the setting in the oscilloscope. Click 

Confirm

 

to connect. Click 

History

 to recall the history settings.   

 

Connect with APP as Wi-Fi Station 

(1) 

In the oscilloscope, push  the 

Utility

  button.  Select 

Function

  in the bottom menu. 

Select 

LAN Set

  in  the left menu. In the bottom menu, set the 

Type

  item

 

as 

WIFI-STA

, and select 

Set 

in the bottom menu.   

(2) 

In the right menu, select 

SSID

, a keyboard will pop up.  Input the name of the Wi-Fi 

hotspot.  You can turn and push the 

M

  knob, or touch the keyboard (if the LCD is 

touchscreen).   

(3) 

In the right menu, select 

Encryption

,  push to set the Encryption  type  of the Wi-Fi 

hotspot. 

Open

  type needs no password. 

WEP

  or 

WPA

  type need to input  the 

password. In the right menu, select 

Password

, use the keyboard to input  the  Wi-Fi 

hotspot password.   

(4) 

In the right menu, select 

IP:PORT

, Turn the 

M

 knob to set the 

PORT

.   

(5) 

After setting, select 

ON

 in the 

Status

 menu. The value of 

IP

 is displayed in 

IP:PORT

 

menu. A changing icon 

  appears on the top right of the screen.   

   

 

 

124 

Summary of Contents for XDS3000 Series

Page 1: ...XDS3000 Dual Channel Series Digital Storage Oscilloscopes User Manual www owon com cn ...

Page 2: ...lace all materials published The information in this manual was correct at the time of printing However LILLIPUT will continue to improve products and reserves the rights to change specification at any time without notice is the registered trademark of the LILLIPUT Company Fujian LILLIPUT Optoelectronics Technology Co Ltd No 19 Heming Road Lantian Industrial Zone Zhangzhou 363005 P R China Tel 86 ...

Page 3: ...d This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care OWON shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than OWON representatives to install repair or service the product b to repair damage resulting from improper use or connection to incomp...

Page 4: ...cal System 14 Introduction to the Horizontal System 15 Introduction to the Trigger System 16 Touchscreen Controls Touchscreen is optional 17 Operate the Menu through Touchscreen 17 Gestures in Normal Mode 18 Gestures in Wave Zoom Mode 21 Other Operations Using Touchscreen 23 Magnifier Function Only for XDS3102AP XDS3202A with touchscreen 24 4 Advanced User Guidebook 26 How to Set the Vertical Syst...

Page 5: ...ow to Print the Screen Image 97 5 Use theArbitrary Function Generator Optional 98 Output Connection 98 To Set Channels 98 To Set Signals 99 To Output Sine Signals 99 To Set the Frequency 99 To Set the Period 100 To Set the Start Phase 100 To Set the Amplitude 100 To Set the Offset 101 To Set the High Level 101 To Set the Low Level 101 To Output Square Signals 101 To Output Ramp Signals 101 To Set ...

Page 6: ...Connect with PC Optional 118 Connect with PC as Wi Fi Access Point 118 Connect with PC as Wi Fi Station 119 8 CommunicationwithAndroidDeviceviaWi Fi Optional 122 How to Connect 122 Connect with APP as Wi Fi Access Point 122 Connect with APP as Wi Fi Station 124 User Interface 126 Gestures Control 128 9 Demonstration 130 Example 1 Measurement a Simple Signal 130 Example 2 Gain of a Amplifier in a M...

Page 7: ...Multimeter Optional 146 General Technical Specifications 147 12 Appendix 148 Appendix A Enclosure 148 Appendix B General Care and Cleaning 148 Appendix C Battery Using Guide 149 iv ...

Page 8: ...tput terminal When powered by AC power it is not allowed to measure AC power source directly because the testing ground and power cord ground conductor are connected together otherwise it will cause short circuit When powered by battery the product must ground connection To avoid electric shock there must be a ground wire connect between ground and the ground port on the back of product panel Chec...

Page 9: ...ccessible potentially Caution It indicates a potential damage to the instrument or other property might occur Safety Symbols Symbols on the product The following symbol may appear on the product Hazardous Voltage Refer to Manual Protective Earth Terminal Chassis Ground Test Ground To avoid body damage and prevent product and connected equipment damage carefully read the following safety informatio...

Page 10: ...the ports Warning To avoid fire or electrical shock when the oscilloscope input signal connected is more than 42V peak 30Vrms or on circuits of more than 4800VA please take note of below items Only use accessory insulated voltage probes and test lead Check the accessories such as probe before use and replace it if there are any damages Remove probes test leads and other accessories immediately aft...

Page 11: ...ace How to implement the general inspection How to implement the function inspection How to make a probe compensation How to set the probe attenuation coefficient How to use the probe safely How to implement an auto calibration Introduction to the vertical system Introduction to the horizontal system Introduction to the trigger system Touchscreen Controls Touchscreen is optional 4 ...

Page 12: ... you can set the different options for the current menu The other buttons are function buttons through which you can enter different function menus or obtain a specific function application directly 8 2 5 3 7 4 6 1 Figure 3 1 Front panel 1 Display area 2 Control button and knob area 3 Probe Compensation Measurement signal 5V 1kHz output 4 EXT Trigger Input 5 Signal Input Channel 6 Copy button You ...

Page 13: ...Menu Buttons Select the right menu item Select the bottom menu item Remove the left and right menu Figure 3 2 Menu Buttons Rear Panel 3 10 9 8 7 13 12 11 1 2 6 4 5 Figure 3 3 Rear Panel 1 Handle 2 Air vents 3 Input terminals of multimeter optional 4 AC power input jack 5 Fuse 6 Foot stool Adjust the tilt angle of the oscilloscope 7 VGA port To connect the oscilloscope with a monitor or a projector...

Page 14: ...rator The output type can be set on the menu Utility menu Output Output 13 Out 1 port Output single channel or CH1 Output dual channel of optional waveform generator Control Area 8 7 6 1 2 3 11 10 5 4 9 Figure 3 4 Control Area Overview 1 Function button area Total 11 buttons 2 Waveform generator controls optional or DAQ Multimeter Recorder see Multimeter Recorder on P111 P F Pass Fail see Pass Fai...

Page 15: ...l function 9 DMM optional Multimeter or Snapshot Shortcut button for measurement snapshot 10 Direction key Move the cursor of the focused parameter 11 M knob Multipurpose knob when a symbol appears on the menu it indicates you can turn the M knob to select the menu or set the value You can push it to close the menu on the left and right User Interface Introduction 1 3 10 21 16 29 26 8 9 11 13 18 1...

Page 16: ... The pointer shows the trigger level position 17 The waveform of CH1 18 The two blue dotted lines indicate the horizontal position of cursor measurement 19 The waveform of CH2 20 It indicates the measured type and value of the corresponding channel T means period F means frequency V means the average value Vp the peak peak value Vr the root mean square value Ma the maximum amplitude value Mi the m...

Page 17: ...e caused by transportation If it is found that the packaging carton or the foamed plastic protection cushion has suffered serious damage do not throw it away first till the complete device and its accessories succeed in the electrical and mechanical property tests 2 Check the Accessories The supplied accessories have been already described in the Appendix A Enclosure of this Manual You can check w...

Page 18: ...select Default in the bottom menu The default attenuation coefficient set value of the probe in the menu is 10X 2 Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel Align the slot in the probe with the plug in the CH1 connector BNC and then tighten the probe with rotating it to the right side Connect the probe tip and the ground clamp to the connector of ...

Page 19: ... with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector and then push the Autoset button on the front panel 2 Check the displayed waveforms and regulate the probe till a correct compensation is achieved see Figure 3 7 and Figure 3 8 Overcompensated Compensated correctly Under compensated Figure 3 7 Displayed Wav...

Page 20: ... 5MHz To use the full bandwidth of the oscilloscope the switch must be set to 10X Identify the Probe Attenuation Coefficient Automatically The oscilloscope can identify the probe attenuation coefficient of the 100 1 impedance 5K 20 or 10 1 impedance 10K 20 probe with the identifying pin When you attach the probe the oscilloscope set the attenuation automatically on the oscilloscope vertical menu f...

Page 21: ...me This program must be executed whenever the change of ambient temperature is 5 or over Before performing a self calibration disconnect all probes or wires from the input connector Push the Utility button select Function in the bottom menu select Adjust in the left menu select Self Cal in the bottom menu run the program after everything is ready Introduction to the Vertical System As shown in Fig...

Page 22: ... the trace position is far out of the screen and want it to get back to the screen center immediately 2 Change the Vertical Setting and Observe the Consequent State Information Change With the information displayed in the status bar at the bottom of the waveform window you can determine any changes in the channel vertical scale factor Turn the Vertical Scale knob and change the Vertical Scale Fact...

Page 23: ...et the triggering displacement back to 0 as a shortcut key 3 Push the Horizontal HOR button to switch between the normal mode and the wave zoom mode Introduction to the Trigger System As shown in Figure 3 13 there are one knob and three buttons make up Trigger Controls The following practices will direct you to be familiar with the setting of the trigger system gradually Figure 3 13 Trigger Contro...

Page 24: ...Touchscreen Select a menu item Touch the menu items in the bottom menu or in the right menu or in the left menu Switch menu items If there are options that can be switched in the menu you can repeatedly touch the area of the menu item to switch or push the corresponding button to switch See figure below Press repeatedly to switch the options Scroll the list If there is a scroll bar in the left men...

Page 25: ...ition knob Swipe up or down on the screen Touch to select the channel Selected Unselected Control the vertical position of the selected channel Set the trigger level of the source in trigger menu Trigger Level knob Swipe up or down on the area nearby the trigger pointer on the right Control the trigger level of the source in trigger menu Set the horizontal position Horizontal Position knob Swipe l...

Page 26: ...tal position Double Zoom and Single Zoom In touchable menu pane if DoubleZoom is selected in the display area pinch and spread horizontally to change the time base pinch and spread vertically to change the voltage division of current channel 19 ...

Page 27: ...click anywhere on the Display area to show the touching control panel Click to increase the voltage division of CH1 Click to decrease the voltage division of CH1 Click to decrease the time base Click to increase the time base Click to increase the voltage division of CH2 Click to decrease the voltage division of CH2 Set the voltage division Vertical Scale knob Click on the left upper area to incre...

Page 28: ...tal Scale knob Click on the left area to increase the voltage division click on the right area to decrease the voltage division Gestures in Wave Zoom Mode Push the Horizontal HOR button to enter into zoom mode The top half of the display shows the Main window and the bottom half displays the Zoom window The Zoom window is a magnified portion of the Main window Control the vertical position of the ...

Page 29: ...uidebook Control the horizontal position of Zoom Window Pinch and spread vertically to change the voltage division of current channel Pinch and spread horizontally to change the time base of Zoom Window Double Zoom 22 ...

Page 30: ...perations Using Touchscreen Measure with Cursors Press nearby any cursor line until both lines are selected drag them up and down simultaneously Switch horizontal or vertical lines If vertical lines are selected drag left and right Press nearby a cursor line to make it selected drag it up and down Selected Unselected Run Stop Double tap on the display area or click the or on the left top of the di...

Page 31: ...agnifier function is enabled Click the icon on the left top of the display area the magnifier window and menu will be shown Click the icon to turn off the magnifier function Voltage division and time base of the magnified wave Magnifier switch icon Wave selection Drag it to move You can move this window by dragging Long press anywhere the selected wave region can be move to here quickly The descri...

Page 32: ...alue the shorter the vertical height of the wave selection Transparent ON OFF Make the magnifier window transparent or not The Cursor Measurement in Magnifier Window When the magnifier function is enabled push Cursor panel key to display the cursor menu In the bottom menu select Window as Main or Extension to make the cursors shown in the main window or the magnifier window Note The magnifier func...

Page 33: ...This chapter will deal with the following topics mainly How to Set the Vertical System How to Set the Horizontal System How to Set the Trigger Decoding System How to Implement the Sampling Setup How to Set the Display System How to Save and Recall Waveform How to Record Playback Waveforms How to Clone and Recall a waveform How to Implement the Auxiliary System Function Setting How to Update your I...

Page 34: ...ption Coupling DC AC GROUND Pass both AC and DC components of the input signal Block the DC component of the input signal Disconnect the input signal Inverted ON OFF Display inverted waveform Display original waveform Probe Attenu 0 001X to 1000X Step by 1 2 5 Match this to the probe attenuation factor to have an accurate reading of vertical scale MeasCurr YES NO If you are measuring current by pr...

Page 35: ...1 as an example if you are measuring current by probing the voltage drop across a 1Ω resistor the operation steps are shown as follows 1 Push the CH1 button to show CH1 SETUP menu 2 Select Probe in the bottom menu In the right menu set MeasCurr as YES the A V radio menu will appear below Select it turn the M knob to set the Amps Volts ratio Amps Volts ratio 1 Resistor value Here the A V radio shou...

Page 36: ...utput impedance The max input voltage cannot exceed 5 Vrms Use Mathematical Manipulation Function The Mathematical Manipulation function is used to show the results of the addition multiplication division and subtraction operations between two channels the FFT operation for a channel advanced math feature including Intg Diff Sqrt user defined function and digital filter Press the Math button to di...

Page 37: ...fined function DIR channel CH1 CH2 Select channel type low pass Only the signals whose frequencies are lower than the current cut off frequency can pass the filter high pass Only the signals whose frequencies are greater than the current cutoff frequency can pass the filter band pass Only the signals whose frequencies are greater than the cutoff frequency down and lower than the current cutoff fre...

Page 38: ...m menu 3 In the right menu select Factor1 as CH1 4 Select Sign as in the right menu 5 In the right menu select Factor2 as CH2 6 Select Vertical div in the right menu turn the M knob to adjust the vertical position of Math waveform 7 Select Vertical V div in the right menu turn the M knob to adjust the vertical division of Math waveform User defined function 1 Press the Math button to display the m...

Page 39: ...input signal on Oscilloscope You can match these frequencies with known system frequencies such as system clocks oscillators or power supplies FFT function in this oscilloscope transforms 8192 data points of the time domain signal into its frequency components mathematically the record length should be 10K or above The final frequency contains 4096 points ranging from 0Hz to Nyquist frequency Taki...

Page 40: ...event are significantly different Rectangle Best solution for frequency worst for magnitude Best type for measuring the frequency spectrum of nonrepetitive signals and measuring frequency components near DC Recommend to use for Transients or bursts the signal level before and after the event are nearly equal Equal amplitude sine waves with frequencies those are very close Broadband random noise wi...

Page 41: ... amplitudes the side lobe level and shape factor are closest to the traditional Gaussian RBW This window is also good for random signals Bartlett The Bartlett window is a slightly narrower variant of the triangular window with zero weight at both ends Figure 4 1 Figure 4 2 Figure 4 3 Figure 4 4 Figure 4 5 Figure 4 6 are examples for measuring sine wave with a frequency of 1kHz under the selection ...

Page 42: ...4 Advanced User Guidebook Figure 4 2 Rectangle window Figure 4 3 Blackman window 35 ...

Page 43: ...4 Advanced User Guidebook Figure 4 4 Hanning window Figure 4 5 Kaiser window 36 ...

Page 44: ...ngle shot events set the oscilloscope acquisition mode to average What is Nyquist frequency The Nyquist frequency is the highest frequency that any real time digitizing oscilloscope can acquire without aliasing This frequency is half of the sample rate Frequencies above the Nyquist frequency will be under sampled which causes aliasing So pay more attention to the relation between the frequency bei...

Page 45: ...al Position knob this knob is used to adjust the horizontal positions of all channels include those obtained from the mathematical manipulation the analytic resolution of which changes with the time base 2 Horizontal Scale knob it is used to set the horizontal scale factor for setting the main time base or the window 3 Horizontal HOR button push it to switch between the normal mode and the wave zo...

Page 46: ...le display to meaningful waveform When DSO starts to acquire data it will collect enough data to draw waveform on left of trigger point DSO continues to acquire data while waiting for trigger condition to occur Once it detects a trigger it will acquire enough data continuously to draw the waveform on right of trigger point Trigger control area consists of 1 knob and 2 menu buttons Trigger Level Th...

Page 47: ...se trigger runt trigger windows trigger timeout trigger and Nth edge trigger Edge Trigger It occurs when the trigger input passes through a specified voltage level with the specified slope Video Trigger Trigger on fields or lines for standard video signal Slope Trigger The oscilloscope begins to trigger according to the signal rising or falling speed Pulse Trigger Find pulses with certain widths R...

Page 48: ...ponent Allow all component pass Block the high frequency signal only low frequency component pass Block the low frequency signal only high frequency component pass LF menu is only for certain models Slope Trigger on rising edge Trigger on falling edge Mode Holdoff Auto Normal Single Holdoff Reset Acquire waveform even no trigger occurs Acquire waveform when trigger occurs When trigger occurs acqui...

Page 49: ...ne Synchronic trigger in video field Synchronic trigger in video odd filed Synchronic trigger in video even field Synchronic trigger in designed video line turn the M knob or click to set the line number Mode Holdoff Auto Acquire waveform even no trigger occurred 3 Slope Trigger Slope trigger sets the oscilloscope as the positive negative slope trigger within the specified time In Slope Trigger mo...

Page 50: ...l before another trigger occur press panel button or click to move cursor to choose which digit to be set Set Holdoff time as 100 ns 4 Pulse Width Trigger Pulse trigger occurs according to the width of pulse The abnormal signals can be detected through setting up the pulse width condition In Pulse Width Trigger mode the trigger setting information is displayed on bottom right of the screen for exa...

Page 51: ... click to set time interval before another trigger occur press panel button or click to move cursor to choose which digit to be set Set Holdoff time as 100 ns 5 Runt Trigger Trigger pulses that pass through one trigger level but fail to pass through the other trigger level Shown as below figure In Runt Trigger mode the trigger setting information is displayed on bottom right of the screen for exam...

Page 52: ...rigger when runt pulse is greater than the set pulse width Trigger when runt pulse equals to the set pulse width Trigger when runt pulse is lower than the set pulse width Mode Holdoff Auto Normal Single Holdoff Reset Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform then stop 100 ns 10 s adjust M knob or click to set time inte...

Page 53: ...ob or click to set the up level threshold Adjust the M knob or click to set the low level threshold Condition Polarity Positive Polarity the oscilloscope triggers on the positive Windows pulse Negative Polarity the oscilloscope triggers on the negative Windows pulse Enter Triggers when the trigger signal enters the specified trigger level range Exit Triggers when the trigger signal exits the speci...

Page 54: ...eater than the timeout time set In Timeout Trigger mode the trigger setting information is displayed on bottom right of the screen for example indicates that trigger type is Timeout trigger source is CH1 edge is positive 0 00mV is up level or low level threshold Timeout Trigger menu list MENU SETTING INSTRUCTION Single Mode Timeout Set vertical channel trigger type as Timeout trigger Source CH1 CH...

Page 55: ...ppears on the specified idle time As figure shown below the oscilloscope should trigger on the second falling edge after the specified idle time and the idle time should be set to P1 P2 P3 P4 Idle Time M Wherein M P1 P2 P3 and P4 are positive or negative pulse width participating in the counting In Nth Edge Trigger mode the trigger setting information is displayed on bottom right of the screen for...

Page 56: ...n no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform then stop 100 ns 10 s adjust M knob or click to set time interval before another trigger occur press panel button or click move cursor to choose which digit to be set Set Holdoff time as 100 ns Alternate Trigger Trigger mode Edge Trigger signal comes from two vertical channels when alternate trigg...

Page 57: ...ON Mode Logic Set vertical channel trigger type as Logic trigger Logic Mode AND OR XNOR XOR Set logic mode as AND Set logic mode as OR Set logic mode as XNOR Set logic mode as XOR Input Mode CH1 CH2 Set CH1 as High Level Low level high or low level Rise and Fall Set CH2 as High Level Low level high or low level Rise and Fall Note When input mode of one channel is set as Rise or Fall the other chan...

Page 58: ...data which consist of 1bit start bit 5 8bits data bits 1bit check bit and 1 2 stop bits In RS232 bus trigger mode the trigger setting information is displayed on bottom right of the screen for example indicates that trigger type is RS232 CH1 trigger level is 0 00mV Format as shown in the figure below RS232 Trigger RS232 Trigger menu list MENU SETTING INSTRUCTION Bus Type RS232 Set vertical channel...

Page 59: ...d Common Baud adjust M knob to choose common baud Custom Baud adjust M knob to choose baud ranges from 50 to 10 000 000 Data Data Bits Set as 5 6 7 8 bits Data Set data according to data bits ranges from 0 31 0 63 0 127 or 0 255 Mode Holdoff Auto Normal Single Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform then stop 2 I2C T...

Page 60: ...he read or write bit when the preset address is met Adr For mat Addr Bits Set Address Bits to be 7 8 or 10 Set address according to the preset address bits address range is 0 127 0 255 0 1023 respectively Set Data Direction to be Read or Write Note The set is not available when Address bits is set to 8 Addr ess Dire ction Data Search for the preset data value on SDA and trigger on the dump edge of...

Page 61: ...nel bus type as SPI trigger Source CH1 CH2 Set CH1 as SCL or SDA Set CH2 as SCL or SDA Time Out Time out Set the minimum time that SCL must be idle that is a period of SCL available range 100ns 10s Time out means SCL keeps idle for a specified time before oscilloscope starts to search for the data SDA on which to trigger adjust M knob or click to set time out press panel button or click move curso...

Page 62: ...etting information is displayed on bottom right of the screen for example indicates that trigger type is CAN CH1 trigger level is 126 mV CAN Trigger menu list MENU SETTING INSTRUCTION Bus Type CAN Set vertical channel bus type as CAN trigger Input Source CH1 CH2 Select CH1 as the trigger source Select CH2 as the trigger source Type CAN_H CAN_L TX RX Actual CAN_H bus signal Actual CAN_L bus signal ...

Page 63: ...he M knob and Direction key on the front panel ID Data Configure Bottom menu Format Select Standard or Extended ID Use the M knob and Direction key on the front panel to set Byte Length Set the number of bytes with the M knob The range is 1 to 8 Data Set the data with the M knob and Direction key on the front panel End Trigger on the end frame of the data frame Missing Ack Trigger on Missing Ack B...

Page 64: ...heck bit error is detected two red error marks will be displayed in the corresponding position in the waveform RS232 Decoding menu list MENU SETTING INSTRUCTION Bus Type RS232 Set bus type of decoding as RS232 Configure Common Baud Turn the M knob to select from the Baud list on the left Custom Baud Turn the M knob or tap on in touchscreen to set the Baud The range is 50 to 10 000 000 Tip You can ...

Page 65: ...igger menu and decoding menu you can set anyone of them the other will be changed synchronously Decoded information interpretation Information Abbreviation Background Read Address R Read or do not display Green Write Address W Write or do not display Green Data D Data or do not display Black Note Use the Trigger Level knob to adjust the thresholds of bus trigger and bus decoding When the ACK ACKno...

Page 66: ...igger Level knob to adjust the thresholds of bus trigger and bus decoding LS First in Bit Order menu item Least Significant Bit First means that the least significant bit will arrive first hence e g the hexadecimal number 0x12 will arrive as the sequence 01001000 in binary representation will be decoded as the reversed sequence 00010010 SPI Decoding menu list MENU SETTING INSTRUCTION Bus Type SPI ...

Page 67: ...by the signal will be displayed Tip If there are repetitive menu items in both trigger menu and decoding menu you can set anyone of them the other will be changed synchronously Decoded information interpretation Information Abbreviation Background Identifier I ID or do not display Green Overload Frame OF Green Error Frame EF Green Data Length code L DLC or do not display Blue Data D Data or do not...

Page 68: ...he Acquire button Acqu Mode Length PERF Mode only for XDS3000 A series and Intrpl is shown in the bottom menu The description of the Acqu Mode menu is shown as follows Function Menu Setting Description Acqu Mode Sample Normal sampling mode Peak Detect Use to capture maximal and minimal samples Finding highest and lowest points over adjacent intervals It is used for the detection of the jamming bur...

Page 69: ...be set to 8 bit or 12 bit in this menu Note For XDS3102AP XDS3202A the selection of 8 bit 12 bit or 14 bit is not affected by sample rate The description of the Intrpl menu is shown as follows Function Menu Setting Description Intrpl Sinx x Use sine x x interpolation x Use linear interpolation Interpolation method is a processing method to connect the sampled points using some points to calculate ...

Page 70: ...4 Advanced User Guidebook Sine x x interpolation Linear interpolation How to Set the Display System Push the Display button and the Display menu is shown as follows 63 ...

Page 71: ...he oscilloscope will start to accumulate acquisitions again Color Only for certain models Color temperature function uses color grading to indicate frequency of occurrence The hot colors like red yellow indicate frequently occurring events and the colder colors like blue green indicate rarely occurring events Function Menu Setting Description Type Dots Vect Only the sampling points are displayed T...

Page 72: ...nnel 1 is displayed in the horizontal axis and Channel 2 in the vertical axis the oscilloscope is set in the un triggered sample mode the data are displayed as bright spots The operations of all control knobs are as follows The Vertical Scale and the Vertical Position knobs of Channel 1 are used to set the horizontal scale and position The Vertical Scale and the Vertical Position knobs of Channel ...

Page 73: ...ps 1 Push Trigger Menu button set the trigger type to Single set the trigger mode to Edge select the signal source 2 Push the Display button 3 Select Counter as ON or OFF in the bottom menu How to Save and Recall a Waveform Push the Save button you can save the waveforms configures screen images record or clone the waveform The description of the Save Function Menu is shown as the following table ...

Page 74: ...Close All Close all the waveforms stored in the object address Save Save the waveform of the source to the selected address Whatever the Type of save menu is set you can save the waveform by just pressing the Copy panel button in any user interface Select Type in the bottom menu in the right Format menu you can select the storage format Storage Internal External Save to internal storage or USB sto...

Page 75: ...1 as object address in the left menu 7 Select Save in the bottom menu to save the waveform 8 Recalling Select Object Show in the bottom menu select 1 in the left menu In the right menu select Show as ON the waveform stored in the address will be shown the address number and relevant information will be displayed at the top left of the screen In order to save the waveform of CH1 and CH2 into the US...

Page 76: ...led push the Save panel button the save menu is displayed at the bottom of the screen 3 Select Type in the bottom menu select Image in the left menu 4 Select Save in the bottom menu an input keyboard used to edit the file name will pop up The default name is current system date and time Select the key in the keyboard to confirm USB disk Requirements The supported format of the USB disk FAT32 file ...

Page 77: ...r 2 red mark area choose Format And system will pop up a warning message click Yes Figure 4 10 Format the USB disk warning 5 Set File System as FAT32 Allocation unit size 4096 Check Perform a quick format to execute a quick format Click OK and then click Yes on the warning message 70 ...

Page 78: ... with allocation unit size 4096 after formatting Use Minitool Partition Wizard to format Download URL http www partitionwizard com free partition manager html Tip There are many tools for the USB disk formatting on the market just take Minitool Partition Wizard for example here 1 Connect the USB disk to the computer 2 Open the software Minitool Partition Wizard 71 ...

Page 79: ...r push keyboard F5 and information about the USB disk will display on the right side with red mark 1 and 2 Figure 4 13 Reload Disk 4 Right click 1 or 2 red mark area choose Format Figure 4 14 Choose format 5 Set File System FAT32 Cluster size 4096 Click OK Figure 4 15 Format setting 72 ...

Page 80: ...igure 4 18 Format successfully How to Record Playback Waveforms Wave Record function can record the input current wave You can set the interval between recorded frames in the range of 10 ms 10 s The max frame number reaches 1000 and you can get better analysis effect with playback and storage function The storage medium contains two kinds Internal and External When the storage medium is Internal W...

Page 81: ...Channel 2 will be recorded If a Channel is turned off while recording the waveform of the channel is invalid in the playback mode Playback Play back the wave recorded or saved Playback menu shows as follows Menu Setting Instruction Playback Mode FrameSet Start frame End frame Cur frame Interval Turn the M knob to select the number of start frame to playback 1 1000 Turn the M knob to select the num...

Page 82: ...ode select Operate as Play 10 To save the wave recorded select Mode in the bottom menu select Storage in the right menu Select FrameSet in the bottom menu to set the range of frames to store select Save in the bottom menu 11 To load the waveform from the internal memory select Load in the bottom menu then enter the Playback of the Mode to analyze the wave Note When playbacking the waveform the sam...

Page 83: ...in the right menu 6 Select FrameSet in the bottom menu set End frame and Interval in the right menu If you want to record wave to external infinitely select Infinity in the right menu the End frame will display 7 In the bottom menu set Refresh 8 In the bottom menu select Operate as Play Connect external device to the computer and wave_record_0 bin is the recorded file Open the software and do as f...

Page 84: ...f CH1 and CH2 Line a b ab x Turn the M knob to move line a Turn the M knob to move line b Two cursors are linked Turn the M knob to move the pair of cursors Set the cursors to select the entire screen automatically Clone Generator is available Clone Clone the waveform between two cursors and output it through the generator AG Output Only for dual channel AG Out1 Out2 If the source is CH1 or CH2 yo...

Page 85: ...d set the record length to a smaller value 5 Select Save in the bottom menu To save the waveform to internal memory select Storage in the right menu as Internal Turn the M knob to select an object in the left menu select Save in the right menu To save the waveform onto a USB memory device insert a USB memory device into the port on the front panel If the icon appears on the top right of the screen...

Page 86: ...right menu To output the selected waveform through the generator directly The generator is optional 1 Push Save button 2 Select Type in the bottom menu in the left menu turn the M knob to select Clone 3 Select Source in the bottom menu select one channel CH1 CH2 or two channels CH1 CH2 in the right menu for Source 4 Select Line in the bottom menu Select the cursor and move it to select the desired...

Page 87: ...eaning Value Comment VOLT Voltage division divided by 400 is ADC resolution When the source is CH1 CH2 it is CH1 voltage division 4 bytes float The value indicates voltage the unit is mV such as 200 mV 6 SAMP Parameter name Meaning Value Comment SAMP Sample rate 4 bytes float The unit is Sa s 7 ADCB Parameter name Meaning Value Comment ADCB ADC bit ADC resolution 4 bytes int 8 bit or 12 bit 8 CHAN...

Page 88: ... menu select Display in the left menu The description of Display Menu is shown as the follows Adjust Function Menu Setting Description Language Choose the display language of the operating system Set Time Display ON OFF On Off the date display Hour Min Setting Hour Minute Day Month Setting Date Month Year Setting Year KeyLock Lock all keys Unlock method push Trigger Menu button in trigger control ...

Page 89: ...lay at the left select Adjust If everything is ready select Self Cal in the bottom menu to enter the self calibration procedure of the instrument Probe checking To check whether probe attenuation is good The results contain three circumstances Overflow compensation Good compensation Inadequate compensation According to the checking result users can adjust probe attenuation to the best Operation st...

Page 90: ...he bottom menu select Operate in the right menu as Start the test will begin 6 Save rule Select SaveRule in the bottom menu Select the save location in the left menu and then select Save in the right menu to save the rules which could be called up at once when need Select Load to call up the rule saved Note 1 When Pass Fail is ON if XY or FFT is ready to run then Pass Fail will be closed under the...

Page 91: ...utput refer to How to Print the Screen Image on page 97 LAN Set Using the LAN port or Wi Fi the oscilloscope can be connected with a computer The oscilloscope also supports communications with Android based smart device via Wi Fi Refer to Communication with PC on page 114 and Communication with Android Device via Wi Fi Optional on page 122 for the operation steps Update Use the front panel USB por...

Page 92: ...rument do not power off the instrument or remove the USB memory device during the update process To update your instrument firmware do the following 1 Push the Utility button select Function in the bottom menu select Configure in the left menu select About in the bottom menu View the model and the currently installed firmware version 2 From a PC visit www owon com cn and check if the website offer...

Page 93: ... on the bottom left of the screen The oscilloscopes provide 30 parameters for auto measurement including Period Frequency Mean PK PK RMS Max Min Top Base Amplitude Overshoot Preshoot Rise Time Fall Time PulseWidth PulseWidth Duty Cycle Duty Cycle Delay A B Delay A B Cycle RMS Cursor RMS Screen Duty Phase PulseCount PulseCount RiseEdgeCnt FallEdgeCnt Area and Cycle Area The Automatic Measurements m...

Page 94: ...1 On the saved waveform 2 On the Dual Wfm Math waveform 3 On the Video trigger mode On the Scan format period and frequency can not be measured Measure the period the frequency of the CH1 following the steps below 1 Push the Measure button to show the automatic measurement function menu 2 Select Add in the bottom menu 3 In the right menu select CH1 in the Source menu item 4 In the left Type menu t...

Page 95: ...The minimum amplitude The most negative peak voltage measured over the entire waveform Vtop Voltage of the waveform s flat top useful for square pulse waveforms Vbase Voltage of the waveform s flat base useful for square pulse waveforms Vamp Voltage between Vtop and Vbase of a waveform OverShoot Defined as Vmax Vtop Vamp useful for square and pulse waveforms PreShoot Defined as Vmin Vbase Vamp use...

Page 96: ...y Cycle defined as Width Period Duty Duty Cycle defined as Width Period Delay A B The delay between the two channels at the rising edge Delay A B The delay between the two channels at the falling edge Screen Duty Defines as the width of the positive pulse Entire period Phase Compare the rising edge of CH1 and CH2 calculate phase difference of two channels Phase difference Delay between channels at...

Page 97: ...measured is 0 How to Measure with Cursors Push the Cursor button to turn cursors on and display the cursor menu Push it again to turn cursors off The Cursor Measurement for normal mode The description of the cursor menu is shown as the following table Function Menu Setting Description Type Voltage Time Time Voltage AutoCursr Display the voltage measurement cursor and menu Display the time measurem...

Page 98: ...tom menu select Line Type as Time to make the vertical cursors active If the Line in the bottom menu is select as a turn the M knob to move line a to the right or left If b is selected turn the M knob to move line b 5 In the bottom menu select Line Type as Voltage to make the horizontal cursors active Select Line in the bottom menu as a or b turn the M knob to move it 6 Push the Horizontal HOR but...

Page 99: ...easurement cursor and menu Freq Display the Freq measurement cursor and menu Freq Vamp or Freq Phase Display the corresponding measurement cursor and menu AutoCursr The horizontal cursors are set as the intersections of the vertical cursors and the waveform Line Type Freq Vamp or Freq Phase type Freq Makes the vertical cursors active Vamp or Phase Makes the horizontal cursors active Window Wave zo...

Page 100: ... as Freq to make the vertical cursors active If the Line in the bottom menu is select as a turn the M knob to move line a to the right or left If b is selected turn the M knob to move line b 6 In the bottom menu select Line Type as Vamp to make the horizontal cursors active Select Line in the bottom menu as a or b turn the M knob to move it 7 In the bottom cursor menu you can select Window as Main...

Page 101: ...o Autoscale function and the symbol AE A will be flickering on the top left of the screen 2 In the mode of Autoscale the oscilloscope can self estimate Trigger Mode Edge Video At this point the trigger menu is not available 3 At the mode of XY and STOP status when entering into Autoscale DSO switches to YT mode and AUTO triggering 4 At the mode of Autoscale DSO is always set as DC coupling with AU...

Page 102: ...ow to Use Executive Buttons Executive Buttons include Autoset Run Stop Single Copy Autoset It s a very useful and quick way to apply a set of pre set functions to the incoming signal and display the best possible viewing waveform of the signal and also works out some measurements for user as well The details of functions applied to the signal when using Autoset are shown as the following table Fun...

Page 103: ...he upper menu and waveform information Note The Autoset function requires that the frequency of signal should be no lower than 20Hz and the amplitude should be no less than 5mv Otherwise the Autoset function may be invalid Run Stop Enable or disable sampling on input signals Notice When there is no sampling at STOP state the vertical division and the horizontal time base of the waveform still can ...

Page 104: ... on the rear panel of the oscilloscope Note The USB Device port supports PictBridge compatible printers 2 Push the Utility button select Function in the bottom menu select Output in the left menu 3 In the bottom menu select Device as PICT When PC is selected you can get an image by Oscilloscope software 4 In the bottom menu select Print Setup In the right menu set up print parameters The On select...

Page 105: ...ct Output in the right menu select AG Output Single channel Connect the BNC cable to the port marked Out in the back of the oscilloscope Dual channel Connect the BNC cable to the port marked Out 1or Out 2 in the back of the oscilloscope Out 1 is the output of CH1 Out 2 is the output of CH2 also can be used as the port of trigger signal output Pass Fail output Figure 5 1 Generator Output Ports Dual...

Page 106: ... menu as On the amplitude of the two channels can be adjusted synchronously Align Phase Select Align Phase in the bottom menu to align the initial phase of two channel signals To Set Signals 1 Push single channel or dual channel button to show the bottom menu of generator 2 Select the desired waveform in the bottom menu the corresponding menu is displayed on the right 3 The parameters can be set i...

Page 107: ...k to increase the value of cursor position Move the cursor Click to decrease the value of cursor position To Set the Period Select Period in the right menu if Period is not displayed select Frequency and select it again to switch to Period Set the parameter in the right menu To Set the Start Phase Select StartPhase in the right menu Set the parameter in the right menu To Set the Amplitude Select A...

Page 108: ... of Ramp waveform are Frequency Period Start Phase Amplitude High Level Offset Low Level Symmetry To set the Frequency Period Start Phase Amplitude High Level Offset Low Level please refer to To Output Sine Signals on page 99 To Set the Symmetry of Ramp Select Symmetry in the right menu of Ramp Set the parameter in the right menu To Output Pulse Signals The parameters of Pulse waveform are Frequen...

Page 109: ...000000 1000000000 The waveform points range is 2 8192 3 Set the interpolation Select Intrpl in the right menu choose between On Off If you choose On the points will be connected with beelines otherwise the voltages between two consecutive points will not change and the waveform looks like a step up one 4 Edit the waveform points Select Edit Points in the right menu Select Point input the number of...

Page 110: ...e To read a waveform stored in internal storage or USB device 1 Push single channel or dual channel button Select Arb in the bottom menu select Others in the right menu and select File Browse 2 Select the desired waveform file in internal storage FLASH or USB device USBDEVICE 3 Select Read in the right menu Built in Waveform There are 46 built in Arbitrary waveforms Steps for selecting the built i...

Page 111: ...r up and stair down waveform Trapezia Trapezoid waveform RoundHalf RoundHalf wave AbsSine Absolute value of a Sine AbsSineHalf Absolute value of half a Sine SineTra Sine transverse cut SineVer Sine vertical cut NegRamp Negative ramp AttALT Gain oscillation curve AmpALT Attenuation oscillation curve CPulse Coded pulse PPulse Positive pulse NPulse Negative pulse Maths ExpRise Exponential rise functi...

Page 112: ...ontrols the built in waveform generator to sweep a sine wave across a range of frequencies while measuring the input to and output from a device under test At each frequency gain and phase are measured and plotted on a frequency response Bode chart When the frequency response analysis completes you can move a marker across the chart to see the measured gain and phase values at each frequency point...

Page 113: ...knob to set waveform generator amplitude Check to enable fine tuning Note The amplitude range is 2 mVpp to 6 Vpp Chart Gain Scale Adjust scale of gain plot range from 5dB to 500dB Gain Offset Adjust offset of gain plot range from 250dB to 250dB Phase Scale Adjust scale of phase plot range from 5 to 180 Phase Offset Adjust offset of phase plot range from 180 to 180 Autoscale Autoscale gain and phas...

Page 114: ... the front panel to enter exit the multimeter function The button backlight will be lighted when the multimeter function is enabled Figure 6 2 Multimeter Button The bottom menu of multimeter is as below Menu Setting Description Current ACA DCA Measuring AC current Measuring DC current Voltage ACV DCV Measuring AC voltage Measuring DC voltage R C Measuring resistance Testing diode Testing for conti...

Page 115: ... current range DMM Information Window The Multimeter Information Window is displayed on the top right of the screen 4 9 1 2 8 5 3 6 7 Figure 6 3 Multimeter Information Window Description 1 Manual Auto range indicators MANUAL refers to the measuring range in manual operation mode and AUTO refers to the measuring range in automatic operation mode 2 Measurement mode indicators A Current measurement V...

Page 116: ...sured circuit and discharge all high voltage capacitors Remove the test leads and restore the circuit to the original condition To measure a AC or DC current which is 400 mA to 10 A do the following 1 Push DMM button on the front panel Select Current in the bottom menu select it again to switch between ACA AC current or DCA DC current 2 Select Configure in the bottom menu select Current as 10A in ...

Page 117: ...ive terminal cathode The cathode of a diode is indicated with a band Read the diode forward bias Testing for Continuity 1 Push DMM button on the front panel Select in the bottom menu select it until switching to 2 Connect the black test lead to the COM terminal on the back of the oscilloscope and the red test lead to the V Ω C terminal 3 Probe the test points to measure the resistance in the circu...

Page 118: ...rmation Display Show hide the information window on the top right of the display 1 Select Configure in the bottom menu select Show Info in the right menu as ON The information window will be shown on the display 2 Select OFF to hide Auto or Manual Range Auto range is set as default To switch auto or manual range do the following steps 1 Select Configure in the bottom menu 2 Select Auto Range in th...

Page 119: ...ile will be named as Multimeter_Recorder csv If a file with the same name already exists in the USB memory device it will be overwritten If you want to keep the existing file back up it to other location in advance Select STRT in the bottom menu to start recording 9 When the record time gets to the defined duration the recording will be ended If you want to end recording prematurely select STOP in...

Page 120: ... the bottom menu to export How to chart the data You can open the CSV file using Microsoft Excel or your favorite spreadsheet application and create charts based on the data In the following steps Microsoft Excel 2010 is used as an example 1 Open Multimeter_Recorder csv in Excel Interval Measurement mode Reference value of the relative measurements 2 Select the data that you want to chart see Figu...

Page 121: ...on to choose from Using USB Port 1 Connection Use a USB data cable to connect the USB Device port in the right panel of the Oscilloscope to the USB port of a PC 2 Install the driver Run the Oscilloscope communication software on PC push F1 to open the help document Follow the steps of title I Device connection in the document to install the driver 3 Port setting of the software Run the Oscilloscop...

Page 122: ...e set the IP address to 192 168 1 71 Figure 7 2 Set the network parameters of the computer 3 Set the network parameters of the OWON Oscilloscope Software Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN About the IP the first three bytes is same as the IP in the step 2 the last byte should be different Here we set it to 192 168 1 7...

Page 123: ...ts Reset to update the config After resetting the oscilloscope if you can get data normally in the oscilloscope software the connection is successful Figure 7 4 Set the network parameters of the oscilloscope Connect through a router 1 Connection Use a LAN cable to connect the oscilloscope with a router the LAN port of the oscilloscope is in the right side panel the computer should be connected to ...

Page 124: ...e computer choose the Ports settings of the Communications menu item Set Connect using to LAN About the IP the first three bytes is same as the IP in the step 2 the last byte should be different Here we set it to 192 168 1 72 The range of the port value is 0 4000 but the port which under 2000 is always used so it is suggested to set it to the value above 2000 Here we set it to 3000 Figure 7 6 Set ...

Page 125: ...ess Point PC should support Wi Fi 1 Set the oscilloscope In the oscilloscope push the Utility button Select Function in the bottom menu Select LAN Set in the left menu In the bottom menu set the Type item as WIFI AP and select Set in the bottom menu 2 In the right menu select SSID a keyboard will pop up You can turn and push the M knob or touch the keyboard if the LCD is touchscreen to set the acc...

Page 126: ...the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN Set IP and Port to the same value in the oscilloscope in the step 4 If you can get data normally the connection is successful Figure 7 8 Set the network parameters of the OWON Oscilloscope Software Connect with PC as Wi Fi Station The oscilloscope and PC should be on the same network LAN...

Page 127: ...In the right menu select Password use the keyboard to input the Wi Fi hotspot password 4 In the right menu select IP PORT Turn the M knob to set the PORT Here we set it to 3000 5 After setting select ON in the Status menu The value of IP is displayed in IP PORT menu Here IP is 10 1 1 126 A changing icon appears on the top right of the screen 6 Select Save set in the bottom menu to save current set...

Page 128: ...illoscope Software Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN Set IP and Port to the same value in the oscilloscope If you can get data normally the connection is successful Figure 7 10 Set the network parameters of the OWON Oscilloscope Software 121 ...

Page 129: ...The help content below may be not completely consistent with the application software only for reference Visit www owon com cn to get the latest version of this user manual How to Connect In Wi Fi settings of the oscilloscope there are two connection types One is WIFI AP Access Point the oscilloscope is set as Wi Fi access point the Android device can be connected to it directly The other type is ...

Page 130: ...lect IP PORT IP is fixed by the oscilloscope and can not be edited Turn the M knob to set the PORT 5 After setting select ON in the Status menu The icon appears on the top right of the screen 6 Select Save set in the bottom menu to save current settings 7 In Android device enter the Wi Fi settings select the oscilloscope access point to connect enter the set password 8 In Android device launch the...

Page 131: ... the right menu select SSID a keyboard will pop up Input the name of the Wi Fi hotspot You can turn and push the M knob or touch the keyboard if the LCD is touchscreen 3 In the right menu select Encryption push to set the Encryption type of the Wi Fi hotspot Open type needs no password WEP or WPA type need to input the password In the right menu select Password use the keyboard to input the Wi Fi ...

Page 132: ...droid device enter the Wi Fi settings select the oscilloscope access point to connect enter the set password 8 In Android device launch the application 9 Click connect on the left 10 Set IP and Port to the same value as the setting in the oscilloscope Click Confirm to connect Click History to recall the history settings 125 ...

Page 133: ... to turn CH2 on off Touch to stop continue to synchronize with the oscilloscope Time Base Horizontal Position Horizontal Position Pointer CH2 Show hide menu see below Trigger Level Share the image wave file via the installed sharing apps Save into wave file bmp png bin csv Exit APP Submit feedback Swipe from the left edge of the screen to show the trigger source menu Trigger Source Menu Trigger So...

Page 134: ...ontrol interface of the generator Turn on off output of selected channel Select channel Select waveform Set the parameters of current waveform Control Interface of the Generator The points will be connected with lines Adjust the brightness of grid Set the auto measurement Display Menu 127 ...

Page 135: ... Cursor Measurement The cursor line can be dragged Check to measure voltage APP Information Enter detail setting Setting Menu Gestures Control Control the vertical position of the selected channel Show the vertical position when gesturing Control the trigger level of the trigger source 128 ...

Page 136: ...ation with Android Device via Wi Fi Optional Control the horizontal position Pinch and spread vertically to change the voltage division of current channel Pinch and spread horizontally to change the time base 129 ...

Page 137: ...e the waveform optimized based on which you can further regulate the vertical and horizontal divisions till the waveform meets your requirement 2 Perform Automatic Measurement The oscilloscope can measure most of the displayed signals automatically To measure the period the frequency of the CH1 following the steps below 1 Push the Measure button to show the Measure menu 2 Select Add in the bottom ...

Page 138: ...e CH2 channel to the output end Operation Steps 1 Push the Autoset button and the oscilloscope will automatically adjust the waveforms of the two channels into the proper display state 2 Push the Measure button to show the Measure menu 3 Select Add in the bottom menu 4 In the right menu select CH1 in the Source menu item 5 In the left Type menu turn the M knob to select PK PK 6 In the right menu s...

Page 139: ...the closest trigger level and trigger type this helps user to make few small adjustments to achieve a proper trigger level and mode Here is how we achieve this The operation steps are as follows 1 Set the probe menu attenuation coefficient to 10X and that of the switch in the probe to 10X see How to Set the Probe Attenuation Coefficient on P12 2 Adjust the Vertical Scale and Horizontal Scale knobs...

Page 140: ...h the Run Stop button and wait a trigger Once there is a burr occurring the instrument will trigger automatically and record the waveform during the period around the trigger time By turning the Horizontal Position knob in the horizontal control area in the panel you can change the horizontal triggering position to obtain the negative delay making an easy observation of the waveform before the bur...

Page 141: ...e level as lower as possible this would enable user to have more details about the signal The Average function offered by our Oscilloscope can help you to achieve this Here are the steps for how to enable Average function 1 Push the Acquire button to display the Acquire menu 2 Select Acqu Mode in the bottom menu 3 Select Average in the right menu turn the M knob and observe the waveform obtained f...

Page 142: ...e operate according to the following steps 1 Set the probe menu attenuation coefficient for 10X and that of the switch in the probe for 10X see How to Set the Probe Attenuation Coefficient on P12 2 Connect the probe of channel 1 to the input of the network and that of Channel 2 to the output of the network 3 Push the Autoset button with the oscilloscope turning on the signals of the two channels a...

Page 143: ...he I and IV quadrants that is in the range of 0 π 2 or 3π 2 2π If the principal axis of the ellipse is in the II and IV quadrants the determined phase difference angle is in the II and III quadrants that is within the range of π 2 π or π 3π 2 Example 6 Video Signal Trigger Observe the video circuit of a television apply the video trigger and obtain the stable video output signal display Video Fiel...

Page 144: ...ight menu 5 Select Modu in the bottom menu Select NTSC in the right menu 6 Select Sync in the bottom menu Select Field in the right menu 7 Turn the Vertical Scale Vertical Position and Horizontal Scale knobs to obtain a proper waveform display see Figure 9 7 Figure 9 7 Waveform Captured from Video Field Trigger 137 ...

Page 145: ...Probe Attenuation Coefficient on P12 4 There is a waveform displayed but it is not stable Check whether the Source item in the TRIG MODE menu is in conformity with the signal channel used in the practical application Check on the trigger Type item The common signal chooses the Edge trigger mode for Type and the video signal the Video If Alternate trigger is selected both of the channel 1 and chann...

Page 146: ...see How to Implement Self calibration on P14 All specification standards can be fulfilled except one s marked with the word Typical Oscilloscope Performance Characteristics Instruction Bandwidth XDS3062A 60 MHz XDS3102 A 100 MHz XDS3102AP 8 bits mode 12 bits mode 100 MHz 14 bits mode 25 MHz XDS3202E 200 MHz XDS3202A 8 bits mode 200 MHz 12 bits mode 150 MHz 14 bits mode 20 MHz XDS3202 200 MHz XDS33...

Page 147: ...Mode Normal Peak detect Averaging Sample rate real time XDS3062A XDS3102A Dual CH 500 MS s Single CH 8 bits mode 1 GS s 12 bits mode 500 MS s XDS3102 Dual CH 500 MS s Single CH 1 GS s XDS3202E Dual CH 500 MS s Single CH 1 GS s XDS3102AP XDS3202A Dual CH 8 bits mode 1 GS s 12 bits mode 500 MS s 14 bits mode 100 MS s Single CH 8 bits mode 1 GS s 12 bits mode 500 MS s 14 bits mode 100 MS s XDS3202 Du...

Page 148: ...S s Single CH 8 bits mode 0 05 S s 1 GS s 12 bits mode 0 05 S s 500 MS s XDS3102 Dual CH 0 05 S s 500 MS s Single CH 0 05 S s 1 GS s XDS3202E Dual CH 0 05 S s 500 MS s Single CH 0 05 S s 1 GS s XDS3102AP XDS3202A Dual CH 8 bits mode 0 05 S s 1 GS s 12 bits mode 0 05 S s 500 MS s 14 bits mode 0 05 S s 100 MS s Single CH 8 bits mode 0 05 S s 1 GS s 12 bits mode 0 05 S s 500 MS s 14 bits mode 0 05 S ...

Page 149: ... XDS3102AP XDS3202A 2 V 1 mV div 50 mV div 20 V 100 mV div 1 V div 200 V 2 V div 10 V div XDS3202E XDS3202 XDS3302 2 V 1 mV div 50 mV div 20 V 100 mV div 500 mV div 250 V 1 V div 10 V div Analog bandwidth 60 MHz 100 MHz 200 MHz 300 MHz Single bandwidth Full bandwidth Low Frequency 10 Hz at input AC coupling 3 dB Rise time at input Typical XDS3062A 5 8 ns XDS3102A 3 5 ns XDS3102AP 8 bits mode 12 bi...

Page 150: ...Cursor RMS Screen Duty Phase Pulse Count Pulse Count Rise Edge Count Fall Edge Count Area and Cycle Area Waveform Math FFT FFTrms Intg Diff Sqrt User Defined Function digital filter low pass high pass band pass band reject Decoding Type optional RS232 I2C SPI CAN Waveform storage 100 waveforms Lissajous figure Bandwidth Full bandwidth Phase difference 3 degrees Communication port Standard USB USB ...

Page 151: ...n Positive pulse Negative pulse Time setting 30 ns to 10 s For XDS3102AP XDS3202A the range is 8 bits mode 2 ns to 10 s 12 bits mode 4 ns to 10 s 14 bits mode 20 ns to 10 s Runt Trigger Polarity Positive Negative Pulse Width Condition Pulse Width Range 30 ns to 10 s For XDS3102AP XDS3202A the range is 8 bits mode 2 ns to 10 s 12 bits mode 4 ns to 10 s 14 bits mode 20 ns to 10 s Windows Trigger Pol...

Page 152: ... 255 0 to 1023 Byte Length 1 to 5 SPI Trigger Trigger Condition Timeout Timeout Value 30 ns to 10 s Data Bits 4 bit to 32 bit Data Line Setting H L X CAN Trigger optional Signal Type CAN_H CAN_L TX RX Trigger Condition Start of Frame Type of Frame Identifier Data ID Data End of Frame Missing Ack Bit Stuffing Error Baud Rate Common Custom Sample Point 5 to 95 Frame Type Data Remote Error Overload W...

Page 153: ...ment 50 30 Ω beeping Capacitance 51 2nF 100uF 3 3 digit Voltage DCV 400mV 4V 40V 400V 1000V 1 1digit Max input DC 1000V ACV 400mV 4V 40V 400V 1 3digit 750V 1 5 3digit Frequency 40Hz 400Hz Max input AC 750V virtual value Current DCA 40mA 400mA 1 5 1 digit 4A 10A 3 3digit ACA 40mA 1 5 3digit 400mA 2 1digit 4A 10A 3 3digit Impedance 400Ω 1 3digit 4KΩ 4MΩ 1 1digit 40MΩ 1 5 3digit 146 ...

Page 154: ... 15 W XDS3102AP XDS3202E XDS3202 A XDS3302 24 W Fuse 2 A T class 250 V Battery optional 3 7V 13200mAh On XDS3062A XDS3102 A the battery can last approximately 3 5 hours after charge fully On XDS3202 E XDS3302 the battery can last approximately 2 5 hours after charge fully Environment Temperature Working temperature 0 40 Storage temperature 20 60 Relative Humidity 90 Height Operating 3 000 m Non op...

Page 155: ...ment where the liquid crystal display will be exposed to direct sunlight for long periods of time Caution To avoid any damage to the instrument or probe do not exposed it to any sprays liquids or solvents Cleaning Inspect the instrument and probes as often as operating conditions require To clean the instrument exterior perform the following steps 1 Wipe the dust from the instrument and probe surf...

Page 156: ...Battery Using Guide Figure 12 1 Battery level indicator Charging the oscilloscope There will have battery power indication show on the top of panel when oscilloscope power supplied by the battery if no indication has appeared see Display on P81 indicates the battery is nearly used up Note To avoid overheat of battery during charging the ambient temperature is not allowed to exceed the permissible ...