3.Functions and Operations
2.
Connect the test lead.
Open or Closed Circuit
I
3.
Set the beeper.
Press the Beeper softkey to enable or disable the beeper. When the beeper is
enabled, the reading is below
30 Ω, the multimeter will beep continuously.
4.
Set the short-circuit resistance.
Press the Threshold softkey to set the short-circuit resistance.
Press
to move the cursor, press
to increase or decrease the value.The
range for the
1 Ω to
10
00 Ω
. The default is 5
0 Ω.
5.
Continuity measurements behave as follows:
Circuit resistance to be measured
Display and beep
≤ Short
-circuit resistance
Displays measured resistance and beeps (if
beeper enabled)
Short-circuit resistance to 1000
Ω
Displays measured resistance without
beeping
> 1000
Ω
Displays "Open" with no beep
Diode Test
This section describes how to configure diode test.
Operating Steps:
1.
Enable the diode test.
Press
on the front panel, press it twice to enter diode test mode.
20