4.Advanced User Guidebook
32
Select pulse width condition and click
or
to set time, click
to move cursor to choose
which digit to be set.
Mode
Holdoff
Auto
Acquire waveform even no trigger occurred
Normal
Acquire waveform when trigger occurred
Single
When trigger occurs, acquire one waveform then
stop
Holdoff
100 ns - 10 s, click
or
to set time interval
before another trigger occur, click
to
move cursor to choose which digit to be set.
Sensitivity
Set the trigger sensitivity.
4. Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within
the specified time.
In Slope Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example,
, indicates that trigger
type is slope, trigger source is CH1, slope is rising, 0.00mV is the differential
between up level and low level threshold.
Slope trigger menu
list:
Menu
Setting
Instruction
Single
Slope
Set vertical channel trigger type as slope
trigger.
Source
CH1
CH2
Select CH1 as the trigger source
Select CH2 as the trigger source
When
slope
Slope selecting
Set slope condition; click
or
to set
slope time, click
to move cursor to
choose which digit to be set.
Threshold
&SlewRate
High level
Low level
Slew rate
Click
or
to set the High level upper limit.
Click
or
to set Low level lower limit.
Slew rate = (High level - Low level) / Settings
Mode
Auto
Acquire waveform even no trigger occurred
Normal
Acquire waveform when trigger occurred
Summary of Contents for TAO3000 Dual-Channel Series
Page 1: ...TAO3000 Dual Channel Series Tablet Oscilloscopes User Manual www owon com ...
Page 93: ...4 Advanced User Guidebook 87 Figure 4 24 Hamming window Figure 4 25 Rectangle window ...
Page 94: ...4 Advanced User Guidebook 88 Figure 4 26 Blackman window Figure 4 27 Hanning window ...
Page 95: ...4 Advanced User Guidebook 89 Figure 4 28 Kaiser window Figure 4 29 Bartlett window ...